scispace - formally typeset
Journal ArticleDOI

Automatic Calibration System for a Laser Power Meter

Takemi Inoue, +2 more
- 01 Sep 1985 - 
- Vol. 34, Iss: 3, pp 431-436
TLDR
In this paper, a microprocessor-based calibration system for a CW laser power meter has been developed, which consists of an isothermal calorimeter with a thermopile unit, an electrical, and an optical subsystem.
Abstract
A microprocessor-based calibration system for a CW laser power meter has been developed. The system consists of an isothermal calorimeter with a thermopile unit, an electrical, and an optical subsystem. Temperature control of the thermopile unit is achieved through a digital technique. The calibration of a power meter is performed using the effective efficiency of the unit as measured by a calorimeter. From error analyses, an overall accuracy has been obtained as ±1 percent in root mean square for a power level of 1 ~ 200 mW.

read more

Citations
More filters
Journal ArticleDOI

Highly sensitive calorimeter for microwatt-level laser power measurements

TL;DR: In this article, a sensitive calorimeter using an isothermal temperature control technique was developed for an accurate measurement of microwatt-level laser power in the visible and infrared spectral region.
Journal ArticleDOI

Calorimetric measurement of absolute terahertz power at the sub-microwatt level.

TL;DR: A calorimeter based on a DC substitution method using an isothermal temperature-control technique for determining absolute terahertz power is developed and is expected to demonstrate the traceability of THz power at a submicrowatt level.
Journal ArticleDOI

A PC-based adaptive software for automatic calibration of power transducers

TL;DR: This paper presents the design and implementation of an adaptive PC-based software on a computer-aided automatic calibration system built for the Instrumentation and Calibration Department, Taiwan Power Company, to calibrate power transducers regularly.
Journal ArticleDOI

A double calorimeter for 10-W level laser power measurements

Makoto Endo, +1 more
TL;DR: In this paper, a double configuration consisting of an absorbing unit made of a disk absorber and an isothermal type calorimeter was fabricated for precise measurement of 10-W level laser power.
References
More filters
Journal ArticleDOI

Calorimetric measurements of laser energy and power

TL;DR: The use of calorimeters for measuring laser output energies and powers is broadly reviewed as mentioned in this paper, including disc, cone, hollow-sphere, bolometer, volume absorption, and partial absorption types.
Journal ArticleDOI

A Laser Microcalorimeter

TL;DR: In this article, a double-calorimeter was developed at the Electrotechnical Laboratory for the measurement of the CW power of the gas laser (He-Ne) with high stability (± 0.05 percent).
Journal ArticleDOI

Advanced Control Method for Calorimetric Power Measurement

TL;DR: In this article, an advanced temperature control system has been developed for calorimetric method of electromagnetic wave power measurement, which can be reduced by repeatedly substituting the controlled variable through a DA converter and an adder.