scispace - formally typeset
Journal ArticleDOI

Digital Wavefront Measuring Interferometer for Testing Optical Surfaces and Lenses

Reads0
Chats0
TLDR
The system has been designed to optimize the collection of data to give higher than usual accuracy in measuring the individual elements and final performance of assembled diffraction limited optical systems, and furthermore, the short loop time of a few minutes makes the system an attractive alternative to constraints imposed by test glasses in the optical shop.
Abstract
A self-scanned 1024 element photodiode array and minicomputer are used to measure the phase (wavefront) in the interference pattern of an interferometer to lambda/100. The photodiode array samples intensities over a 32 x 32 matrix in the interference pattern as the length of the reference arm is varied piezoelectrically. Using these data the minicomputer synchronously detects the phase at each of the 1024 points by a Fourier series method and displays the wavefront in contour and perspective plot on a storage oscilloscope in less than 1 min (Bruning et al. Paper WE16, OSA Annual Meeting, Oct. 1972). The array of intensities is sampled and averaged many times in a random fashion so that the effects of air turbulence, vibrations, and thermal drifts are minimized. Very significant is the fact that wavefront errors in the interferometer are easily determined and may be automatically subtracted from current or subsequent wavefrots. Various programs supporting the measurement system include software for determining the aperture boundary, sum and difference of wavefronts, removal or insertion of tilt and focus errors, and routines for spatial manipulation of wavefronts. FFT programs transform wavefront data into point spread function and modulus and phase of the optical transfer function of lenses. Display programs plot these functions in contour and perspective. The system has been designed to optimize the collection of data to give higher than usual accuracy in measuring the individual elements and final performance of assembled diffraction limited optical systems, and furthermore, the short loop time of a few minutes makes the system an attractive alternative to constraints imposed by test glasses in the optical shop.

read more

Citations
More filters
Journal ArticleDOI

Phase-shifting digital holography

TL;DR: A new method is proposed in which the distribution of complex amplitude at a plane is measured by phase-shifting interferometry and then Fresnel transformed by a digital computer, which can reconstruct an arbitrary cross section of a three-dimensional object with higher image quality and a wider viewing angle than from conventional digital holography using an off-axis configuration.
Book ChapterDOI

V Phase-Measurement Interferometry Techniques

TL;DR: The phase modulation in an interferometer can be induced by moving a mirror, tilting a glass plate, moving a grating, rotating a half-wave plate or analyzer, using an acoustooptic or electro-optic modulator, or using a Zeeman laser as mentioned in this paper.
Journal ArticleDOI

Automated phase-measuring profilometry of 3-D diffuse objects

TL;DR: The high resolution surface profile of a 3-D diffuse object is obtained by measurement of the phase distribution across the image of a projected sinusoidal grating deformed by the surface, based on phase-shifting interferometric techniques.
Journal ArticleDOI

Digital phase-shifting interferometry: a simple error-compensating phase calculation algorithm.

TL;DR: La difference de phase entre les 2 faisceaux interferant varie de maniere connue et on fait des mesures de the distribution d'intensite a travers la pupille correspondant a au moins 3 dephasages differents.
Journal ArticleDOI

Digital wave-front measuring interferometry: some systematic error sources.

TL;DR: To study the occurrence of wave-front irregularities caused by dust particles a model has been developed and countermeasures derived which assure sufficient regularity of contour line plots, and the repeatability of the present experimental setup was better than λ/200 within the 3σ limits.
References
More filters
Related Papers (5)