EWOD (electrowetting on dielectric) digital microfluidics powered by finger actuation

17 Feb 2014-Lab on a Chip (The Royal Society of Chemistry)-Vol. 14, Iss: 6, pp 1117-1122
TL;DR: In this paper, a finger-actuated digital microfluidics (F-DMF) based on the manipulation of discrete droplets via the electrowetting on dielectric (EWOD) phenomenon is reported.
Abstract: We report finger-actuated digital microfluidics (F-DMF) based on the manipulation of discrete droplets via the electrowetting on dielectric (EWOD) phenomenon. Instead of requiring an external power supply, our F-DMF uses piezoelectric elements to convert mechanical energy produced by human fingers to electric voltage pulses for droplet actuation. Voltage outputs of over 40 V are provided by single piezoelectric elements, which is necessary for oil-free EWOD devices with thin (typically <1 μm) dielectric layers. Higher actuation voltages can be provided using multiple piezoelectric elements connected in series when needed. Using this energy conversion scheme, we confirmed basic modes of EWOD droplet operation, such as droplet transport, splitting and merging. Using two piezoelectric elements in series, we also successfully demonstrated applications of F-DMF for glucose detection and immunoassay. Not requiring power sources, F-DMF offers intriguing paths for various portable and other microfluidic applications.

Summary (2 min read)

Introduction

  • Microfluidic devices are attractive for portable applications, including point-of-care (POC) diagnosis, bio-surveillance, environment sampling, and forensic science.
  • Previous studies proposed different approaches to circumventing these challenges.
  • Battery-powered microfluidic devices with integrated microvalves and micropumps, for example, have been reported for carrying out biological assays.

Device design and fabrication

  • To successfully manipulate micro-droplets using EWOD, one must generate voltage pulses of sufficient amplitude to overcome capillary, inertial, and viscous forces.
  • Because the capillary resistance originated from the contact angle hysteresis of an aqueous droplet surrounded in air (as opposed to the popular oil environment) is larger than the inertial or viscous resistance in most cases, performance of EWOD devices is often measured in the air environment without resorting to the filler oil or oil impregnation.
  • The bottom glass plate contains an array of 1 × 1 mm2 gold electrodes, which was fabricated using standard micro-fabrication processes.
  • A shadow mask was used to define electrical contact pads.
  • The 1118 | Lab Chip, 2014, 14, 1117–1122 dielectric layer thickness is further limited by the capacitance allowed per EWOD electrode, which must be kept below that of the piezoelectric element to minimize the voltage dividing effect.

EWOD actuation voltage

  • To characterize the threshold actuation voltage required for EWOD actuation as a function of the dielectric layer thickness, an external programmable power source was used to apply precisely defined voltage pulses.
  • A water droplet of ~0.3 μL in volume was spotted onto the EWOD device and subsequently split into two nominally identical daughter droplets.
  • After one of these droplets was positioned on one of the electrodes, the amplitude of the voltage pulse applied to the adjacent electrode was gradually increased until the droplet was successfully transported.
  • The threshold actuation voltage was recorded for each dielectric layer thickness and the results are plotted in Fig. 2 .
  • Also shown as a solid line in Fig. 2 are the voltages required theoretically (eqn (1)) for the contact angle to change from 120° to 70°, which is an empirically determined range for droplet actuation in the given EWOD device.

Mechanical energy conversion

  • The prediction (straight line) agrees reasonably well with the experimental results over the entire range.
  • The results are shown for three different bending angles.
  • At relatively small tip bending angles (~15°), the total voltage output increases linearly with each additional piezoelectric element.

Basic droplet operations

  • The authors next demonstrated successful finger-powered actuation of water droplets on EWOD devices with 0.8 μm-thick PECVD SiNx dielectric layers.
  • Referring to Fig. 6, element 2 was not entirely released when element 3 was deflected, such that the front contact line of the droplet would stay across the gap between electrodes 2 and 3.
  • The authors also demonstrated droplet splitting by simultaneously deflecting two non-adjacent piezoelectric elements while keeping the middle one non-deflected.
  • Elongated in the longitudinal direction by a wetting force exerted at the two ends while keeping the middle non-wetting, as shown in Fig.
  • The actuation voltage on either side is approximately 40–50 V, produced by one single piezoelectric element with a bending angle <90°.

Applications to biological assays

  • In these experiments, where higher EWOD voltages are necessary to actuate droplets, the authors used silicon nitride layers of thickness 2.5 μm because they help prevent electrolysis of water under their low-frequency (~1 Hz) finger-driven actuation scheme.
  • Since the reagent solution may chemically attack the hydrophobic coatings, the authors pre-treated both the upper plate and bottom substrate of their EWOD devices with silicone oil.
  • The reagent droplet was then transported towards the glucose sample droplet and merged, as shown in Fig. 9(b–d).
  • In their experiment, the enzyme substrate was used to detect an ALP conjugated antibody.

Conclusions

  • In summary, the authors demonstrated finger-actuated digital microfluidics based on the electrowetting-on-dielectric (EWOD) The Royal Society of Chemistry 2014 phenomenon by using piezoelectric energy conversion of human power.
  • The authors demonstrated generation of voltage pulses of amplitudes >100 V using laminated polymer piezoelectric elements connected in series.
  • Using the scheme, the authors confirmed basic EWOD droplet operations such as droplet transport, splitting and merging and further demonstrated an implementation of basic assay steps in glucose detection and immunoassay.
  • Due to the low-frequency nature of finger actuation, a relatively thick dielectric layer was used to help prevent possible electrolysis.
  • The authors work offers a promising solution for expanded applications of EWOD-based digital microfluidics in portable systems.

Did you find this useful? Give us your feedback

Content maybe subject to copyright    Report

UCLA
UCLA Previously Published Works
Title
EWOD (electrowetting on dielectric) digital microfluidics powered by finger actuation.
Permalink
https://escholarship.org/uc/item/2nr7z58v
Journal
Lab on a chip, 14(6)
ISSN
1473-0197
Authors
Peng, Cheng
Zhang, Zhongning
Kim, Chang-Jin CJ
et al.
Publication Date
2014-03-01
DOI
10.1039/c3lc51223a
Peer reviewed
eScholarship.org Powered by the California Digital Library
University of California

Lab on a Chip
PAPER
Cite this: Lab Chip,2014,14,1117
Received 30th October 2013,
Accepted 17th December 2013
DOI: 10.1039/c3lc51223a
www.rsc.org/loc
EWOD (electrowetting on dielectric) digital
microfluidics powered by finger actuation
Cheng Peng,
a
Zhongning Zhang,
a
Chang-Jin CJ Kim
a
and Y. Sungtaek Ju
*
ab
We report finger-actuated digital microfluidics (F-DMF) based on the manipulation of discrete droplets via
the electrowetting on dielectric (EWOD) phenomenon. Instead of requiring an external power supply, our
F-DMF uses piezoelectric elements to convert mechanical energy produced by human fingers to electric
voltage pulses for droplet actuation. Voltage outputs of over 40 V are provided by single piezoelectric
elements, which is necessary for oil-free EWOD devices with thin (typically <1 μm) dielectric layers. Higher
actuation voltages can be provided using multiple piezoelectric elements connected in series when needed.
Using this energy conversion scheme, we confirmed basic modes of EWOD droplet operation, such as
droplet transport, splitting and merging. Using two piezoelectric elements in series, we also successfully
demonstrated applications of F-DMF for glucose detection and immunoassay. Not requiring power sources,
F-DMF offers intriguing paths for various portable and other microfluidic applications.
Introduction
Microfluidic devices are attractive for portable applications,
including point-of-care (POC) diagnosis, bio-surveillance, envi-
ronment sampling, and forensic science. They offer advantages
of low sample volume requirements and high reaction rates for
rapid analyses. However, most microfluidic systems require
bulky pumping and control systems as well as electrical outlets
to power them, limiting their field applications.
1
Previous studies proposed different approaches to
circumventing these challenges. Battery-powered microfluidic
devices with integrated microvalves and micropumps, for
example, have been reported for carrying out biological assays.
2
The resulting systems, however, were still too bulky for porta-
ble operations. Passive pumping mechanisms were also
explored. Paper-based microfluidic devices based on capillary
wicking attracted a lot of attention due to its perceived low
cost, lightweight and disposability.
3,4
Other passive pumping
mechanisms investigated include osmotic pressure,
5
built-in
vacuum
6
and human powered hydraulic pumping.
7,8
The fixed
layouts of continuous microfluidics employed in these devices,
however, limit their scalability and flexibility, making it diffi-
cult to adapt them on the fly to different protocols.
In digital microfluidics (DMF), droplets containing samples
or reagents are manipulated to carry out a range of discrete
fluidic operations. Precise control of individual droplets can be
achieved, thus providing greater flexibility and facilitating the
implementation of multi-step reactions.
9,10
The electrowetting on dielectric (EWOD) phenomenon
11
is
one of the most common actuation mechanisms used in
DMF and has been widely adopted in biomedical
1216
and,
sometimes, other applications.
17,18
Traditional EWOD
devices, however, still require an external high-voltage supply
and switching circuitry, which require significant custom
development to realize compact systems.
19
We report EWOD-based DMF with finger-actuated voltage
generation on chip, or F-DMF, to eliminate the need for
external voltage sources. Fig. 1 schematically illustrates one
possible implementation of our device concept, which uses
an array of piezoelectric elements to convert mechanical
energy pulses provided by human fingers to voltage pulses.
Using this scheme, we confirmed basic modes of droplet
Lab Chip,2014,14,11171122 | 1117This journal is © The Royal Society of Chemistry 2014
a
Department of Mechanical and Aerospace Engineering, University of California,
Los Angeles, CA, USA. E-mail: just@seas.ucla.edu
b
Department of Bioengineering, University of California, Los Angeles, CA, USA
Fig. 1 Schematic of one implementation for the finger-actuated digital
microfluidic platform. The piezoelectric elements convert mechanical
energy imparted by human fingers to electrical energy to actuate droplets
confined between two parallel plates through the electrowetting on
dielectric (EWOD) phenomenon.
Published on 18 December 2013. Downloaded on 20/02/2014 17:10:00.
View Article Online
View Journal
| View Issue

1118 | Lab Chip,2014,14,11171122 This journal is © The Royal Society of Chemistry 2014
manipulation, such as transport, splitting and merging
of water droplets.
20
Furthermore, to demonstrate its capabil-
ity for biomedical applications, we successfully performed
the key assay steps involved in glucose detection and an
immunoassay.
EWOD device
Device design and fabrication
To successfully manipulate micro-droplets using EWOD, one
must generate voltage pulses of sufficient amplitude to over-
come capillary, inertial, and viscous forces. The EWOD-
induced contact angle change is related to the applied voltage
by the LippmannYoung equation:
cos ( ) cos


V
t
V
0
0
LG
2
2
(1)
where θ
0
denotes the equilibrium contact angle at V =0V,ε
0
is the permittivity of vacuum, ε is the dielectric constant of
the dielectric layer separating the droplet from the electrode,
t is its thickness, and γ
LG
is the surface tension between the
droplet and surrounding. Since the contact angle change rep-
resents the actuation force along the surface, the higher the
applied voltage the stronger the actuation force that will drive
the droplet against the above-mentioned resistant forces.
Because the capillary resistance originated from the contact
angle hysteresis of an aqueous droplet surrounded in air
(as opposed to the popular oil environment) is larger than
the inertial or viscous resistance in most cases, performance
of EWOD devices is often measured in the air environment
without resorting to the filler oil or oil impregnation.
11
To
manipulate water droplets in air, typical EWOD devices
require a voltage source of about 40 V.
20
Our EWOD devices consist of two parallel glass plates that
were separated by approximately 100 μm. The bottom glass
plate contains an array of 1 × 1mm
2
gold electrodes, which
was fabricated using standard micro-fabrication processes.
We first deposited a 20 nm Cr/100 nm Au layer on a glass
wafer and patterned the layer by wet etching. Then, we depos-
ited a dielectric layer of silicon nitride by PECVD (plasma
enhanced chemical vapor deposition). We next spin coated a
solution of Teflon® AF (2% wt/wt in Fluorinert FC-40) at
2000 rpm for 30 seconds and baked it at 180 °C for
10 minutes to obtain a ~100 nm-thick hydrophobic topcoat. A
shadow mask was used to define electrical contact pads. The
top glass plate was coated with a transparent conductive ITO
(indium tin oxide, <15 ohm per square) layer to form a
counter electrode for EWOD. The counter electrode was also
coated with a ~100-nm-thick layer of Teflon®.
Due to the finite energy conversion efficiency of our piezo-
electric elements and safety consideration in portable appli-
cations, a low threshold actuation voltage and therefore a
thin dielectric layer is desired. However, a thin dielectric layer
is less robust and prone to pinholes and other defects, which
lead to electrolysis-induced failure of EWOD devices. The
dielectric layer thickness is further limited by the capacitance
allowed per EWOD electrode, which must be kept below that
of the piezoelectric element to minimize the voltage dividing
effect. With the above considerations in mind, we examined
SiNx layers of thickness ranging from 0.8 μm to 2.5 μmas
our dielectric layers. The estimated capacitance is approxi-
mately 60 pF per electrode for the EWOD devices with a
0.8 μm-thick dielectric layer, which is much smaller than that
of the piezoelectric element (~1 nF).
EWOD actuation voltage
To characterize the threshold actuation voltage required for
EWOD actuation as a function of the dielectric layer thick-
ness, an external programmable power source was used
to apply precisely defined voltage pulses. A water droplet of
~0.3 μL in volume was spotted onto the EWOD device and
subsequently split into two nominally identical daughter
droplets. After one of these droplets was positioned on one of
the electrodes, the amplitude of the voltage pulse applied to
the adjacent electrode was gradually increased until the drop-
let was successfully transported.
The threshold actuation voltage was recorded for each
dielectric layer thickness and the results are plotted in Fig. 2
(symbols). Also shown as a solid line in Fig. 2 are the voltages
required theoretically (eqn (1)) for the contact angle to
change from 120° to 70°, which is an empirically determined
range for droplet actuation in the given EWOD device. In
general, the threshold actuation voltage increases with the
increasing dielectric layer thickness. For EWOD devices with
a PECVD SiNx dielectric layer of thickness approximately
2.5 μm, a voltage as large as 70 V is required for successful
actuation of a water droplet.
Mechanical energy conversion
Piezoelectric elements of 13 × 25 mm
2
were used to convert a
mechanical energy input by human fingers into voltage
pulses to actuate micro-droplets in our EWOD DMF devices.
Fig. 2 Threshold actuation voltage of a water droplet on EWOD
devices as a function of the thickness of PECVD SiNx dielectric layers.
Lab on a ChipPaper
Published on 18 December 2013. Downloaded on 20/02/2014 17:10:00.
View Article Online

Lab Chip,2014,14,11171122 | 1119This journal is © The Royal Society of Chemistry 2014
Each piezoelectric element consists of a PVDF (polyvinylidene
fluoride) layer of thickness 28 μm (piezoelectric layer) lami-
nated on a polyester layer of thickness 125 μm (substrate layer)
as illustrated in Fig. 3(a) to maximize the average strain in the
piezoelectric layer and hence its output voltage. The length of
the piezoelectric layer, L
p
, is approximately 20.5 mm.
We modelled each piezoelectric element as an Euler
Bernoulli beam. The element (beam) is mounted vertically
with the clamped end fixed. The bending angle, α(s), is
defined as the rotation of the piezoelectric element beam
measured in radians at distance s from the fixed end, as illus-
trated in Fig. 3(b). For a single piezoelectric element, the
open circuit voltage can be expressed as:
21,22
V
eh
S
s
0
31
1
p
(2)
Here, e
31
is the electromechanical coupling coefficient of the
piezoelectric layer, ε
s
is the permittivity of the piezoelectric
layer under a constant strain and h
p
is the thickness of the
piezoelectric layer.
The average strain, S
¯
1
, is defined as:
S
s
yys
Lh
yy
Lh
Lh
h
h
h
1
0
1
2
1
2


p
dd
d
pp
tip
pp
()
()

(3)
where h
1
and h
2
are the distances from the neutral axis of
the whole beam to the bottom and top of the piezoelectric
layer, respectively; α
tip
and α
0
are the bending angles, α(s), at
the tip and at the starting end of the piezoelectric layer,
respectively (Fig. 3(b)).
To characterize the energy conversion capability of our
piezoelectric elements, the open-circuit voltage outputs were
measured using an electrometer of input impedance >200 TΩ.
Optical images of the side views of the element were captured
to extract the bending angles along the beam.
Fig. 4 shows the measured and predicted output voltages
as a function of the tip bending angle, α
tip
, for a single piezo-
electric element. The prediction (straight line) agrees reason-
ably well with the experimental results (symbols) over the
entire range. Output voltages greater than 40 V, which are
sufficient to reliably actuate a water droplet in the EWOD
device with a ~0.8 μm-thick SiNx dielectric layer, can be gener-
ated at tip bending angles greater than 80°. However, larger
deformations may not be desirable for field operations and
may also lead to degradation of the piezoelectric elements. We
estimated that the force required to achieve a bending angle of
36, 70 and 108° is approximately 0.06, 0.12 and 0.18 N,
respectively.
If needed, multiple piezoelectric elements may be
connected electrically in series and mechanically in parallel to
increase the output voltage while limiting required deflection
to an acceptable range. Fig. 5 shows the total output voltage of
single and two or three piezoelectric elements connected in
series. The results are shown for three different bending
angles. At relatively small tip bending angles (~15°), the total
voltage output increases linearly with each additional piezo-
electric element. At larger tip bending angles (45° and 90°),
adding more elements does not lead to a proportionate
increase in the output voltage. This is in part due to asynchro-
nous bending of each element and in part to finite leakage cur-
rents. Nevertheless, we demonstrated that output voltages on
Fig. 3 (a) Cross section of the main functional layers of a laminated
piezoelectric element used in the present study; (b) definitions of the
length, L
p
, and the bending angle, α.
Fig. 4 Open-circuit output voltages of a single piezoelectric element
as a function of the tip bending angle.
Fig. 5 Total voltage outputs from multiple piezoelectric elements
connected in series under different tip bending angles.
Lab on a Chip Paper
Published on 18 December 2013. Downloaded on 20/02/2014 17:10:00.
View Article Online

1120 | Lab Chip,2014,14, 11171122 This journal is © The Royal Society of Chemistry 2014
the order of 100 V can be reliably generated using piezoelectric
elements connected in series with tip bending angles <90°.
EWOD actuation of micro droplets
Basic droplet operations
We next demonstrated successful finger-powered actuation of
water droplets on EWOD devices with 0.8 μm-thick PECVD
SiNx dielectric layers. An array of single piezoelectric ele-
ments is used to convert mechanical energy to voltage pulses.
Fig. 6 shows optical images of a single water droplet (~0.15 μL)
being transported over adjacent electrodes through a sequence
of finger-driven deflection of the piezoelectric elements. To
avoid a droplet being trapped on an inactive zone between two
adjacent electrodes, we delayed the release of the previously
bent piezoelectric element while deflecting the neighbouring
element. For example, referring to Fig. 6, element 2 was not
entirely released when element 3 was deflected, such that the
front contact line of the droplet would stay across the gap
between electrodes 2 and 3. For such a delayed release, a
holding force of approximately 0.06 N is sufficient.
We also demonstrated droplet splitting by simultaneously
deflecting two non-adjacent piezoelectric elements while
keeping the middle one non-deflected. Fig. 7 illustrates a
splitting of a water droplet (~0.3 μL) into two daughter
droplets of similar sizes through simultaneous deflection of
element 1 and 3. When splitting was in order, the droplet was
elongated in the longitudinal direction by a wetting force
exerted at the two ends while keeping the middle non-wetting,
as shown in Fig. 6. The actuation voltage on either side is
approximately 4050 V, produced by one single piezoelectric
element with a bending angle <90°. For a large droplet that
covers multiple electrodes, asymmetric splitting can also be
achieved by simultaneously deflecting a set of piezoelectric
elements connected to the electrodes in an asymmetric man-
ner, for example, elements 1 and 3, 4. Fig. 8 shows a merging
of two droplets of similar sizes by asynchronous bending and
release of three piezoelectric elements (2, 4, and 3).
Applications to biological assays
As proof of principle demonstration of biological applications
of our F-DMF, we performed basic steps of glucose detection
and immunoassay. In these experiments, where higher
EWOD voltages are necessary to actuate droplets, we used
silicon nitride layers of thickness 2.5 μm because they help
prevent electrolysis of water under our low-frequency (~1 Hz)
finger-driven actuation scheme. Two piezoelectric elements
connected in series were used to provide an actuation voltage
of up to 100 V.
We demonstrated glucose detection based on an enzy-
matic oxidation, where the color of an assay solution changes
from clear to brown in the presence of glucose. Since the
reagent solution may chemically attack the hydrophobic coat-
ings, we pre-treated both the upper plate and bottom sub-
strate of our EWOD devices with silicone oil. A reagent
solution was prepared by adding 0.8 mL of o-dianisidine
reagent to an amber bottle containing 39.2 mL of a 1 : 5
horseradish peroxidase/glucose oxidase solution (15 units of
protein per mL of solution). We spotted a droplet of a 1 mg mL
1
Fig. 6 Finger-actuated EWOD transport of a water droplet where
actuation voltage pulses were provided by bending a series of
piezoelectric elements.
Fig. 7 Finger-actuated EWOD droplet splitting. The piezoelectric
elements 1 and 3 are bent simultaneously for splitting.
Fig. 8 Finger-actuated EWOD droplet merging.
Lab on a ChipPaper
Published on 18 December 2013. Downloaded on 20/02/2014 17:10:00.
View Article Online

Citations
More filters
Journal ArticleDOI
TL;DR: In this paper, a simple and effective ratchet-like strategy (relaxing and squeezing actions) is introduced to transfer liquid droplets using a parallel plate structure (PPS) to maintain the stable and efficient operation of the system.
Abstract: Due to the superior operability and good anti-interference, the prospect of controlling microdroplets using a parallel plate structure (PPS) is very promising. However, in practical applications, droplets in such structures are often affected by various factors, resulting in deformation, evaporation, stress rupture and other phenomena, leading to equipment failure. Therefore, how to simply and effectively transfer liquid droplets to PPS to maintain the stable and efficient operation of the system has become an urgent problem to be solved. In this paper, a simple and effective ratchet-like strategy (relaxing and squeezing actions) is introduced to transfer droplets. To analyze the mechanism of the strategy and optimize the control, we conduct this study from three aspects. First, the droplet movement trend is obtained by analyzing the pressure between SPS and PPS. Second, the reasons why the droplet can achieve this inward motion are investigated. Through theoretical analysis, which is also proven by simulations and experiments, we creatively put forward that the asymmetric change of the contact angle (CA) induced by the asymmetric structure is the fundamental cause of this kind of motion. Due to the asymmetric change of the contact angle, the CA in the PPS will reach the advancing angle first in the squeezing process, and the CA in the SPS will reach the receding angle first in the relaxing process, thus causing the inward movement of the droplet. Third, to optimize this strategy, the effects of the following governing parameters are researched individually based on the corresponding simulations and experiments: the control parameters (the initial gap width of the PPS H0 and the amount of squeezing and relaxing of ΔH) and the thickness of the top plate. Subsequently, an optimized ratchet-like cycle is achieved. In summary, these findings not only provide a new method by which to realize the movement of droplets toward hydrophobic PPSs but also creatively point out the cause of the ratchet strategy, which can be applied in many microfluidics fields.

2 citations

Book ChapterDOI
01 Jan 2020
TL;DR: In this chapter, DMF technology from the aspects of fundamentals, actuation mechanisms, and device fabrication is described and current progress in biomedical devices of DMF for healthcare is reviewed and evaluated.
Abstract: Digital microfluidics (DMF) breaks up the continuous flow of analysts’ solution into discrete droplets and offers radical improvements on the manipulation of pL-sized to µL-sized droplets moving, merging, splitting, and dispensing from reservoirs. Due to its unique advantages including precise control, high throughput, less cross-contamination, simple instrumentation, flexible device geometry, and easy coupling with other technologies, its applications have been extremely promoted in the last two decades. A variety of biomedical devices have been implemented and demonstrated for healthcare ranging from biochemical synthesis to molecular diagnostics and immunoassays. In this chapter, DMF technology from the aspects of fundamentals, actuation mechanisms, and device fabrication is described. Current progress in biomedical devices of DMF for healthcare is reviewed and evaluated. Download : Download full-size image

2 citations

Proceedings ArticleDOI
02 Mar 2015
TL;DR: In this paper, the authors introduce an advantageous self-mixing phenomenon created during the formation of aqueous two-phase systems in a digital microfluidic device, for the first time.
Abstract: This paper introduces an advantageous self-mixing phenomenon created during the formation of aqueous two-phase systems in a digital microfluidic device, for the first time The influence of such self-mixing on enhancement of on-chip extraction kinetics is quantified and reported, highlighting a significant mixing capability in the absence of forced mixing Results suggest potential applications of the self-mixing concept in portable microfluidics, where implementing complicated mixing sequences or device geometries are problematic

2 citations


Cites methods from "EWOD (electrowetting on dielectric)..."

  • ...EWOD DMF using finger actuation[10] shows the potential of developing portable DMF devices that does not require power input in complicated sequences....

    [...]

Journal ArticleDOI
TL;DR: The experimental results demonstrate that the electric field plays an important role in patterning nanoparticles, which is important to explain the mechanism that SSAWs enhance reaction and crystallization in droplet.
Abstract: In this paper, the particle movements in a sessile droplet induced by standing surface acoustic waves (SSAWs) are studied. Tritoroidal particle rings are formed under the interaction of acoustic field and electric field. The experimental results demonstrate that the electric field plays an important role in patterning nanoparticles. The electric field can define the droplet shape due to electrowetting. When the droplet approximates a hemisphere, the acoustic radiation force induced by SSAWs drives the particles to form tritoroidal particle rings. When the droplet approximates a convex plate, the drag force induced by acoustic steaming drives the particle to move. The results will be useful for better understanding the nanoparticle movements in a sessile droplet, which is important to explain the mechanism that SSAWs enhance reaction and crystallization in droplet.

2 citations

Journal ArticleDOI
TL;DR: A comprehensive review of recent advances in the programmable actuation of liquid matter controlled by unique substrate structures and various physical fields is presented in this paper , where the fundamental theories of asymmetric surface microstructure and specific external effects such as electric, magnetic, acoustic, light, and thermal fields are illustrated.
Abstract: Intelligent responses, transformation and desired delivery of liquids involve the intersection among materials, physics, and chemistry. The precise programmed operation has been found rather significant in many emerging fields, such as microfluidic devices, digital fluids, intelligent matter, responsive liquid machines, interactive display, soft robotic system and drug delivery, etc. This article is dedicated to presenting a comprehensive review of recent advances in the programmable actuation of liquid matter controlled by unique substrate structures and various physical fields. First, the fundamental theories of asymmetric surface microstructure and specifically administrated external effects such as electric, magnetic, acoustic, light, and thermal fields to drive droplets are illustrated. Then, we compare the advantages and disadvantages of these typical technological strategies. Furthermore, the practical multifunctional extension of programmable liquids based on different actuation mechanisms is presented. Finally, we summarize the design schemes of programming units for liquid matter and give perspective on possible future research directions, providing a brief reference for researchers interested in the field.

1 citations

References
More filters
Journal ArticleDOI
TL;DR: Microfluidic paper-based analytical devices are a new class of point-of-care diagnostic devices that are inexpensive, easy to use, and designed specifically for use in developing countries.
Abstract: Microfluidic paper-based analytical devices (μPADs) are a new class of point-of-care diagnostic devices that are inexpensive, easy to use, and designed specifically for use in developing countries. (To listen to a podcast about this feature, please go to the Analytical Chemistry multimedia page at pubs.acs.org/page/ancham/audio/index.html.)

2,373 citations

Journal ArticleDOI
TL;DR: This review includes challenges to scaling up, commercialisation and regulatory issues, and the factors which limit paper-based microfluidic devices to become real world products and future directions are also identified.
Abstract: Dipstick and lateral-flow formats have dominated rapid diagnostics over the last three decades. These formats gained popularity in the consumer markets due to their compactness, portability and facile interpretation without external instrumentation. However, lack of quantitation in measurements has challenged the demand of existing assay formats in consumer markets. Recently, paper-based microfluidics has emerged as a multiplexable point-of-care platform which might transcend the capabilities of existing assays in resource-limited settings. However, paper-based microfluidics can enable fluid handling and quantitative analysis for potential applications in healthcare, veterinary medicine, environmental monitoring and food safety. Currently, in its early development stages, paper-based microfluidics is considered a low-cost, lightweight, and disposable technology. The aim of this review is to discuss: (1) fabrication of paper-based microfluidic devices, (2) functionalisation of microfluidic components to increase the capabilities and the performance, (3) introduction of existing detection techniques to the paper platform and (4) exploration of extracting quantitative readouts via handheld devices and camera phones. Additionally, this review includes challenges to scaling up, commercialisation and regulatory issues. The factors which limit paper-based microfluidic devices to become real world products and future directions are also identified.

1,658 citations

Journal ArticleDOI
TL;DR: In this paper, the authors report the completion of four fundamental fluidic operations considered essential to build digital microfluidic circuits, which can be used for lab-on-a-chip or micro total analysis system (/spl mu/TAS): 1) creating, 2) transporting, 3) cutting, and 4) merging liquid droplets, all by electrowetting.
Abstract: Reports the completion of four fundamental fluidic operations considered essential to build digital microfluidic circuits, which can be used for lab-on-a-chip or micro total analysis system (/spl mu/TAS): 1) creating, 2) transporting, 3) cutting, and 4) merging liquid droplets, all by electrowetting, i.e., controlling the wetting property of the surface through electric potential. The surface used in this report is, more specifically, an electrode covered with dielectrics, hence, called electrowetting-on-dielectric (EWOD). All the fluidic movement is confined between two plates, which we call parallel-plate channel, rather than through closed channels or on open surfaces. While transporting and merging droplets are easily verified, we discover that there exists a design criterion for a given set of materials beyond which the droplet simply cannot be cut by EWOD mechanism. The condition for successful cutting is theoretically analyzed by examining the channel gap, the droplet size and the degree of contact angle change by electrowetting on dielectric (EWOD). A series of experiments is run and verifies the criterion.

1,522 citations

Journal ArticleDOI
TL;DR: This work presents an alternative paradigm--a fully integrated and reconfigurable droplet-based "digital" microfluidic lab-on-a-chip for clinical diagnostics on human physiological fluids, and demonstrates reliable and repeatable high-speed transport of microdroplets.
Abstract: Clinical diagnostics is one of the most promising applications for microfluidic lab-on-a-chip systems, especially in a point-of-care setting. Conventional microfluidic devices are usually based on continuous-flow in microchannels, and offer little flexibility in terms of reconfigurability and scalability. Handling of real physiological samples has also been a major challenge in these devices. We present an alternative paradigm—a fully integrated and reconfigurable droplet-based “digital” microfluidic lab-on-a-chip for clinical diagnostics on human physiological fluids. The microdroplets, which act as solution-phase reaction chambers, are manipulated using the electrowetting effect. Reliable and repeatable high-speed transport of microdroplets of human whole blood, serum, plasma, urine, saliva, sweat and tear, is demonstrated to establish the basic compatibility of these physiological fluids with the electrowetting platform. We further performed a colorimetric enzymatic glucose assay on serum, plasma, urine, and saliva, to show the feasibility of performing bioassays on real samples in our system. The concentrations obtained compare well with those obtained using a reference method, except for urine, where there is a significant difference due to interference by uric acid. A lab-on-a-chip architecture, integrating previously developed digital microfluidic components, is proposed for integrated and automated analysis of multiple analytes on a monolithic device. The lab-on-a-chip integrates sample injection, on-chip reservoirs, droplet formation structures, fluidic pathways, mixing areas and optical detection sites, on the same substrate. The pipelined operation of two glucose assays is shown on a prototype digital microfluidic lab-on-chip, as a proof-of-concept.

1,124 citations

Journal ArticleDOI
TL;DR: The performance of magnetic bead-based immunoassays (cardiac troponin I) on a digital microfluidic cartridge in less than 8 minutes using whole blood samples and the capability to perform sample preparation for bacterial infectious disease pathogen, methicillin-resistant Staphylococcus aureus and for human genomic DNA using magnetic beads are demonstrated.
Abstract: Point of care testing is playing an increasingly important role in improving the clinical outcome in health care management. The salient features of a point of care device are rapid results, integrated sample preparation and processing, small sample volumes, portability, multifunctionality and low cost. In this paper, we demonstrate some of these salient features utilizing an electrowetting-based Digital Microfluidic platform. We demonstrate the performance of magnetic bead-based immunoassays (cardiac troponin I) on a digital microfluidic cartridge in less than 8 minutes using whole blood samples. Using the same microfluidic cartridge, a 40-cycle real-time polymerase chain reaction was performed within 12 minutes by shuttling a droplet between two thermal zones. We further demonstrate, on the same cartridge, the capability to perform sample preparation for bacterial infectious disease pathogen, methicillin-resistant Staphylococcus aureus and for human genomic DNA using magnetic beads. In addition to rapid results and integrated sample preparation, electrowetting-based digital microfluidic instruments are highly portable because fluid pumping is performed electronically. All the digital microfluidic chips presented here were fabricated on printed circuit boards utilizing mass production techniques that keep the cost of the chip low. Due to the modularity and scalability afforded by digital microfluidics, multifunctional testing capability, such as combinations within and between immunoassays, DNA amplification, and enzymatic assays, can be brought to the point of care at a relatively low cost because a single chip can be configured in software for different assays required along the path of care.

559 citations

Frequently Asked Questions (1)
Q1. What are the contributions in "Ewod (electrowetting on dielectric) digital microfluidics powered by finger actuation" ?

In this paper, a finger-actuated digital microfluidics based on electrowetting-on-dielectric ( EWOD ) was demonstrated by using piezoelectric energy conversion of human power.