Journal ArticleDOI
Intra-rater reliability of lower leg subcutaneous skin fold thickness derived from hand-held a-mode ultrasound
Brian T. McGirr,Jake L. Martin,Chris E. Colborn,Alex C.S. Shefflette,Steve Ryan Soltysiak,Elisabeth J. Dichiara,John F. Caruso,John F. Caruso +7 more
TLDR
Despite the relatively long period between measurements, the hand-held ultrasound device exhibited a high degree of intra-rater reliability, and ultrasound measurements may be a useful tool to quantify skin fold thickness.Abstract:
BACKGROUND: Ultrasound is an important tool to diagnose many clinical conditions. Yet hand-held devices may be prone to more data variability in part from the greater likelihood of human error. OBJECTIVE: Quantify intra-rater reliability of subcutaneous skin fold thickness from a hand-held ultrasound device. PARTICIPANTS: College-age subjects (18 men, 14 women) submitted to two sets of ultrasound subcutaneous skin fold measurements spaced (mean + sem) 10.6 + 2.2 days apart. Per measurement, they stood relaxed as ultrasound measured the subcutaneous skin fold thickness that covered their left leg’s calf muscle group. Measurements occurred with a hand-held device (BodyMetrix Pro System BX2000; Livermore, CA) used in accordance with the manufacturer’s guidelines. Four subcutaneous measurements were made 90∘ apart (anterior, medial, posterior, lateral) at the portion of the left calf with the largest circumference. To assess intra-rater reliability, we used intraclass correlation coefficients, limits of agreement, coefficient of variation and the smallest real difference. RESULTS: Intra-rater reliability was high for most of our statistical tests. CONCLUSION: Despite the relatively long period between measurements, our hand-held ultrasound device exhibited a high degree of intra-rater reliability. Given our results, ultrasound measurements may be a useful tool to quantify skin fold thickness.read more
References
More filters
Journal ArticleDOI
Measures of reliability in sports medicine and science.
TL;DR: A wider understanding of reliability and adoption of the typical error as the standard measure of reliability would improve the assessment of tests and equipment in the authors' disciplines.
Journal ArticleDOI
Reliability of gait performance tests in men and women with hemiparesis after stroke
Ulla-Britt Flansbjer,Anna Maria Holmbäck,David Downham,Carolynn Patten,Carolynn Patten,Jan Lexell,Jan Lexell +6 more
TL;DR: These commonly used gait performance tests are highly reliable and can be recommended to evaluate improvements in various aspects of gaitperformance in individuals with chronic mild to moderate hemiparesis after stroke.
Journal ArticleDOI
Smallest real difference, a link between reproducibility and responsiveness.
Heleen Beckerman,Marij E. Roebroeck,Gustaaf J. Lankhorst,Jules G. Becher,P.D. Bezemer,André L. M. Verbeek +5 more
TL;DR: The responsiveness to change of a health status measurement instrument is closely related to its test-retest reproducibility, and this relationship becomes more evident when the SEM and the SRD are used to quantify reproducible, than when ICC or other correlation coefficients are used.
Journal ArticleDOI
Design and analysis of research on sport performance enhancement.
TL;DR: Assessment of research aimed at measuring performance enhancements that affect success of individual elite athletes in competitive events found that estimates of enhancement of performance in laboratory or field tests in most previous studies may not apply to elite athletesIn competitive events.
Journal ArticleDOI
Skeletal muscle ultrasound
Sigrid Pillen,Nens van Alfen +1 more
TL;DR: With improving resolution and frame rate it has recently become clear that also smaller scale spontaneous muscle activity such as fibrillations can be detected by ultrasound, which opens the way to a broader use of muscle ultrasound in the diagnosis of peripheral nerve and muscle disorders.