scispace - formally typeset
Search or ask a question
Proceedings ArticleDOI

Wire scanners for beam size and emittance measurements at the SLC

06 May 1991-Vol. 910506, pp 1201-1203
TL;DR: In this article, the authors describe the design, construction, commissioning and ultimate uses of wire scanners in the SLAC Linear Collider (SLAC) focusing on the linear accelerator and upstream systems scanners.
Abstract: The authors describe the design, construction, commissioning and ultimate uses of wire scanners in the SLC (SLAC Linear Collider), focusing on the linear accelerator and upstream systems scanners. Of particular interest is the interaction between the wire and the scattered radiation from the wire with the extreme electric field of the beam. As this field reaches the level of several volts/angstrom, as it does easily at the SLC interaction point (and may in upstream parts of SLC), field emission from the wire may occur. A key feature of SLC operation is the degree of high level active control required to keep it optimized. Advanced high level control software allows the use of wire scanner data in feedback and beam optimization procedures. Non-invasive scans are performed almost continually and the results are logged so that long term trends in emittance can be examined. >

Summary (2 min read)

INTRODUCTION

  • Measurements of the beam size and associated optical parameters are key to SLC.
  • In the past progress has been slowed by their inability to measure beam size accurately in a rapid, non-invasive manner.
  • Construction, commissioning and ultimate uses of wire scanners in the SLC focusing on the linac and upstream systems scanners.the authors.
  • Of particular interest is the interaction between the wire and the scattered radiation from the wire with the extreme electric Field of the beam.
  • As this field reaches the level of several volts/angstrom, as it does easily at the SLC interaction point (and may in upstream parts of SLC), field emission from the wire may occur.

Tungsten Wire

  • The high level of demand takes the instrument out of the category of a device primarily used for machine development or failure diagnosis purposes and elevates it to an online device.
  • Feedback requires a fast, non-invasive (or minimal impact) device which in turn means that the wire, not the beam, must be moved during the scan.
  • The speed, range, vibration and other mechanical specifications can be generated from this requirement and from the expected beam sizes and rates.
  • Tabic 1 shows the expected performance of the scanners.

CONTROLS

  • A scan consists of three steps: 1) move from PARK (near, but not in, any of the beams) to the start of scan at maximum speed, 2) scan at the speed corresponding to the desired interpoint spacing and beam rate and 3) return to PARK at full speed.
  • Only brief pauses, to allow the device to come to a complete stop, occur during the scan sequence.
  • The control system software built around these devices allows use of the wire scanner at several levels.
  • At the next level higher, the fit results can be used in the SLC control system correlation plot ulilily.'' allowing the acquisition of scan data with other beam diagnostic data and machine parameters, has been invaluable for commissioning the scanners.

SIGNAL DETECTOR

  • The purpose of the wire signal detector is to indicate the amount of charge striking the wire.
  • Secondary emission and forward scattering, used in many wire scanner systems, are often not practical at the SLC, die first because of problems discussed below and the second because or the very different beam line areas in which these devices have ID operate.
  • Because of these problems, tests were made to determine if a significant signal was present at 90' to the beam direction, directly opposite the wire support card.
  • In order to cross check PMT performance small fast ion chambers were built and installed downstream of some scanners 6.
  • The linearity of die ADC is also very important.

PERFORMANCE AND CONCLUSIONS

  • The SLC wire scanners provide beam cmittance dr.ta that is reliable enough to have already yielded new insights into the performance of the SLC.
  • Piezo-elcctric motors, with their very small step size and ultra-high vacuum compatibility, may prove to be an appropriate technology for future wire scanners.
  • Two such scanners are installed for use at die SLC IP.

Did you find this useful? Give us your feedback

Content maybe subject to copyright    Report

1?
'9s$
SLAC-PUB—5556 *' t
DE91 015482
Wire Scanners for Beam Size and Emittance Measurements at the SLC*
M.C.
Ross, J.T. Seeman, E. Bong, L. Hendrickson, D. McComiick, L. Sanchez-Chopitea
Stanford Linear Accelerator Center,
Stanford University, Stanford, Ca. 94309
ABSTRACT
The SLC wire scanner beam profile monitors provide
accurate beam size
and
emittance measurements
for each
bunch
in the three bunch SLC beam. The beam size measurement
error for typical 50GcV SLC linac beams (lOOum
o-(x.y))
is
better than 5p.m. Beam profile measurements can
be
performed
throughout much of the SLC with no interruption to normal
machine operation
and
no adverse impact on interaction region
detector backgrounds. The linac input and output omittance is
determined using sets of four scanners spaced by -45' betatron
phase advance. Each scanner contains three wires, x, y and u
(45*),
from which an estimate of the x - y coupling can be
obtained. Advanced high level control software allows the use
of wire scanner data in feedback and beam optimization
procedures. Non-invasive scans are performed almost
continually and the results arc logged so that long term trends
in emiliancc can be examined. In this paper we describe the
design, construction, performance and uses of SLC wire
scanners.
INTRODUCTION
Measurements of the beam size and associated optical
parameters are key to SLC. In the past progress has been
slowed by our inability to measure beam size accurately in a
rapid, non-invasive manner.
Tungsten
Wire
Figure
1:
SLC wire scanner wire support
card
viewed
in
the
beam direction. The translation stage and vacuum chamber are
not shown.
work supported by Department of Energy contract
DE-AC03-76SF00515
In this paper we describe the design, construction,
commissioning and ultimate uses of wire scanners in the SLC
focusing on the linac and upstream systems scanners. Of
particular interest is the interaction between the wire and the
scattered radiation from the wire with the extreme electric Field
of the beam. As this field reaches the level of several
volts/angstrom, as it does easily at the SLC interaction point
(and may in upstream parts of SLC), field emission from the
wire may occur.
A key feature of
SLC
operation is the degree of high level
active control required to keep it optimized. The high level of
demand takes the instrument out of the category of a device
primarily used for machine development or failure diagnosis
purposes
and
elevates it to an online device.
Feedback requires a fast, non-invasive (or minimal impact)
device which in turn means that the wire, not the beam, must
be moved during the scan. The speed, range, vibration and
other mechanical specifications can be generated from this
requirement
and
from the expected beam sizes and rates. Tabic
1 shows
the
expected performance of
the
scanners.
MECHANICAL
The mechanical design effort addressed the following
problems: 1) wire and wire retention, 2) vibration over the
large speed range and 3) positioning errors and position
transducers.
A
particular concern
was radiation
damage.
A schematic diagram of the scanner is shown in figure 1.
Several labs hnve built scanners of a similar design
1
>
2
>'.
The
wire is strung around 1.5mm stainless steel studs set in a
3/16in thick alumina fork in such a way so that it can carry
wires of three different orientations across the beam and
provide x, y and u (45') scans. The carriage motion is actuated
by a stepping motor through a 2mm pitch ball screw, chosen
because of the expected large number of cycles. Some
difficulty was experienced obtaining the small pitch, high
quality ball screw with no plastic parts. A 125um thick
stainless steel vacuum window opposite the wire allows low
energy wide angle scattered radiation to emerge from the
vacuum chamber.
Both the cantilever nature of the wire support and the
stepping motor contribute to wire vibration. We have used a
piezoelectric accelcrometcr to quantify the motor related
system vibration.
The wire chosen for the scanner was gold plated tungsten
with a diameter of 0.3 Obcam- ^
nc w
'
rc nas an
effective
'a'
of
radius/2 which, when added in quadrature to the beam size,
causes a
3%
apparent increase when 0"bcam =
w
irc diameter.
Under normal conditions the wire size can be subtracted in
quadrature from the measured size.
Presented at the IEEE Particle Accelerator Conference, San Francisco
_ CiC '.'i"i,0'-.7IGjN O"
Tj|
«. 'VlPi "s«cv
;.JASP
#>

At Ml SLC currents and rates, the beam can heat the wire
substantially reaching a steady state temperature of lOOO'C. In
the fast scan mode, the wire is subjected to continuous beam
for no more than a few seconds. However, the wire may be
parked in the beam in error or
Tor
diagnostic purposes and musl
be able to withstand continuous beam. By measuring the
increase in resistance of the wire assembly an estimate of die
wire temperature rise can be made. Resistance tests show good
agreement with
a
calculated rise of
about
4' C/pulse.
Table 1: Wire scanner performance specifications.
Beam size resolution
<3%
a (= 3 -
10
urn)
Systematic error
<3%
a
(=
3
- 10
um)
Emitlancc
(E)
error
10% (for
IE=3
x 10"
5
m-rad)
Dynamic range
10
9
- 10
n
particles/pulse
Vibration Peak amplitude < 0.2a
Relative positioning
20um
Speed
and
acceleration
lem/s max
0.3mm/s min; 0.2m/s^ acccl.
Multibunch operation
<5%
signal contamination from
nearby bunch (60ns)
Radiation resistant
lOMrad/ycar
Lifetime 100,000 cycles/year
As the wire is scanned through the beam its position is not
encoded on each successive beam pulse, rather the current
position of me wire is inferred from a check of me remaining
step count. A position measurement using a radiation hard
LVDT (linear differential transformer) is done at die limits of
ihc scan to check mat the expected position was reached.
CONTROLS
A scan consists of
three
steps: 1) move from PARK (near,
but not in, any of the beams) to the start of scan at maximum
speed, 2) scan at the speed corresponding to the desired inter-
point spacing and beam rate and 3) return to PARK at full
speed. Only brief pauses, to allow the device to come to a
complete stop, occur during the scan sequence. During
muliiwire emillancc or skew scans, the wire moves to die next
appropriate PARK. An important feature of this scheme is die
use of machine wide data acquisition codes which coordinate
die readback of
the
scanner step count and the signals from the
wire scanner detector and therefore allow a great deal of
flexibility in the choice of detector including,
Tor
example, the
use of detectors several miles away.
Because die scanners are to be used for feedback, the
application software that controls them must have
sophisticated exception handling, error logging and status
reporting. The control system software built around these
devices allows use of the wire scanner at several levels. The
lowest is the single scan and associated single pulse detector
signal readout. This information allows checking the Tit
quality, scan ranges and other details. At the next level higher,
the fit results can be used in the SLC control system
correlation plot ulilily.'' This extremely powerful tool.
allowing the acquisition of scan data with other beam
diagnostic data and machine parameters, has been invaluable
for commissioning the scanners. All aspects of
the
gaussian fit
to the scan data
arc
available and arc automatically acquired as
an upstream device setting is controlled in
a
programmed way.
Most automated optimization procedures arc built around this
facility. The next level of software docs multiple scans and
accumulates these results in the correlation plot utility. This
includes four wire emittancc scan results and skew scan results.
Finally, feedback can perform the scans as a background task
and implement the desired corrections.
SIGNAL DETECTOR
The purpose of the wire signal detector is to indicate the
amount of charge striking the wire. Secondary emission and
forward scattering, used in many wire scanner systems, are
often not practical at the SLC, die first because of problems
discussed below and the second because or the very different
beam line areas in which these devices have
ID
operate. The
most difficult region is just upstream of a high power
collimator system. Radiation scattered by the wire in the
forward direction is completely overwhelmed by the scattering
from the collimator jaws thus making the use of small angle
scattered radiation impossible.
When me SLC interaction point (IP) wire scanners
5
were
first tested it was found that the secondary emission signal
would increase dramatically when either the beam intensity
exceeded about 5 x 10' or the Cbeam
<
10p.m. The onset of
ihis dramatic increase is indicated by a very unstable signal. In
die SLC linac, where the beam sizes are about 100pm, a beam
intensity of about 2.5 x lO
1
^ is required. This effect appears to
be
field
emission induced by die field of
the
beam which peaks
at about 20V/A.
Because of
these
problems, tests were made to determine if
a significant signal was present at 90' to the beam direction,
directly opposite the wire support card. A strong, very low
energy election signal was seen in a bare phoiomuliiplicr
(PMT) placed about 30cm from the wire. A ihin window is
required so mat this scattered radiation is not absorbed by the
vacuum chamber wall. Substantial shielding (±50 radiation
lengths) is required in some locations to protect the PMTs
from background generated by upstream beam losses. In
regions where no collimators follow the scanners, small angle
scattering monitors have been placed about 10m downstream
to use for comparison with the PMT.
The detector linearity musl be better than a few percent. A
moderate gain, excellent linearity lube was chosen. In order to
cross check PMT performance small fast ion chambers were
built
and
installed downstream of some scanners
6
The linearity
of die ADC is also very important.
PERFORMANCE AND CONCLUSIONS
Performance tcsls have focused
on
understanding systematic
errors. These tests Tall into two broad categories: 1) tests made
by varying beam size and/or intensity and 2)
ICSL>
made with
different detectors sensing scattered radiation from a single

-1000 0
Wire position (|im)
Figure
2:
A typical single wire scan. The data arc fit with a
gaussian with an offset.
wire.
Figure 3 shows the beam emittance measured using
quadrupole scans
7
on three different scanners and multiwirc
scans at
1
GeV and 4,5 x 10
1
" e/'puise, near the SLC nominal
operating intensity. This is a good test of saturation effects
since the beam sizes and signal strengths vary considerably
over the scan range and from scanner to scanner. These tests
were done using the downstream fast ion chamber.
The SLC wire scanners provide beam cmittance dr.ta that
is
reliable
enough to have already yielded new insights into the
performance of the SLC. By late 1991, 32 scanners will be in
use (figure 4). Future linear colliders will have a tighter
emittance budget and will require improved resolution
a b c
Emittancc measurement method
Figure 3: Comparison of emiltance measurements made
with: multi wire (a) and quadmpolc scans
(b-d)
using different
quadrupole magnet / wire scanner combinations. The results
are in good agreement.
scanners. Piezo-elcctric motors, with their very small step size
and ultra-high vacuum compatibility, may prove to be an
appropriate technology for future wire scanners. Two such
scanners are installed for use at die SLC IP.
ACKNOWLEDGEMENTS
We would like to acknowledge the efforts of the linac
group for help in commissioning the scanners, the mechanical
design and engineering group for mechanical design, A.
Tilghman for controls, and C. Field, K. Bouldin and C. Young
for PMT testing and data analysis.
Special
Purpose
Figure
4:
SLC schematic showing the locations of wire scanners. Wire scanners are used for cmittance (e), energy spread (AE/E)
and special purpose functions in the final focus.
1
R. Jung and RJ. Colchester, 'Development of Beam Profile
and Fast Position Monitors for the LEP Injector Linacs', IEEE
Trans.
Nucl. Sci. NS32-5:1917,1985.
2
R.I. Cutler et. al., "Performance of Wire Scanner Beam
Profile Monitors to Determine the Emittance and Position of
High Power CW Electron Beams of the NBS-Los Alamos
Racetrack Microtron', Proceedings of the 1987 IEEE Particle
Accelerator Conference,
p.
625,1987.
3
K.D. Jacobs el.al., 'The Beam Profile Measurement System
at the Bates Linac', Proceedings of the 1989 IEEE Particle
Accelerator Conference,
p.
1523,1989.
4
L. Sanchez-Chopitea et. al., 'Correlation Plot Facility in the
SLC Control System', proceedings of this conference.
5
R. Fulton et. al., ' A High Resolution Wire Scanner for
Micron Size Profile Measurements at the SIX', Nucl. Instr.
Meth. A274:37,1989.
6
D.
McCormick, "Fast Ion Chambers for SLC, proceedings of
this conference.
7
M.C.
Ross ct. al.,'Automated Emiltance Measurements in
the SLC, Proceedings of the 1987 IEEE Particle Accelerator
Conference, p. 725,1987.
Citations
More filters
Journal ArticleDOI
TL;DR: In this article, a 2 to 4 nm SASE FEL was used for X-ray microscopy in the water window around 2.5 kA, while reducing the bunch length to the subpicosecond range.
Abstract: We report the results of preliminary studies of a 2 to 4 nm SASE FEL, using a photoinjector to produce the electron beam, and the SLAC linac to accelerate it to an energy up to 10 GeV. Longitudinal bunch compression is used to increase ten fold the peak current to 2.5 kA, while reducing the bunch length to the subpicosecond range. The saturated output power is in the multi-gigawatt range, producing about 1014 coherent photons within a bandwidth of about 0.2% rms, in a pulse of several millijoules. At 120 Hz repetition rate the average power is about 1 W. The system is optimized for X-ray microscopy in the water window around 2 to 4 nm, and will permit imaging a biological sample in a single subpicosecond pulse.

38 citations

DOI
01 Dec 1992
TL;DR: This introductory course aims at a reasonably complete coverage of beam diagnostic devices used in linear and circular accelerators and in primary beam lines.
Abstract: This introductory course aims at a reasonably complete coverage of beam diagnostic devices used in linear and circular accelerators and in primary beam lines. The weight is on the concepts and the indication of variants, while for technical details the reader is referred to the literature. The present updated version replaces those from previous General Accelerator Physics Courses.

29 citations

Journal ArticleDOI
TL;DR: In this paper, beam size estimates made using beam-beam deflections are used for optimization of the Stanford Linear Collider (SLC) electron-positron beam sizes for 1996 operations are 2.1 × 0.6 μm at 4.0 × 1010 particles per pulse.
Abstract: Beam size estimates made using beam-beam deflections are used for optimization of the Stanford Linear Collider (SLC) electron-positron beam sizes. Typical beam sizes and intensities expected for 1996 operations are 2.1 × 0.6 μm (x, y) at 4.0 × 1010 particles per pulse. Conventional profile monitors, such as scanning wires, fail at charge densities well below this. The laser-based profile monitor uses a finely-focused 350-nm wavelength tripled YLF laser pulse that traverses the particle beam path about 29 cm away from the e+/e− IP. Compton scattered photons and degraded e+/e− are detected as the beam is steered across the laser pulse. The laser pulse has a transverse size of 380 nm and a Rayleigh range of about 5 μm.

22 citations

Proceedings ArticleDOI
P. Emma1
01 May 1995
TL;DR: The Stanford Linear Collider (SLC) as mentioned in this paper is the only high-energy e/sup +/e/sup -/ linear collider in the world that produces high intensity, submicron sized, polarized beams at a single interaction point.
Abstract: The Stanford Linear Collider (SLC) is the first and only high-energy e/sup +/e/sup -/ linear collider in the world. Its most remarkable features are high intensity, submicron sized, polarized (e/sup -/) beams at a single interaction point. The main challenges posed by these unique characteristics include machine-wide emittance preservation, consistent high intensity operation, polarized electron production and transport, and the achievement of a high degree of beam stability on all time scales. In addition to serving as an important machine for the study of Z/sup 0/ boson production and decay using polarized beams, the SLC is also an indispensable source of hands-on experience for future linear colliders. Each new year of operation has been highlighted with a marked improvement in performance. The most significant improvements for the 1994-95 run include new low impedance vacuum chambers for the damping rings, an upgrade to the optics and diagnostics of the final focus systems, and a higher degree of polarization from the electron source. As a result, the average luminosity has nearly doubled over the previous year with peaks approaching 10/sup 30/ cm/sup -2/ s/sup -1/ and an 80% electron polarization at the interaction point. These developments as well as the remaining identifiable performance limitations will be discussed.

18 citations

Posted Content
N. Phinney1
TL;DR: The Stanford Linear Collider (SLC) was the first prototype of a new type of accelerator, the electron-positron linear collider as mentioned in this paper, and many years of dedicated effort were required to understand the physics of this new technology and to develop the techniques for maximizing performance.
Abstract: The Stanford Linear Collider (SLC) was the first prototype of a new type of accelerator, the electron-positron linear collider. Many years of dedicated effort were required to understand the physics of this new technology and to develop the techniques for maximizing performance. Key issues were emittance dilution, stability, final beam optimization and background control. Precision, non-invasive diagnostics were required to measure and monitor the beams throughout the machine. Beam-based feedback systems were needed to stabilize energy, trajectory, intensity and the final beam size at the interaction point. A variety of new tuning techniques were developed to correct for residual optical or alignment errors. The final focus system underwent a series of refinements in order to deliver sub-micron size beams. It also took many iterations to understand the sources of backgrounds and develop the methods to control them. The benefit from this accumulated experience was seen in the performance of the SLC during its final run in 1997-98. The luminosity increased by a factor of three to 3*10**30 and the 350,000 Z data sample delivered was nearly double that from all previous runs combined.

16 citations

References
More filters
Journal ArticleDOI
TL;DR: In this article, a beam profile is obtained by scanning a fiber across the beam in steps as small as 1 μm, and recording the secondary emission signal at each step, using a charge sensitive amplifier.
Abstract: Fine conductive fibers have been used to measure transverse beam dimensions of a few microns at the Stanford Linear Collider (SLC). The beam profile is obtained by scanning a fiber across the beam in steps as small as 1 μm, and recording the secondary emission signal at each step, using a charge sensitive amplifier. We first outline the mechanical construction and the analogue electronics of the wire scanner. We then describe its performance in test beams and in actual operation. The article closes with a brief discussion of performance limitations of such a beam profile monitor.

32 citations

Proceedings Article
01 Mar 1987
TL;DR: In this paper, the emittance of the SLC beam is determined from measurements of the beam width on a profile monitor as a quadrupole field is varied using an automated system.
Abstract: The emittance of the SLC beam is determined from measurements of the beam width on a profile monitor as a quadrupole field is varied An automated system has been developed to allow this to be done rapidly and accurately The image on a fluorescent screen profile monitor (resolution about 20 pm) is read out through an electronic interface and digitized by a transient recorder A high level software package has been developed to set up the hardware for the measurements, acquire data, fit the beam width, and calculate the emittance

32 citations

01 Jan 1991
TL;DR: The Correlation Plot facility as discussed by the authors is a powerful interactive tool for data acquisition and analysis throughout the SLC, allowing the user to perform a range of operations or machine physics experiments without the need for any specialized analysis software.
Abstract: The Correlation Plot facility is a powerful interactive tool foi data acquisition and analysis throughout the SLC. This generalized interface slows the user to perform a range of operations or machine physics experiments without the need for any specialized analysis software. The user may step one or more independent parameters, such as magnet or feedback setpoints, while measuring or calculating up to 160 other parameters. Measured vapiables include all analog signals available to the control system, as well as calculated parameters such as beam size, luminosity, or emittance. Various fitting algorithms and display options are provided. A software-callable interface has been provided so that a host of applications can call this package for analysis and display. Such applications regularly phase klystrons, measure emittance and dispersion, minimize beam size, and maintain beam collisions at the interaction point.

11 citations

Proceedings ArticleDOI
06 May 1991
TL;DR: Applications based on this facility are used to phase klystrons, measure emittance and dispersion, minimize beam size at the interaction point and maintain beam collisions.
Abstract: The correlation plot facility is a powerful interactive tool for data acquisition and analysis throughout the SLC (Stanford Linear Collider). A generalized interface allows the user to perform a wide variety of machine physics experiments. The user may step one or two independent parameters such as magnet or feedback setpoints while measuring or calculating up to 160 others. Measured variables include all analog signals available to the control system as well as a variety of derived parameters such as beam size or emittance. Various fitting algorithms and display options are provided for data analysis. A software-callable interface is also provided. Applications based on this facility are used to phase klystrons, measure emittance and dispersion, minimize beam size at the interaction point and maintain beam collisions. >

10 citations

Proceedings ArticleDOI
06 May 1991
TL;DR: In this paper, the authors describe the construction and instrumentation of very fast ion chambers with pulse duration of less than 60 ns, referred to as the Panofsky's long ion chamber, which is the primary diagnostic used to locate losses in all the SLC (Stanford Linear Collider) transport lines.
Abstract: Construction and instrumentation of very fast ion chambers with pulse duration of less than 60 ns are detailed. A long ion chamber referred to as the PLIC (Panofsky's long ion chamber) is the primary diagnostic used to locate losses in all the SLC (Stanford Linear Collider) transport lines. Accurately locating beam loss with the use of fudicial cables and coaxial switches is discussed. The design, the pulse length, the performance and operation are considered. >

7 citations