scispace - formally typeset
Journal ArticleDOI

Absorbed radiation dose in mammography.

G R Hammerstein, +5 more
- 01 Feb 1979 - 
- Vol. 130, Iss: 2, pp 485-491
Reads0
Chats0
TLDR
Analysis of data from the ACS/NCI Screening Centers shows current average midbreast doses to be 25 times lower (film/screen) and 3 times lower than the 2 rads previously estimated.
Abstract
Radiation dose from mammographic techniques was determined as a function of surface exposure, beam quality, and depth. Relative exposure vs. depth was measured in tissue-substitute materials by thermoluminescent dosimetry. The f-factors were calculated from elemental compositions of mastectomy specimens. Dose at depth depends on beam quality as well as exposure and tissue composition. Analysis of data from the ACS/NCI Screening Centers shows current average midbreast doses to be 25 times lower (film/screen) and 3 times lower (Xerox) than the 2 rads previously estimated. Quantitative risk indicators other than midbreast dose are also discussed.

read more

Citations
More filters

sources and effects of ionizing radiation

TL;DR: This annex is aimed at providing a sound basis for conclusions regarding the number of significant radiation accidents that have occurred, the corresponding levels of radiation exposures and numbers of deaths and injuries, and the general trends for various practices, in the context of the Committee's overall evaluations of the levels and effects of exposure to ionizing radiation.
Journal ArticleDOI

IARC Handbooks of Cancer Prevention

TL;DR: The International Agency for Research on Cancer conducts a programme of research concentrating particularly on the epidemiology of cancer and the study of potential carcinogens in the human environment.
Journal ArticleDOI

The composition of body tissues

TL;DR: A reassessment of the composition data given by ICRP (1975) has been completed, augmented with results from appropriate recent studies, and the compositions presented apply to healthy, adult humans, except for some new material on the bones of children.
Journal ArticleDOI

Physics of Radiology

F. W. Spiers
- 01 Dec 1968 - 
References
More filters
Book

The physics of radiology

H. E. Johns
Journal ArticleDOI

Physics of Radiology

F. W. Spiers
- 01 Dec 1968 - 
Journal ArticleDOI

Photon mass attenuation and mass energy-absorption coefficients for H, C, N, O, Ar, and seven mixtures from 0. 1 keV to 20 MeV

J. H. Hubbell
- 01 Apr 1977 - 
TL;DR: From this cross-section data base, mass energy-absorption coefficients pen/p, taking into account bremsstrahlung losses and positron annihilation in flight, Compton-scattered and fluorescence photons, and mass-attenuation coefficients p/p are tabulated for the above elements and for the mixtures.
Journal ArticleDOI

Epoxy resin based tissue substitutes

TL;DR: A series of epoxy resin based tissue substitutes, simulating adipose tissue, bone, breast, lung, muscle and skin, and the versatility of the range of substitutes is discussed in terms of both investigated and projected applications in experimental radiation studies.
Journal ArticleDOI

Mammography: a contrary view.

TL;DR: There is good evidence that annual history, physical examination, and mammography can reduce short-term and midrange breast cancer mortality by about one third and promotion of mammography as a general public health measure is premature.
Related Papers (5)