scispace - formally typeset
Search or ask a question
Author

C. Varatharaj

Bio: C. Varatharaj is an academic researcher from Kidwai Memorial Institute of Oncology. The author has contributed to research in topics: Multileaf collimator & Dosimetry. The author has an hindex of 4, co-authored 14 publications receiving 50 citations.

Papers
More filters
Journal ArticleDOI
TL;DR: Among all detectors used in the study, the unshielded diodes were found to be an appropriate choice of detector for the measurement of beam parameters in small fields.
Abstract: The advent of modern technologies in radiotherapy poses an increased challenge in the determination of dosimetric parameters of small fields that exhibit a high degree of uncertainty. Percent depth dose and beam profiles were acquired using different detectors in two different orientations. The parameters such as relative surface dose (D S), depth of dose maximum (D max), percentage dose at 10 cm (D 10), penumbral width, flatness, and symmetry were evaluated with different detectors. The dosimetric data were acquired for fields defined by jaws alone, multileaf collimator (MLC) alone, and by MLC while the jaws were positioned at 0, 0.25, 0.5, and 1.0 cm away from MLC leaf-end using a Varian linear accelerator with 6 MV photon beam. The accuracy in the measurement of dosimetric parameters with various detectors for three different field definitions was evaluated. The relative D S(38.1%) with photon field diode in parallel orientation was higher than electron field diode (EFD) (27.9%) values for 1 cm ×1 cm field. An overestimation of 5.7% and 8.6% in D 10 depth were observed for 1 cm ×1 cm field with RK ion chamber in parallel and perpendicular orientation, respectively, for the fields defined by MLC while jaw positioned at the edge of the field when compared to EFD values in parallel orientation. For this field definition, the in-plane penumbral widths obtained with ion chamber in parallel and perpendicular orientation were 3.9 mm, 5.6 mm for 1 cm ×1 cm field, respectively. Among all detectors used in the study, the unshielded diodes were found to be an appropriate choice of detector for the measurement of beam parameters in small fields.

21 citations

Journal ArticleDOI
TL;DR: Due to simplicity and fast evaluation process of array detectors, it can be routinely used in busy departments without compromising the measurement accuracy.
Abstract: Background: The evaluation of the agreement between measured and calculated dose plays an essential role in the quality assurance (QA) procedures of intensity-modulated radiation therapy (IMRT). Aim: The purpose of this study is to compare performances of the two dosimetric systems (EDR2 and I'matriXX) in the verification of the dose distributions calculated by the TPS for brain and head and neck dynamic IMRT cases. Materials and Methods: The comparison of cumulative fluence by using Kodak extended dose rate (EDR2) and I'matriXX detectors has been done for the evaluation of 10 brain, 10 head and neck IMRT cases treated with 6 MV beams. The parameter used to assess the quality of dose calculation is the gamma-index (g -index) method. The acceptance limits for g calculation we have used are 3% and 3 mm respectively for dose agreement and distance to agreement parameters. Statistical analyses were performed by using the paired, two-tailed Student t-test, and P Results: The qualitative dose distribution comparison was performed using composite dose distribution in the measurement plane and profiles along various axes for TPS vs. EDR2 film and TPS Vs I'matriXX. The quantitative analysis between the calculated and measured dose distribution was evaluated using DTA and g-index. The percentage of pixels matching with the set DTA and g values are comparable for both with EDR2 film and I'matriXX array detectors. Statistically there was no significant variation observed between EDR2 film and I'matriXX in terms of the mean percentage of pixel passing g for brain cases (98.77 ± 1.03 vs 97.62 ± 1.66, P = 0.0218) and for head and neck cases (97.39 ± 2.13 vs 97.17 ± 1.52%, P = 0.7404). Conclusion: Due to simplicity and fast evaluation process of array detectors, it can be routinely used in busy departments without compromising the measurement accuracy.

7 citations

Journal ArticleDOI
TL;DR: The newly developed portal dosimetry software using independent dose prediction algorithm EPIDose™ is implemented and this new tool for the pre-treatment IMRT plan quality assurance of Whole Pelvis with Simultaneous Integrated Boost (WP-SIB-IMRT) of prostate cases is evaluated by comparing with routine two-dimensional array detector system (MapCHECK™).
Abstract: Electronic portal imaging devices (EPIDs) are extensively used for obtaining dosimetric information of pre-treatment field verification and in-vivo dosimetry for intensity-modulated radiotherapy (IMRT). In the present study, we have implemented the newly developed portal dosimetry software using independent dose prediction algorithm EPIDose TM and evaluated this new tool for the pre-treatment IMRT plan quality assurance of Whole Pelvis with Simultaneous Integrated Boost (WP-SIB-IMRT ) of prostate cases by comparing with routine two-dimensional (2D) array detector system (MapCHECK TM ). We have investigated 104 split fields using g-distributions in terms of predefined g frequency parameters. The mean γ values are found to be 0.42 (SD: 0.06) and 0.44 (SD: 0.06) for the EPIDose and MapCHECK TM , respectively. The average g∆ for EPIDose and MapCHECK TM are found as 0.51 (SD: 0.06) and 0.53 (SD: 0.07), respectively. Furthermore, the percentage of points with g 2 are 97.4%, 99.3%, and 0.56%, respectively for EPIDose and 96.4%, 99.0% and 0.62% for MapCHECK TM . Based on our results obtained with EPIDose and strong agreement with MapCHECK TM , we may conclude that the EPIDose portal dosimetry system has been successfully implemented and validated with our routine 2D array detector

6 citations

Journal ArticleDOI
TL;DR: The aim of this study was to optimize dose distribution for Gammamed plus vaginal cylinders and found that the effect of source travel step size on the optimized dose distributions for vaginal cylinders was also evaluated.

5 citations

Journal ArticleDOI
TL;DR: The purpose of this study was to compare the dosimetric characteristics of a physical and enhanced dynamic wedge from a dual-energy (6 and 18 MV) linear accelerator such as surfaces doses with different source to surface distances (SSD), half value layer (HVL) in water and peripheral doses for both available energies.
Abstract: The use of megavoltage X-ray sources of radiation, with their skin-sparing qualities in radiation therapy, has been proved useful in relieving patient discomfort and allowing higher tumor doses to be given with fewer restrictions due to radiation effects in the skin. The purpose of this study was to compare the dosimetric characteristics of a physical and enhanced dynamic wedge from a dual-energy (6 and 18 MV) linear accelerator such as surfaces doses with different source to surface distances (SSD), half value layer (HVL) in water and peripheral doses for both available energies. At short SSD such as 85 cm, higher surface doses are produced by the lower wedges by the short wedge-to-skin distance. For physical wedged field, at heel edge side HVL value was high (17 cm) compared with the measured that of EDW (15.1 cm). It was noticed that, the HVL variation across the beam was significantly higher for 6 MV X-rays than for 18 MV X-rays. The lower wedge has the maximum variation of peripheral dose compared to other wedges. The three wedge systems discussed in this work possess vastly different dosimetric characteristics. These differences will have a direct impact on the choice of the wedge system to be used for a particular treatment. Complete knowledge of the dosimetric characterisitics, including the surface and peripheral doses, is crucial in proper choice of particular wedge systems in clinical use.

4 citations


Cited by
More filters
Journal Article
TL;DR: It is demonstrated that film dosimetry for IMRT involves sources of error due to its over-response to low-energy photons, with the error most transparent in the low-dose region.
Abstract: X-ray film has been used for the dosimetry of intensity modulated radiation therapy (IMRT). However, the over-response of the film to low-energy photons is a significant problem in photon beam dosimetry, especially in regions outside penumbra. In IMRT, the radiation field consists of multiple small fields and their outside-penumbra regions; thus, the film dosimetry, for it involves the source of over-response in its radiation field. In this study we aim to verify and possibly improve film dosimetry for IMRT. Two types of modulated beams were constructed by combining five to seven different static radiation fields using 6 MV x rays. For verifying film dosimetry, x-ray films and an ion chamber were used to measure dose profiles at various depths in a phantom. The film setups include both parallel and perpendicular arrangements against the beam incident direction. In addition, to reduce an over-response, we placed 0.01 in. (0.25 mm) thick lead filters on both sides of the film. Compared with ion-chamber measurement, measured dose profiles showed the film over-response at outside-penumbra and low-dose regions. The error increased with depths and approached 15% as a maximum for the field size of 15 cm x 15 cm at 10 cm depth. The use of filters reduced the error down to 3%. In this study we demonstrated that film dosimetry for IMRT involves sources of error due to its over-response to low-energy photons, with the error most transparent in the low-dose region. The use of filters could enhance the accuracy in film dosimetry for IMRT. In this regard, the use of an optimal filter condition is recommended.

94 citations

Journal ArticleDOI
TL;DR: The results indicate that the use of the EPID, combined with MLC log files and a TPS, is a viable method for QA of VMAT plans.
Abstract: In this project, we investigated the use of an electronic portal imaging device (EPID), together with the treatment planning system (TPS) and MLC log files, to determine the delivered doses to the patient and evaluate the agreement between the treatment plan and the delivered dose distribution. The QA analysis results are presented for 15 VMAT patients using the EPID measurements, the ScandiDos Delta4 dosimeter, and the beam fluence calculated from the multileaf collimator (MLC) log file. EPID fluence images were acquired in continuous acquisition mode for each of the patients and they were processed through an in-house MATLAB program to create an opening density matrix (ODM), which was used as the input fluence for the dose calculation in the TPS (Pinnacle3). The EPID used in this study was the aSi1000 Varian on a Novalis TX linac equipped with high-definition MLC. The actual MLC positions and gantry angles were retrieved from the MLC log files and the data were used to calculate the delivered dose distributions in Pinnacle. The resulting dose distributions were then compared against the corresponding planned dose distributions using the 3D gamma index with 3 mm/3% passing criteria. The ScandiDos Delta4 phantom was also used to measure a 2D dose distribution for all the 15 patients and a 2D gamma was calculated for each patient using the Delta4 software. The average 3D gamma using the EPID images was 96.1% ± 2.2%. The average 3D gamma using the log files was 98.7% ± 0.5%. The average 2D gamma from the Delta4 was 98.1% ± 2.1%. Our results indicate that the use of the EPID, combined with MLC log files and a TPS, is a viable method for QA of VMAT plans.

35 citations

Journal ArticleDOI
TL;DR: Vaginal HDR brachytherapy using a multichannel vaginal applicator and inverse planning provides dosimetric advantages over single channel cylinder, by reducing the dose to organs at risk without compromising the target volume coverage, but at the expense of an increased vaginal mucosa dose.
Abstract: Purpose: To retrospectively compare the potential dosimetric advantages of a multichannel vaginal applicator vs. a single channel one in intracavitary vaginal high-dose-rate (HDR) brachytherapy after hysterectomy, and evaluate the dosimetric advantage of fractional re-planning. Material and methods: We randomly selected 12 patients with endometrial carcinoma, who received adjuvant vaginal cuff HDR brachytherapy using a multichannel applicator. For each brachytherapy fraction, two inverse treatment plans (for central channel and multichannel loadings) were performed and compared. The advantage of fractional re-planning was also investigated. Results: Dose-volume-histogram (DVH) analysis showed limited, but statistically significant difference (p = 0.007) regarding clinical-target-volume dose coverage between single and multichannel approaches. For the organs-at-risk rectum and bladder, the use of multichannel applicator demonstrated a noticeable dose reduction, when compared to single channel, but statistically significant for rectum only (p = 0.0001). For D 2cc of rectum, an average fractional dose of 6.1 ± 0.7 Gy resulted for single channel vs. 5.1 ± 0.6 Gy for multichannel. For D 2cc of bladder, an average fractional dose of 5 ± 0.9 Gy occurred for single channel vs. 4.9 ± 0.8 Gy for multichannel. The dosimetric benefit of fractional re-planning was demonstrated: DVH analysis showed large, but not statistically significant differences between first fraction plan and fractional re-planning, due to large inter-fraction variations for rectum and bladder positioning and filling. Conclusions: Vaginal HDR brachytherapy using a multichannel vaginal applicator and inverse planning provides dosimetric advantages over single channel cylinder, by reducing the dose to organs at risk without compromising the target volume coverage, but at the expense of an increased vaginal mucosa dose. Due to large inter-fraction dose variations, we recommend individual fraction treatment plan optimization. J Contemp Brachytherapy 2014; 6, 4: 362–370 DOI: 10.5114/jcb.2014.47816

24 citations

Journal ArticleDOI
TL;DR: The results of this investigation found that the uncorrected transit doses lead to small errors in the delivered dose at the first dwell position, but the transit dose correction for other dwells was accurate within 0.2 cGy.
Abstract: A comprehensive system characterisation was performed of the Eckert & Ziegler BEBIG GmbH MultiSource® High Dose Rate (HDR) brachytherapy treatment unit with an 192Ir source. The unit is relatively new to the UK market, with the first installation in the country having been made in the summer of 2009. A detailed commissioning programme was devised and is reported including checks of the fundamental parameters of source positioning, dwell timing, transit doses and absolute dosimetry of the source. Well chamber measurements, autoradiography and video camera analysis techniques were all employed. The absolute dosimetry was verified by the National Physical Laboratory, UK, and compared to a measurement based on a calibration from PTB, Germany, and the supplied source certificate, as well as an independent assessment by a visiting UK centre. The use of the 'Krieger' dosimetry phantom has also been evaluated. Users of the BEBIG HDR system should take care to avoid any significant bend in the transfer tube, as this will lead to positioning errors of the source, of up to 1.0 mm for slight bends, 2.0 mm for moderate bends and 5.0 mm for extreme curvature (depending on applicators and transfer tube used) for the situations reported in this study. The reason for these errors and the potential clinical impact are discussed. Users should also note the methodology employed by the system for correction of transit doses, and that no correction is made for the initial and final transit doses. The results of this investigation found that the uncorrected transit doses lead to small errors in the delivered dose at the first dwell position, of up to 2.5 cGy at 2 cm (5.6 cGy at 1 cm) from a 10 Ci source, but the transit dose correction for other dwells was accurate within 0.2 cGy. The unit has been mechanically reliable, and source positioning accuracy and dwell timing have been reproducible, with overall performance similar to other existing HDR equipment. The unit is capable of high quality brachytherapy treatment delivery, taking the above factors into account.

24 citations

Journal ArticleDOI
TL;DR: It is confirmed that conventional IMRT QA performance metrics are not predictive of dose errors in PTV and organs-at-risk and the clinically-relevant-dose QA has allowed us to predict the patient dose-volume relationships.
Abstract: Recent studies have demonstrated that per-beam planar intensity-modulated radiation therapy (IMRT) quality assurance (QA) passing rates may not predict clinically relevant patient dose errors. This work is to evaluate the effect of dose variations introduced in dynamic multi-leaf collimator (DMLC) modeling and delivery processes on clinically relevant metrics for IMRT. Ten head and neck (HN) IMRT plans were randomly selected for this study. The conventional per-beam IMRT QA was performed for each plan by 2 different methods: (1) with gantry angle of 0 (gantry pointing downward) for all IMRT fields and (2) with gantry at specific angles as designed in the IMRT plan. For each patient, a batch analysis was done for each scenario and then imported to the 3DVH (Sun Nuclear Corp.) for processing. A "corrected DVH" was generated and compared to the DVH from the treatment plan. Their differences represented errors introduced from the combination of the treatment planning system (TPS) dose calculation algorithm and beam-delivery. The dose metrics from the two scenarios were compared with the corresponding calculated doses, and then their differences were analyzed. Although all per-beam planar IMRT QA had high Gamma passing rates 99.3 ± 1.3% (92.3-100%) for "2%/3 mm" criteria, there were significant errors in some of the calculated clinical dose metrics. Such as, for all the plans studied, there were as much as 3.2%, 5.7%, 5.6%, 2.3%, 4.1%, and 23.8% errors found in max cord dose, max brainstem dose, mean parotid dose, larynx dose, oral cavity dose, and PTV(D95) dose, respectively. The differences in errors for clinical metrics obtained between the two scenarios (zero gantry angle vs. true gantry angles) can also be significant: max cord dose (2.9% vs. 0.2%), max brainstem dose (3.8% vs. 0.4%), mean parotid dose (2.3% vs. 4.5%), mean larynx dose (3.9% vs. 2.0%), mean oral cavity dose (1.6% vs. 3.9%), and PTV(D95) dose (-0.4% vs. -2.6%). However, in the two scenarios, a strong and clear correlation between the dose differences for each of the organ structures was observed. This study confirms that conventional IMRT QA performance metrics are not predictive of dose errors in PTV and organs-at-risk. The clinically-relevant-dose QA has allowed us to predict the patient dose-volume relationships.

23 citations