scispace - formally typeset
Journal ArticleDOI

A high-efficiency real-time digital signal averager for time-of-flight mass spectrometry

Reads0
Chats0
TLDR
This work presents a high-efficiency real-time digital signal averager, consisting of a signal conditioner, a data conversion module and a signal processing module, combined with a vacuum ultraviolet single-photon ionization time-of-flight mass spectrometer (VUV-SPI-TOFMS).
Abstract
RATIONALE Analog-to-digital converter (ADC)-based acquisition systems are widely applied in time-of-flight mass spectrometers (TOFMS) due to their ability to record the signal intensity of all ions within the same pulse. However, the acquisition system raises the requirement for data throughput, along with increasing the conversion rate and resolution of the ADC. It is therefore of considerable interest to develop a high-performance real-time acquisition system, which can relieve the limitation of data throughput. METHODS We present in this work a high-efficiency real-time digital signal averager, consisting of a signal conditioner, a data conversion module and a signal processing module. Two optimization strategies are implemented using field programmable gate arrays (FPGAs) to enhance the efficiency of the real-time processing. A pipeline procedure is used to reduce the time consumption of the accumulation strategy. To realize continuous data transfer, a high-efficiency transmission strategy is developed, based on a ping-pong procedure. RESULTS The digital signal averager features good responsiveness, analog bandwidth and dynamic performance. The optimal effective number of bits reaches 6.7 bits. For a 32 μs record length, the averager can realize 100% efficiency with an extraction frequency below 31.23 kHz by modifying the number of accumulation steps. In unit time, the averager yields superior signal-to-noise ratio (SNR) compared with data accumulation in a computer. CONCLUSIONS The digital signal averager is combined with a vacuum ultraviolet single-photon ionization time-of-flight mass spectrometer (VUV-SPI-TOFMS). The efficiency of the real-time processing is tested by analyzing the volatile organic compounds (VOCs) from ordinary printed materials. In these experiments, 22 kinds of compounds are detected, and the dynamic range exceeds 3 orders of magnitude.

read more

Citations
More filters
Journal ArticleDOI

Development of a miniature time-of-flight mass/charge spectrometer for ion beam source analyzing

TL;DR: In this article, a miniature time-of-flight mass spectrometer (TOFMS) with compact structure was developed for cathodic vacuum arc discharge ion source, which could be used to diagnose the ion discharge process.
Journal ArticleDOI

The Analog Revolution and Its On-Going Role in Modern Analytical Measurements.

TL;DR: The multiple roles op amps play in modern instrumentation are described along with a number of examples of new areas of analysis that have been enabled by their improvements.
References
More filters
Book

Discrete-Time Signal Processing

TL;DR: In this paper, the authors provide a thorough treatment of the fundamental theorems and properties of discrete-time linear systems, filtering, sampling, and discrete time Fourier analysis.
Journal ArticleDOI

Analog-to-digital converter survey and analysis

TL;DR: The state-of-the-art of ADCs is surveyed, including experimental converters and commercially available parts, and the distribution of resolution versus sampling rate provides insight into ADC performance limitations.
Journal ArticleDOI

An introduction to quadrupole-time-of-flight mass spectrometry.

TL;DR: A brief introduction to the basic principles and application of a quadrupole-time-of-flight (TOF) tandem mass spectrometer and basic algorithms for liquid chromatography/MS/MS automation are discussed and illustrated by two applications.
Book

Design with Operational Amplifiers and Analog Integrated Circuits

TL;DR: In this article, the authors present a design-oriented course for operational amplifiers and analog ICs, which includes additional problems, more in-depth coverage of negative feedback, more effective layout, updated technology (current feedback and folded-cascode amplifiers, and low-voltage amplifiers), and increased topical coverage (current-feedback amplifier, switching regulators and phase-locked loops).
Journal ArticleDOI

A New Time-of-Flight Aerosol Mass Spectrometer (TOF-AMS)—Instrument Description and First Field Deployment

TL;DR: In this article, the authors report the development and first field deployment of a new version of the AMS, which is capable of measuring non-refractory aerosol mass concentrations, chemically speciated mass distributions and single particle information.
Related Papers (5)