This work describes a method wherein true RF A-lines are acquired and augmented at subpitch locations using CLA transducer, instead of interpolating the data, and using this new frame data for further image formation and/or processing to yield better lateral resolution and LDE.
Abstract
In elastography, conventional linear array (CLA)-based RF data acquisition provides more accurate displacement measurements in the direction of beam propagation (axial direction) when compared to the perpendicular direction (lateral). Obtaining good quality lateral displacement estimates in ultrasound (US) elastography will lead to several benefits such as obtaining accurate inverse solutions, improving shear strain elastogram quality, getting good quality poroelastograms, and obtaining reliable rotation elastograms. For accomplishing high-precision lateral displacement estimation (LDE), one of the popular methods is by interpolating additional A-lines in between neighboring RF A-lines. We describe a method wherein true RF A-lines (not interpolated) are acquired and augmented at subpitch locations using CLA transducer, instead of interpolating the data, and using this new frame data for further image formation and/or processing to yield better lateral resolution and LDE. We demonstrate the proposed method by translating the US beam of CLA transducer in subpitch range by the following two approaches: 1) actuator-assisted beam translation and 2) electronic translation of subaperture of a CLA by activating odd and even number of consecutive elements sequentially, referred to as electronic beam translation. The performances of the different methods were studied through simulations and experiments on phantoms. The results demonstrate that these methods yield better quality LDE compared to those obtained from interpolation of RF A-lines. These methods may provide affordable ways to obtain subpitch precision LDE using CLA.

read more

Citations
More filters
Journal ArticleDOI

3D normalized cross-correlation for estimation of the displacement field in ultrasound elastography.

TL;DR: A novel technique to estimate tissue displacement in quasi-static elastography is introduced and two approaches are proposed to get third dimension, which substantially outperforms NCC using simulation, phantom and in-vivo experiments.
Journal ArticleDOI

Towards quantitative quasi-static ultrasound elastography using a reference layer for liver imaging application: A preliminary assessment.

TL;DR: A method to make ultrasound elastography quantitative using a calibrated reference layer, sandwiched between the transducer and the tissue surface, to quantitatively image the unknown modulus of the examined tissue is reported.
Proceedings ArticleDOI

Incorporating Multiple Observations in global Ultrasound Elastography

TL;DR: This paper proposes a novel framework for time delay estimation in ultrasound elastography using multiple observations (mGLUE), and primary validation of mGLUE against soft and hard inclusion simulation phantoms proves that m GLUE is capable of obtaining high quality strain map while dealing with noisy ultrasound data.
Journal ArticleDOI

Enhancing Image Quality of Photoacoustic Tomography Using Sub-Pitch Array Translation Approach: Simulation and Experimental Validation

TL;DR: The proposed linear translation-based approach has the potential to serve as an easy-to-integrate and simple way of achieving better image quality without requiring to increase the system complexity with existing transducer array probe technology in regular clinical scanners.
Journal ArticleDOI

Lateral Strain Imaging Using Self-Supervised and Physically Inspired Constraints in Unsupervised Regularized Elastography

TL;DR: Wang et al. as mentioned in this paper proposed Physically Inspired ConsTraint for Unsupervised Regularized Elastography (PICTURE), where they imposed a constraint on the Effective Poisson's ratio (EPR) to improve the lateral strain estimation.
References
More filters
Journal ArticleDOI

Effect of skin impedance on image quality and variability in electrical impedance tomography: a model study

TL;DR: It is concluded that the patient's skin should be abraded to reduce impedance, and measurements should be avoided in the first 10 min after electrode placement, to allow satisfactory images.
Journal ArticleDOI

Statistics of Speckle in Ultrasound B-Scans

TL;DR: In this article, the authors derived autocorrelation functions and power spectra derived from B-scans of a scattering phantom containing many scatterers per resolution cell, leading naturally to the definition of the average speckle spot or cell sue, and this inturn is comparable to the resolution cell.
Journal ArticleDOI

Synthetic Aperture Ultrasound Imaging

TL;DR: The paper describes the use of synthetic aperture (SA) imaging in medical ultrasound, where data is acquired simultaneously from all directions over a number of emissions, and the full image can be reconstructed from this data.
Journal ArticleDOI

Tissue response to mechanical vibrations for “sonoelasticity imaging”

TL;DR: Preliminary results support the postulate that sonoelasticity imaging can provide useful information concerning tissue properties that are not otherwise obtainable.
Journal ArticleDOI

Phantom materials for elastography

TL;DR: In this article, a device and procedure for measuring elastic properties of gelatin for elasticity imaging (elastography) was described. And the measured compression forces were comparable to results obtained from finite element analysis when linear elastic media are assumed.