Electrochemical Detection for Paper-Based Microfluidics

01 Jun 2009-Analytical Chemistry (American Chemical Society)-Vol. 81, Iss: 14, pp 5821-5826
TL;DR: This study shows the successful integration of paper-based microfluidics and electrochemical detection as an easy-to-use, inexpensive, and portable alternative for point of care monitoring.
Abstract: We report the first demonstration of electrochemical detection for paper-based microfluidic devices. Photolithography was used to make microfluidic channels on filter paper, and screen-printing technology was used to fabricate electrodes on the paper-based microfluidic devices. Screen-printed electrodes on paper were characterized using cyclic voltammetry to demonstrate the basic electrochemical performance of the system. The utility of our devices was then demonstrated with the determination of glucose, lactate, and uric acid in biological samples using oxidase enzyme (glucose oxidase, lactate oxidase, and uricase, respectively) reactions. Oxidase enzyme reactions produce H2O2 while decomposing their respective substrates, and therefore a single electrode type is needed for detection of multiple species. Selectivity of the working electrode for H2O2 was improved using Prussian Blue as a redox mediator. The determination of glucose, lactate, and uric acid in control serum samples was performed using chron...
Citations
More filters
Journal ArticleDOI
TL;DR: Microfluidic paper-based analytical devices are a new class of point-of-care diagnostic devices that are inexpensive, easy to use, and designed specifically for use in developing countries.
Abstract: Microfluidic paper-based analytical devices (μPADs) are a new class of point-of-care diagnostic devices that are inexpensive, easy to use, and designed specifically for use in developing countries. (To listen to a podcast about this feature, please go to the Analytical Chemistry multimedia page at pubs.acs.org/page/ancham/audio/index.html.)

2,373 citations

Journal ArticleDOI
TL;DR: This review includes challenges to scaling up, commercialisation and regulatory issues, and the factors which limit paper-based microfluidic devices to become real world products and future directions are also identified.
Abstract: Dipstick and lateral-flow formats have dominated rapid diagnostics over the last three decades. These formats gained popularity in the consumer markets due to their compactness, portability and facile interpretation without external instrumentation. However, lack of quantitation in measurements has challenged the demand of existing assay formats in consumer markets. Recently, paper-based microfluidics has emerged as a multiplexable point-of-care platform which might transcend the capabilities of existing assays in resource-limited settings. However, paper-based microfluidics can enable fluid handling and quantitative analysis for potential applications in healthcare, veterinary medicine, environmental monitoring and food safety. Currently, in its early development stages, paper-based microfluidics is considered a low-cost, lightweight, and disposable technology. The aim of this review is to discuss: (1) fabrication of paper-based microfluidic devices, (2) functionalisation of microfluidic components to increase the capabilities and the performance, (3) introduction of existing detection techniques to the paper platform and (4) exploration of extracting quantitative readouts via handheld devices and camera phones. Additionally, this review includes challenges to scaling up, commercialisation and regulatory issues. The factors which limit paper-based microfluidic devices to become real world products and future directions are also identified.

1,658 citations

Journal ArticleDOI
TL;DR: The goal of this study is to review the fundamental structures and chemistries of wood and wood-derived materials, which are essential for a wide range of existing and new enabling technologies.
Abstract: With the arising of global climate change and resource shortage, in recent years, increased attention has been paid to environmentally friendly materials. Trees are sustainable and renewable materials, which give us shelter and oxygen and remove carbon dioxide from the atmosphere. Trees are a primary resource that human society depends upon every day, for example, homes, heating, furniture, and aircraft. Wood from trees gives us paper, cardboard, and medical supplies, thus impacting our homes, school, work, and play. All of the above-mentioned applications have been well developed over the past thousands of years. However, trees and wood have much more to offer us as advanced materials, impacting emerging high-tech fields, such as bioengineering, flexible electronics, and clean energy. Wood naturally has a hierarchical structure, composed of well-oriented microfibers and tracheids for water, ion, and oxygen transportation during metabolism. At higher magnification, the walls of fiber cells have an interes...

1,031 citations

Journal ArticleDOI
TL;DR: This chapter discusses the development of personalized medicine and home testing in the developing world, and some of the strategies used to achieve this goal have not yet been developed.
Abstract: Introduction A Why POC Diagnostics? B Time B Patient Responsibility and Compliance B Cost B Diagnostic Targets C Proteins C Metabolites and Other Small Molecules C Nucleic Acids C Human Cells D Microbes/Pathogens D Drugs and Food Safety D Current Context of POC Assays E POC Glucose Assays E Lateral Flow Assays E Limitations of “Traditional” POC Approaches F Enabling Technologies G Printing and Laminating G Microfluidic Technologies and Approaches: “Unit Operations” for POC Devices G Pumping and Valving H Mixing I Separation I Reagent Storage J Sample Preparation K Surface Chemistry and Device Substrates L Physical Adsorption L Bioaffinity Attachment L Covalent Attachment M Substrate Materials M Detection M Electrochemical Detection N Optical Detection N Magnetic Detection N Label-Free Methods O Enabling Multiplexed Assays O Recent Innovation O Lateral Flow Assay Technologies O Proteins P Antibodies P Protein Expression and Purification Q Nucleic Acids Q Aptamers R Infectious Diseases and Food/Water Safety R Blood Chemistry S Coagulation Markers S Whole Cells S Trends, Unmet Needs, Perspectives T Glucose T Global Health and the Developing World T Personalized Medicine and Home Testing U Technology Trends U Multiplexing V Author Information V Biographies V Acknowledgment W References W

983 citations

References
More filters
Journal ArticleDOI
TL;DR: This communication describes a simple method for patterning paper to create well-defined, millimeter-sized channels, comprising hydrophilic paper bounded by hydrophobic polymer, that will become the basis for low-cost, portable, and technically simple multiplexed bioassays.
Abstract: This communication describes a simple method for patterning paper to create well-defined, millimeter-sized channels, comprising hydrophilic paper bounded by hydrophobic polymer. We believe that this type of patterned paper will become the basis for low-cost, portable, and technically simple multiplexed bioassays. We demonstrate this capability by the simultaneous detection of glucose and protein in 5 μL of urine. The assay system is small, disposable, easy to use (and carry), and requires no external equipment, reagents, or power sources. We believe this kind of system is attractive for uses in less-industrialized countries, in the field, or as an inexpensive alternative to more advanced technologies already used in clinical settings.[1-4] The analysis of biological fluids is necessary for monitoring the health of populations,[2] but these measurements are difficult to implement in remote regions such as those found in less-industrialized countries, in emergency situations, or in home health-care settings.[3] Conventional laboratory instruments provide quantitative measurements of biological samples, but they are unsuitable for these situations since they are large, expensive, and require trained personnel and considerable volumes of biological samples.[2] Other bioassay platforms provide alternatives to more expensive instruments,[5-7] but the need remains for a platform that uses small volumes of sample and that is sufficiently inexpensive to be used widely for measuring samples from large populations. We believe that paper may serve as a particularly convenient platform for running bioassays in the remote situations locations. As a prototype for a mthod we believe to be particularly promosing, we patterned photoresist onto chromatography paper to form defined areas of hydrophilic paper separated by hydrophobic lines or “walls”; these patterns provide spatial control of biological fluids and enable fluid transport, without pumping, due to capillary action in the millimeter-sized channels produced. This method for patterning paper makes it possible to run multiple diagnostic assays on one strip of paper, while still using only small volumes of a single sample. In a fully developed technology, patterned photoresist would be replaced by an appropriate printing technology, but patterning paper with photoresist is: i) convenient for prototyping these devices, and ii) a useful new micropatterning technology in its own right. We patterned chromatography paper with SU-8 2010 photoresist as shown in Figure 1a and as described below: we soaked a 7.5-cm diameter piece of chromatography paper in 2 mL of SU-8 2010 for 30 s, spun it at 2000 rpm for 30 s, and then baked it at 95 °C for 5 min to remove the cyclopentanone in the SU-8 formula. We then exposed the photoresist and paper to 405 nm UV light (50 mW/cm2) for 10 s through a photo-mask (CAD/Art Services, Inc.) that was aligned using a mask aligner (OL-2 Mask Aligner, AB-M, Inc). After exposure, we baked the paper a second time at 95 °C for 5 min to cross-link the exposed portions of the resist. The unpolymerized photoresist was removed by soaking the paper in propylene glycol monomethyl ether acetate (PGMEA) (5 min), and by washing the pattern with propan-2-ol (3 × 10 mL). The paper was more hydrophobic after it was patterned, presumably due to residual resist bound to the paper, so we exposed the entire surface to an oxygen plasma for 10 s at 600 millitorr (SPI Plasma-Prep II, Structure Probe, Inc) to increase the hydrophilicity of the paper (Figures 2a and 2b). Figure 1 Chromatography paper patterned with photoresist. The darker lines are cured photoresist; the lighter areas are unexposed paper. (a) Patterned paper after absorbing 5 μL of Waterman red ink by capillary action. The central channel absorbs the sample ... Figure 2 Assays contaminated with (a) dirt, (b) plant pollen, and (c) graphite powder. The pictures were taken before and after running an artificial urine solution that contained 550 mM glucose and 75 μM BSA. The particulates do not move up the channels ... The patterned paper can be derivatized for biological assays by adding appropriate reagents to the test areas (Figures 1b and ​and2b).2b). In this communication, we demonstrate the method by detecting glucose and protein,[8] but the surface should be suitable for measuring many other analytes as well.[7] The glucose assay is based on the enzymatic oxidation of iodide to iodine,[9] where a color change from clear to brown is associated with the presence of glucose.[10] The protein assay is based on the color change of tetrabromophenol blue (TBPB) when it ionizes and binds to proteins;[11] a positive result in this case is indicated by a color change from yellow to blue. For the glucose assay, we spotted 0.3 μL of a 0.6 M solution of potassium iodide, followed by 0.3 μL of a 1:5 horseradish peroxidase/glucose oxidase solution (15 units of protein per mL of solution). For the protein assay, we spotted 0.3 μL of a 250-mM citrate buffer (pH 1.8) in a well separate from the glucose assay, and then layered 0.3 μL of a 3.3 mM solution of tetrabromophenol blue (TBPB) in 95% ethanol over the citrate buffer. The spotted reagents were allowed to air dry at room temperature. This pre-loaded paper gave consistent results for the protein assay regardless of storage temperature and time (when stored for 15 d both at 0 °C and at 23 °C, wrapped in aluminum foil). The glucose assay was sensitive to storage conditions, and showed decreased signal for assays run 24 h after spotting the reagents (when stored at 23 °C); when stored at 0 °C, however, the glucose assay was as sensitive after day 15 as it was on day 1. We measured artificial samples of glucose and protein in clinically relevant ranges (2.5-50 mM for glucose and 0.38-7.5 μM for bovine serum albumin (BSA))[12, 13] by dipping the bottom of each test strip in 5 μL of a pre-made test solution (Figure 2d). The fluid filled the entire pattern within ca. one minute, but the assays required 10-11 min for the paper to dry and for the color to fully develop.[14] In all cases, we observed color changes corresponding roughly in intensity to the amount of glucose and protein in the test samples, where the lowest concentrations define the lower limits to which these assays can be used (Figure 2e). For comparison, commercially-available dipsticks detect glucose at concentrations as low as 5 mM[7, 9] and protein as low as 0.75 μM;[6, 15] these limits indicate that these paper-based assays are comparable in sensitivity to commercial dipstick assays. Our assay format also allows for the measurement of multiple analytes. This paper-based assay is suitable for measuring multiple samples in parallel and in a relatively short period of time. For example, in one trial, one researcher was able to run 20 different samples (all with 550 mM glucose and 75 μM BSA) within 7.5 min (followed by another 10.5 min for the color to fully develop). An 18-min assay of this type—one capable of measuring two analytes in 20 different sample—may be efficient enough to use in high-throughput screens of larger sample pools. In the field, samples will not be measured under sterile conditions, and dust and dirt may contaminate the assays. The combination of paper and capillary action provides a mechanism for separating particulates from a biological fluid. As a demonstration, we purposely contaminated the artificial urine samples with quantities of dirt, plant pollen, and graphite powder at levels higher than we might expect to see in the samples in the field. These particulates do not move up the channels under the action of capillary wicking, and do not interfere with the assay (Figure 3). Paper strips have been used in biomedical assays for decades because they offer an inexpensive platform for colorimetric chemical testing.[1] Patterned paper has characteristics that lead to miniaturized assays that run by capillary action (e.g., without external pumping), with small volumes of fluids. These methods suggest a path for the development of simple, inexpensive, and portable diagnostic assays that may be useful in remote settings, and in particular, in less-industrialized countries where simple assays are becoming increasingly important for detecting disease and monitoring health,[16, 17], for environmental monitoring, in veterinary and agricultural practice and for other applications.

2,580 citations

Journal ArticleDOI
16 Oct 1998-Science
TL;DR: A device was developed that uses microfabricated fluidic channels, heaters, temperature sensors, and fluorescence detectors to analyze nanoliter-size DNA samples to facilitate the use of DNA analysis in applications such as rapid medical diagnostics and point-of-use agricultural testing.
Abstract: A device was developed that uses microfabricated fluidic channels, heaters, temperature sensors, and fluorescence detectors to analyze nanoliter-size DNA samples. The device is capable of measuring aqueous reagent and DNA-containing solutions, mixing the solutions together, amplifying or digesting the DNA to form discrete products, and separating and detecting those products. No external lenses, heaters, or mechanical pumps are necessary for complete sample processing and analysis. Because all of the components are made using conventional photolithographic production techniques, they operate as a single closed system. The components have the potential for assembly into complex, low-power, integrated analysis systems at low unit cost. The availability of portable, reliable instruments may facilitate the use of DNA analysis in applications such as rapid medical diagnostics and point-of-use agricultural testing.

1,486 citations

Journal ArticleDOI
TL;DR: A prototype system for quantifying bioassays and for exchanging the results of the assays digitally with physicians located off-site offers new opportunities for inexpensive monitoring of health, especially in situations that require physicians to travel to patients to obtain diagnostic information that might be obtained more effectively by less valuable personnel.
Abstract: This article describes a prototype system for quantifying bioassays and for exchanging the results of the assays digitally with physicians located off-site. The system uses paper-based microfluidic devices for running multiple assays simultaneously, camera phones or portable scanners for digitizing the intensity of color associated with each colorimetric assay, and established communications infrastructure for transferring the digital information from the assay site to an off-site laboratory for analysis by a trained medical professional; the diagnosis then can be returned directly to the healthcare provider in the field. The microfluidic devices were fabricated in paper using photolithography and were functionalized with reagents for colorimetric assays. The results of the assays were quantified by comparing the intensities of the color developed in each assay with those of calibration curves. An example of this system quantified clinically relevant concentrations of glucose and protein in artificial uri...

1,393 citations

Journal ArticleDOI
TL;DR: The prospects and challenges of electrochemical biosensors for next-generation cancer diagnostics are discussed, with a focus on portable devices that can be delivered at patient bedside or physician office within few minutes.

1,185 citations

Journal ArticleDOI
TL;DR: In this article, fundamental aspects of deposition, structure and electrochemistry of Prussian Blue and its analogues are discussed, and the future prospects of chemical and biological sensors based on metal hexacyanoferrates are outlined.
Abstract: This article reviews fundamental aspects of deposition, structure and electrochemistry of Prussian Blue and its analogues. Special attention is given to the metal hexacyanoferrates with potential analytical applications. Prussian Blue and its analogues as advanced sensing materials for nonelectroactive ions are discussed. In contrast to common ‘smart materials’, the sensitivity and selectivity of metal hexacyanoferrates to such ions is provided by thermodynamic background. Prussian Blue itself is recognized as the most advantageous low-potential transducer for hydrogen peroxide over all known systems. Both high sensitivity (ca. 1 A M−1 cm−2) and selectivity in relation to oxygen reduction are more than three orders of magnitude higher, than for platinum electrodes. Biosensors based on different transducing principles containing enzymes oxidases are compared, and the devices operated due to hydrogen peroxide detection with the Prussian Blue based transducer are shown to be the most advantageous ones. The future prospects of chemical and biological sensors based on metal hexacyanoferrates are outlined.

827 citations