Physical performance of the new hybrid PET∕CT Discovery-690.

Reads0
Chats0
TLDR
The Discovery-690 shows very good PET physical performance for all the standard NEMA NU-2-2007 measurements and the new reconstruction algorithms available for PET data (TOF and PSF) allow further improvements of the D-690 image quality performance both qualitatively and quantitatively.
Abstract
Purpose The aim of this work was the assessment of the physical performance of the new hybrid PET∕CT system: Discovery-690. Methods The Discovery-690 combines a lutetium-yttrium-orthosilicate (LYSO) block detector designed PET tomograph with a 64-slice CT scanner. The system is further characterized by a dedicated powerful computing platform implementing fully 3D-PET iterative reconstruction algorithms. These algorithms can account for time of flight (TOF) information and∕or a 3D model of the PET point spread function (PSF). PET physical performance was measured following NEMA NU-2-2007 procedures. Furthermore, specific tests were used: (i) to measure the energy and timing resolution of the PET system and (ii) to evaluate image quality, by using phantoms representing different clinical conditions (e.g., brain and whole body). Data processing and reconstructions were performed as required by standard procedures. Further reconstructions were carried out to evaluate the performance of the new reconstruction algorithms. In particular, four algorithms were considered for the reconstruction of the PET data: (i) HD = standard configuration, without TOF and PSF, (ii) TOF = HD + TOF, (iii) PSF = HD + PSF, and (iv) TOFPSF = HD + TOF + PSF. Results The transverse (axial) spatial resolution values were 4.70 (4.74) mm and 5.06 (5.55) mm at 1 cm and 10 cm off axis, respectively. Sensitivity (average between 0 and 10 cm) was 7.5 cps∕kBq. The noise equivalent count rate (NECR) peak was 139.1 kcps at 29.0 kBq∕ml. The scatter fraction at the NECR peak was 37%. The correction accuracy for the dead time losses and random event counts had a maximum absolute error below the NECR peak of 2.09%. The average energy and timing resolution were 12.4% and 544.3 ps, respectively. PET image quality was evaluated with the NEMA IEC Body phantom by using four reconstruction algorithms (HD, TOF, PSF, and TOFPSF), as previously described. The hot contrast (after 3 iterations and for a lesion∕background activity ratio of 4:1) for the spheres of 10, 13, 17, and 22 mm was (HD) 29.8, 45.4, 55.4, and 68.1%; (TOF) 39.9, 53.5, 62.7, and 72.2%; (PSF) 28.3, 47.3, 60.4, and 71.8%; (TOFPSF) 43.8, 62.9, 70.6, and 76.4%. The cold contrast for the spheres of 28 and 37 mm was (HD) 62.4 and 65.2%; (TOF) 77.1 and 81.4%; (PSF) 62.0 and 65.2%; (TOFPSF) 77.3 and 81.6%. Similar hot and cold contrast trends were found during the analyses of other phantoms representing different clinical conditions (brain and whole body). Nevertheless, the authors observed a predominant role of either TOF or PSF, depending on the specific characteristics and dimensions of the phantoms. Conclusions Discovery-690 shows very good PET physical performance for all the standard NEMA NU-2-2007 measurements. Furthermore, the new reconstruction algorithms available for PET data (TOF and PSF) allow further improvements of the D-690 image quality performance both qualitatively and quantitatively.

read more

Citations
More filters
Journal ArticleDOI

Recent developments in time-of-flight PET

TL;DR: Recent developments in system design, image reconstruction, corrections, and the potential in new applications for TOF-PET are reviewed to introduce the reader in an educational way into the topic of time-of-flight-PET.
Journal ArticleDOI

NEMA NU 2-2012 performance studies for the SiPM-based ToF-PET component of the GE SIGNA PET/MR system.

TL;DR: Performance measurements of the ToF-PET whole body GE SIGNA PET/MR system indicate that it is a promising new simultaneous imaging platform.
Journal ArticleDOI

Studies of a Next Generation Silicon-Photomultiplier-Based Time-of-Flight PET/CT System

TL;DR: Excellent performance was achieved with the Discovery MI, including 375 ps FWHM coincidence time resolution and sensitivity of 14 cps/kBq, and comparisons between reconstruction algorithms and other multimodal silicon photomultiplier and non–silicon photomULTiplier PET detector system designs indicated that performance can be substantially enhanced with this next-generation system.
Journal ArticleDOI

PET Image Denoising Using a Deep Neural Network Through Fine Tuning

TL;DR: A deep convolutional neural network was trained to improve PET image quality by employingceptual loss based on features derived from a pretrained VGG network instead of the conventional mean squared error to preserve image details.
Journal ArticleDOI

Understanding and simulating SiPMs

TL;DR: In this paper, the authors consider the structure and electrical model of a single photon avalanche diode (SPAD), and the integration in an array, i.e., the SiPM.
References
More filters
Journal Article

A combined PET/CT scanner for clinical oncology.

TL;DR: A combined PET and CT scanner is a practical and effective approach to acquiring co-registered anatomical and functional images in a single scanning session.
Journal ArticleDOI

Analytic 3D image reconstruction using all detected events

TL;DR: In this paper, the authors present an algorithm for three-dimensional image reconstruction that uses all gamma-ray coincidence events detected by a PET (positron emission tomography) volume imaging scanner.
Journal ArticleDOI

High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner

TL;DR: Reconstructions obtained from point source data using the accurate system model demonstrate a potential for near-isotropic FWHM resolution of approximately 1.2 mm at the center of the field of view compared with approximately 2 mm when using an analytic 3D reprojection method with a ramp filter.
Journal Article

Performance of Philips Gemini TF PET/CT Scanner with Special Consideration for Its Time-of-Flight Imaging Capabilities

TL;DR: The Gemini TF whole-body scanner represents the first commercially available fully 3-dimensional PET scanner that achieves time-of-flight capability as well as conventional imaging capabilities.
Journal ArticleDOI

Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents counts

TL;DR: It is demonstrated that, compared to the Alderson brain phantom, the standard 20-cm cylinder is a poor predictor of count rate performance for PET brain imaging.
Related Papers (5)