Evaluation of a rapid antigen test (Panbio™ COVID-19 Ag rapid test device) for SARS-CoV-2 detection in asymptomatic close contacts of COVID-19 patients.

01 Apr 2021-Clinical Microbiology and Infection (Elsevier)-Vol. 27, Iss: 4
TL;DR: In this article, the Panbio™ COVID-19 Ag Rapid Test Device (Abbott Diagnostics, Jena, Germany) was evaluated for detecting SARS-CoV-2-infected asymptomatic individuals.
About: This article is published in Clinical Microbiology and Infection.The article was published on 2021-04-01 and is currently open access. It has received 116 citations till now. The article focuses on the topics: Rapid antigen test.
Citations
More filters
Journal ArticleDOI
TL;DR: A systematic review and meta-analysis of commercially available rapid diagnostic tests (Ag-RDTs) for SARS-CoV-2 up until 30 April 2021 was conducted in this paper.
Abstract: Background SARS-CoV-2 antigen rapid diagnostic tests (Ag-RDTs) are increasingly being integrated in testing strategies around the world. Studies of the Ag-RDTs have shown variable performance. In this systematic review and meta-analysis, we assessed the clinical accuracy (sensitivity and specificity) of commercially available Ag-RDTs. Methods and findings We registered the review on PROSPERO (registration number: CRD42020225140). We systematically searched multiple databases (PubMed, Web of Science Core Collection, medRvix, bioRvix, and FIND) for publications evaluating the accuracy of Ag-RDTs for SARS-CoV-2 up until 30 April 2021. Descriptive analyses of all studies were performed, and when more than 4 studies were available, a random-effects meta-analysis was used to estimate pooled sensitivity and specificity in comparison to reverse transcription polymerase chain reaction (RT-PCR) testing. We assessed heterogeneity by subgroup analyses, and rated study quality and risk of bias using the QUADAS-2 assessment tool. From a total of 14,254 articles, we included 133 analytical and clinical studies resulting in 214 clinical accuracy datasets with 112,323 samples. Across all meta-analyzed samples, the pooled Ag-RDT sensitivity and specificity were 71.2% (95% CI 68.2% to 74.0%) and 98.9% (95% CI 98.6% to 99.1%), respectively. Sensitivity increased to 76.3% (95% CI 73.1% to 79.2%) if analysis was restricted to studies that followed the Ag-RDT manufacturers’ instructions. LumiraDx showed the highest sensitivity, with 88.2% (95% CI 59.0% to 97.5%). Of instrument-free Ag-RDTs, Standard Q nasal performed best, with 80.2% sensitivity (95% CI 70.3% to 87.4%). Across all Ag-RDTs, sensitivity was markedly better on samples with lower RT-PCR cycle threshold (Ct) values, i.e., <20 (96.5%, 95% CI 92.6% to 98.4%) and <25 (95.8%, 95% CI 92.3% to 97.8%), in comparison to those with Ct ≥ 25 (50.7%, 95% CI 35.6% to 65.8%) and ≥30 (20.9%, 95% CI 12.5% to 32.8%). Testing in the first week from symptom onset resulted in substantially higher sensitivity (83.8%, 95% CI 76.3% to 89.2%) compared to testing after 1 week (61.5%, 95% CI 52.2% to 70.0%). The best Ag-RDT sensitivity was found with anterior nasal sampling (75.5%, 95% CI 70.4% to 79.9%), in comparison to other sample types (e.g., nasopharyngeal, 71.6%, 95% CI 68.1% to 74.9%), although CIs were overlapping. Concerns of bias were raised across all datasets, and financial support from the manufacturer was reported in 24.1% of datasets. Our analysis was limited by the included studies’ heterogeneity in design and reporting. Conclusions In this study we found that Ag-RDTs detect the vast majority of SARS-CoV-2-infected persons within the first week of symptom onset and those with high viral load. Thus, they can have high utility for diagnostic purposes in the early phase of disease, making them a valuable tool to fight the spread of SARS-CoV-2. Standardization in conduct and reporting of clinical accuracy studies would improve comparability and use of data.

181 citations

Posted ContentDOI
01 Mar 2021-medRxiv
TL;DR: In this article, a systematic review and meta-analysis of commercially available rapid diagnostic tests (Ag-RDTs) is presented, where the clinical accuracy (sensitivity and specificity) of these tests are assessed.
Abstract: Background SARS-CoV-2 antigen rapid diagnostic tests (Ag-RDTs) are increasingly being integrated in testing strategies around the world. Studies of the Ag-RDTs have shown variable performance. In this systematic review and meta-analysis, we assessed the clinical accuracy (sensitivity and specificity) of commercially available Ag-RDTs. Methods We registered the review on PROSPERO (Registration number: CRD42020225140). We systematically searched multiple databases (PubMed, Web of Science Core Collection, medRvix and bioRvix, FINDdx) for publications up until December 11th, 2020. Descriptive analyses of all studies were performed and when more than four studies were available, a random-effects meta-analysis was used to estimate pooled sensitivity and specificity in comparison to reverse transcriptase polymerase chain reaction testing. We assessed heterogeneity by subgroup analyses ((1) performed con-form with manufacturer’s instructions for use (IFU) or not, (2) symptomatic vs. asymptomatic, (3) duration of symptoms less than seven days vs. more than seven days, (4) Ct-value Results From a total of 11,715 articles, we extracted 98 analytical and clinical data sets. 74 clinical accuracy data sets were evaluated that included 31,202 samples. Across all meta-analyzed samples, the pooled Ag-RDT sensitivity was 73.8% (CI 68.6 to 78.5). If analysis was restricted to studies that followed the Ag-RDT manufacturers’ instructions using fresh upper respiratory swab samples, the sensitivity increased to 79.1% (95%CI 75.0 to 82.8). The SD Biosensor Standard Q and Abbott Panbio showed the highest sensitivity with 81.7% and 72.7%, respectively. The best Ag-RDT performance was found with nasopharyngeal sampling (77.3%, CI 72.0 to 81.9) in comparison to other sample types (e.g., anterior nasal or mid turbinate 63.5%, CI 49.5 to 75.5). Testing in the first week from symptom onset resulted in higher sensitivity (87.5%, CI 86.0 to 89.1) compared to testing after one week (64.1%, CI 54.4 to 73.8). The tests performed markedly better on samples with lower Ct-values, i.e., Conclusion As Ag-RDTs detect most cases within the first week of symptom onset and those with high viral load, they can have high utility for screening purposes in the early phase of disease, and thus can be a valuable tool to fight the spread of SARS-CoV-2. Standardization of conduct and reporting of clinical accuracy studies would improve comparability and use of data. Summary In this living systematic review we analyzed 98 data sets for performance of SARS-CoV-2 Ag-RDTs compared to RT-PCR. Best-performing tests achieved a sensitivity of 81.7%. Highest sensitivity was found in patients within seven days of symptom onset when NP swabs were utilized.

166 citations

Journal ArticleDOI
TL;DR: LFDs could support the scaling up of mass testing to aid track and trace methodology and break the chain of transmission of COVID-19 with the additional benefit of providing individuals with the results in a much shorter time frame.
Abstract: Lateral flow devices (LFDs) are viral antigen tests for the detection of SARS-CoV-2 that produce a rapid result, are inexpensive and easy to operate. They have been advocated for use by the World Health Organisation to help control outbreaks and break the chain of transmission of COVID-19 infections. There are now several studies assessing their accuracy but as yet no systematic review. Our aims were to assess the sensitivity and specificity of LFDs in a systematic review and summarise the sensitivity and specificity of these tests. A targeted search of Pubmed and Medxriv, using PRISMA principles, was conducted identifying clinical studies assessing the sensitivity and specificity of LFDs as their primary outcome compared to reverse transcriptase polymerase chain reaction (RT-PCR) for the detection of SARS-CoV-2. Based on extracted data sensitivity and specificity was calculated for each study. Data was pooled based on manufacturer of LFD and split based on operator (self-swab or by trained professional) and sensitivity and specificity data were calculated. Twenty-four papers were identified involving over 26,000 test results. Sensitivity from individual studies ranged from 37.7% (95% CI 30.6–45.5) to 99.2% (95% CI 95.5–99.9) and specificity from 92.4% (95% CI 87.5–95.5) to 100.0% (95% CI 99.7–100.0). Operation of the test by a trained professional or by the test subject with self-swabbing produced comparable results. This systematic review identified that the performance of lateral flow devices is heterogeneous and dependent on the manufacturer. Some perform with high specificity but a great range of sensitivities were shown (38.32–99.19%). Test performance does not appear dependent on the operator. Potentially, LFDs could support the scaling up of mass testing to aid track and trace methodology and break the chain of transmission of COVID-19 with the additional benefit of providing individuals with the results in a much shorter time frame.

55 citations

Journal ArticleDOI
TL;DR: In this article , the second update of this review was published, which was first published in 2020, and included 152 evaluations of single test applications including 100,462 unique samples (16,822 with confirmed SARS-CoV-2 infection or known absence of infection).
Abstract: Accurate rapid diagnostic tests for SARS-CoV-2 infection would be a useful tool to help manage the COVID-19 pandemic. Testing strategies that use rapid antigen tests to detect current infection have the potential to increase access to testing, speed detection of infection, and inform clinical and public health management decisions to reduce transmission. This is the second update of this review, which was first published in 2020.To assess the diagnostic accuracy of rapid, point-of-care antigen tests for diagnosis of SARS-CoV-2 infection. We consider accuracy separately in symptomatic and asymptomatic population groups. Sources of heterogeneity investigated included setting and indication for testing, assay format, sample site, viral load, age, timing of test, and study design.We searched the COVID-19 Open Access Project living evidence database from the University of Bern (which includes daily updates from PubMed and Embase and preprints from medRxiv and bioRxiv) on 08 March 2021. We included independent evaluations from national reference laboratories, FIND and the Diagnostics Global Health website. We did not apply language restrictions.We included studies of people with either suspected SARS-CoV-2 infection, known SARS-CoV-2 infection or known absence of infection, or those who were being screened for infection. We included test accuracy studies of any design that evaluated commercially produced, rapid antigen tests. We included evaluations of single applications of a test (one test result reported per person) and evaluations of serial testing (repeated antigen testing over time). Reference standards for presence or absence of infection were any laboratory-based molecular test (primarily reverse transcription polymerase chain reaction (RT-PCR)) or pre-pandemic respiratory sample.We used standard screening procedures with three people. Two people independently carried out quality assessment (using the QUADAS-2 tool) and extracted study results. Other study characteristics were extracted by one review author and checked by a second. We present sensitivity and specificity with 95% confidence intervals (CIs) for each test, and pooled data using the bivariate model. We investigated heterogeneity by including indicator variables in the random-effects logistic regression models. We tabulated results by test manufacturer and compliance with manufacturer instructions for use and according to symptom status.We included 155 study cohorts (described in 166 study reports, with 24 as preprints). The main results relate to 152 evaluations of single test applications including 100,462 unique samples (16,822 with confirmed SARS-CoV-2). Studies were mainly conducted in Europe (101/152, 66%), and evaluated 49 different commercial antigen assays. Only 23 studies compared two or more brands of test. Risk of bias was high because of participant selection (40, 26%); interpretation of the index test (6, 4%); weaknesses in the reference standard for absence of infection (119, 78%); and participant flow and timing 41 (27%). Characteristics of participants (45, 30%) and index test delivery (47, 31%) differed from the way in which and in whom the test was intended to be used. Nearly all studies (91%) used a single RT-PCR result to define presence or absence of infection. The 152 studies of single test applications reported 228 evaluations of antigen tests. Estimates of sensitivity varied considerably between studies, with consistently high specificities. Average sensitivity was higher in symptomatic (73.0%, 95% CI 69.3% to 76.4%; 109 evaluations; 50,574 samples, 11,662 cases) compared to asymptomatic participants (54.7%, 95% CI 47.7% to 61.6%; 50 evaluations; 40,956 samples, 2641 cases). Average sensitivity was higher in the first week after symptom onset (80.9%, 95% CI 76.9% to 84.4%; 30 evaluations, 2408 cases) than in the second week of symptoms (53.8%, 95% CI 48.0% to 59.6%; 40 evaluations, 1119 cases). For those who were asymptomatic at the time of testing, sensitivity was higher when an epidemiological exposure to SARS-CoV-2 was suspected (64.3%, 95% CI 54.6% to 73.0%; 16 evaluations; 7677 samples, 703 cases) compared to where COVID-19 testing was reported to be widely available to anyone on presentation for testing (49.6%, 95% CI 42.1% to 57.1%; 26 evaluations; 31,904 samples, 1758 cases). Average specificity was similarly high for symptomatic (99.1%) or asymptomatic (99.7%) participants. We observed a steady decline in summary sensitivities as measures of sample viral load decreased. Sensitivity varied between brands. When tests were used according to manufacturer instructions, average sensitivities by brand ranged from 34.3% to 91.3% in symptomatic participants (20 assays with eligible data) and from 28.6% to 77.8% for asymptomatic participants (12 assays). For symptomatic participants, summary sensitivities for seven assays were 80% or more (meeting acceptable criteria set by the World Health Organization (WHO)). The WHO acceptable performance criterion of 97% specificity was met by 17 of 20 assays when tests were used according to manufacturer instructions, 12 of which demonstrated specificities above 99%. For asymptomatic participants the sensitivities of only two assays approached but did not meet WHO acceptable performance standards in one study each; specificities for asymptomatic participants were in a similar range to those observed for symptomatic people. At 5% prevalence using summary data in symptomatic people during the first week after symptom onset, the positive predictive value (PPV) of 89% means that 1 in 10 positive results will be a false positive, and around 1 in 5 cases will be missed. At 0.5% prevalence using summary data for asymptomatic people, where testing was widely available and where epidemiological exposure to COVID-19 was suspected, resulting PPVs would be 38% to 52%, meaning that between 2 in 5 and 1 in 2 positive results will be false positives, and between 1 in 2 and 1 in 3 cases will be missed.Antigen tests vary in sensitivity. In people with signs and symptoms of COVID-19, sensitivities are highest in the first week of illness when viral loads are higher. Assays that meet appropriate performance standards, such as those set by WHO, could replace laboratory-based RT-PCR when immediate decisions about patient care must be made, or where RT-PCR cannot be delivered in a timely manner. However, they are more suitable for use as triage to RT-PCR testing. The variable sensitivity of antigen tests means that people who test negative may still be infected. Many commercially available rapid antigen tests have not been evaluated in independent validation studies. Evidence for testing in asymptomatic cohorts has increased, however sensitivity is lower and there is a paucity of evidence for testing in different settings. Questions remain about the use of antigen test-based repeat testing strategies. Further research is needed to evaluate the effectiveness of screening programmes at reducing transmission of infection, whether mass screening or targeted approaches including schools, healthcare setting and traveller screening.

54 citations

Journal ArticleDOI
TL;DR: In this paper, the overall pooled specificity and sensitivity of RAT were detected as 99.4% (95% CI: 99.1-99.8; I2 = 90%) and 68.4%, respectively.
Abstract: Early diagnosis is still as crucial as the initial stage of the COVID-19 pandemic. As RT-PCR sometimes is not feasible in developing nations or rural areas, health professionals may use a rapid antigen test (RAT) to lessen the load of diagnosis. However, the efficacy of RAT is yet to be investigated thoroughly. Hence, we tried to evaluate the overall performance of RAT in SARS-CoV-2 diagnosis. Based on our PROSPERO registered protocol (CRD42021231432), we searched online databases (i.e., PubMed, Google Scholar, Scopus, and Web of Science) and analysed overall pooled specificity and sensitivity of RAT along with study quality, publication bias, heterogeneity and more. The overall pooled specificity and sensitivity of RAT were detected as 99.4% (95% CI: 99.1–99.8; I2 = 90%) and 68.4% (95% CI: 60.8–75.9; I2 = 98%), respectively. In subgroup analyses, nasopharyngeal specimens and symptomatic patient’s samples were more sensitive in RAT, while cycle threshold (Ct) values were found to have an inverse relationship with sensitivity. In the European and American populations, RAT showed better performance. Although the sensitivity of RAT is yet to be improved, it could still be an alternative in places with poor laboratory set up. Nevertheless, the negative samples of RAT can be re-tested using RT-PCR to reduce false negative results.

52 citations

References
More filters
Journal ArticleDOI
01 Apr 2020-Nature
TL;DR: Detailed virological analysis of nine cases of coronavirus disease 2019 (COVID-19) provides proof of active replication of the SARS-CoV-2 virus in tissues of the upper respiratory tract.
Abstract: Coronavirus disease 2019 (COVID-19) is an acute infection of the respiratory tract that emerged in late 20191,2. Initial outbreaks in China involved 13.8% of cases with severe courses, and 6.1% of cases with critical courses3. This severe presentation may result from the virus using a virus receptor that is expressed predominantly in the lung2,4; the same receptor tropism is thought to have determined the pathogenicity—but also aided in the control—of severe acute respiratory syndrome (SARS) in 20035. However, there are reports of cases of COVID-19 in which the patient shows mild upper respiratory tract symptoms, which suggests the potential for pre- or oligosymptomatic transmission6–8. There is an urgent need for information on virus replication, immunity and infectivity in specific sites of the body. Here we report a detailed virological analysis of nine cases of COVID-19 that provides proof of active virus replication in tissues of the upper respiratory tract. Pharyngeal virus shedding was very high during the first week of symptoms, with a peak at 7.11 × 108 RNA copies per throat swab on day 4. Infectious virus was readily isolated from samples derived from the throat or lung, but not from stool samples—in spite of high concentrations of virus RNA. Blood and urine samples never yielded virus. Active replication in the throat was confirmed by the presence of viral replicative RNA intermediates in the throat samples. We consistently detected sequence-distinct virus populations in throat and lung samples from one patient, proving independent replication. The shedding of viral RNA from sputum outlasted the end of symptoms. Seroconversion occurred after 7 days in 50% of patients (and by day 14 in all patients), but was not followed by a rapid decline in viral load. COVID-19 can present as a mild illness of the upper respiratory tract. The confirmation of active virus replication in the upper respiratory tract has implications for the containment of COVID-19. Detailed virological analysis of nine cases of coronavirus disease 2019 (COVID-19) provides proof of active replication of the SARS-CoV-2 virus in tissues of the upper respiratory tract.

5,840 citations

Journal ArticleDOI
TL;DR: The data suggested that patients with RT-PCR-proven COVID-19 testing negative by RAD are unlikely to be infectious, and the Panbio™ CO VID-19 Ag Rapid Test Device performed well as a POCT for early diagnosis in primary healthcare centers.

266 citations

Journal ArticleDOI
TL;DR: The rapid antigen test evaluated in this study showed a high sensitivity and specificity in samples obtained during the first week of symptoms and with high viral loads, and seems to be an effective strategy for controlling the COVID-19 pandemic.

216 citations

Journal ArticleDOI
TL;DR: Asymptomatic SARS-CoV-2 infection is common and can be detected by analysis of saliva or NTS, but they appear able to transmit the virus to others.
Abstract: BACKGROUND: Little is known about the natural history of asymptomatic SARS-CoV-2 infection or its contribution to infection transmission. METHODS: We conducted a prospective study at a quarantine center for COVID-19 in Ho Chi Minh City, Vietnam. We enrolled quarantined people with RT-PCR-confirmed SARS-CoV-2 infection, collecting clinical data, travel and contact history, and saliva at enrolment and daily nasopharyngeal throat swabs (NTS) for RT-PCR testing. We compared the natural history and transmission potential of asymptomatic and symptomatic individuals. RESULTS: Between March 10th and April 4th, 2020, 14,000 quarantined people were tested for SARS-CoV-2; 49 were positive. Of these, 30 participated in the study: 13(43%) never had symptoms and 17(57%) were symptomatic. 17(57%) participants acquired their infection outside Vietnam. Compared with symptomatic individuals, asymptomatic people were less likely to have detectable SARS-CoV-2 in NTS samples collected at enrolment (8/13 (62%) vs. 17/17 (100%) P=0.02). SARS-CoV-2 RNA was detected in 20/27 (74%) available saliva; 7/11 (64%) in the asymptomatic and 13/16 (81%) in the symptomatic group (P=0.56). Analysis of the probability of RT-PCR positivity showed asymptomatic participants had faster viral clearance than symptomatic participants (P<0.001 for difference over first 19 days). This difference was most pronounced during the first week of follow-up. Two of the asymptomatic individuals appeared to transmit the infection to up to four contacts. CONCLUSIONS: Asymptomatic SARS-CoV-2 infection is common and can be detected by analysis of saliva or NTS. NTS viral loads fall faster in asymptomatic individuals, but they appear able to transmit the virus to others.

204 citations

Journal ArticleDOI
TL;DR: Coronavirus disease 2019 (COVID-19), declared a pandemic by the WHO on March 11th, 2020 (1), requires an early diagnosis to optimize patient management and limit further transmission.
Abstract: Coronavirus disease 2019 (COVID-19), declared a pandemic by the WHO on March 11th, 2020 (1), requires an early diagnosis to optimize patient management and limit further transmission.….

152 citations

Related Papers (5)