Correcting astigmatism with toric intraocular lenses: Effect of posterior corneal astigmatism

TLDR
Corneal astigmatism was overestimated in WTR by all devices and underestimated in ATR by all except the Placido–dual Scheimpflug analyzer.
Abstract
Purpose To evaluate the impact of posterior corneal astigmatism on outcomes with toric intraocular lenses (IOLs). Setting Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA. Design Case series. Methods Corneal astigmatism was measured using 5 devices before and 3 weeks after cataract surgery. Toric IOL alignment was recorded at surgery and at the slitlamp 3 weeks postoperatively. The actual corneal astigmatism was calculated based on refractive astigmatism 3 weeks postoperatively and the effective toric power calculated with the Holladay 2 formula. The prediction error was calculated as the difference between the astigmatism measured by each device and the actual corneal astigmatism. Vector analysis was used in all calculations. Results With the IOLMaster, Lenstar, Atlas, manual keratometer, and Galilei (combined Placido–dual Scheimpflug analyzer), the mean prediction errors (D) were, respectively, 0.59 @ 89.7, 0.48 @ 91.2, 0.51 @ 78.7, 0.62 @ 97.2, and 0.57 @ 93.9 for with-the-rule (WTR) astigmatism (60 to 120 degrees), and 0.17 @ 86.2, 0.23 @ 77.7, 0.23 @ 91.4, 0.41 @ 58.4, and 0.12 @ 7.3 for against-the-rule (ATR) astigmatism (0 to 30 degrees and 150 to 180 degrees). In the WTR eyes, there were significant WTR prediction errors (0.5 to 0.6 diopters [D]) by all devices. In ATR eyes, WTR prediction errors were 0.2 to 0.3 D by all devices except the Placido–dual Scheimpflug analyzer (all P Conclusions Corneal astigmatism was overestimated in WTR by all devices and underestimated in ATR by all except the Placido–dual Scheimpflug analyzer. A new toric IOL nomogram is proposed. Financial Disclosure Drs. Koch, Weikert, and Wang received research support from Ziemer USA, Inc. Dr. Koch has a financial interest with Alcon Laboratories, Inc., Abbott Medical Optics, Inc., Optimedica Corp., and Ziemer USA, Inc. No other author has a financial or proprietary interest in any material or method mentioned.

read more

Citations
More filters
Journal ArticleDOI

Comparison of accuracy of a toric calculator with predicted vs measured posterior corneal astigmatism

TL;DR: Accuracy of residual astigmatism prediction is improved using the Barrett toric calculator with measured PCA rather than predicted PCA.
Journal ArticleDOI

Comparison of Barrett Toric Calculations Using Measured and Predicted Posterior Corneal Astigmatism in Cataract Surgery Patients

TL;DR: B Barrett toric calculations using predicted posterior corneal astigmatism (PCA) and PCA measured using the IOL Master 700 with total keratometry (IOLM) produced comparable residualAstigmatism prediction errors.
Journal ArticleDOI

Total Keratometry Measured With a Swept-Source Optical Biometer Versus Anterior Keratometry: From Planning to Postoperative Results.

TL;DR: In this article , the authors compared the influence of total keratometry (TK) versus anterior Keratometry measured with the swept-source optical biometer IOLMaster 700 (Carl Zeiss Meditec AG) on the planning of toric intraocular lenses (IOLs), and the error in predicted residual astigmatism (PRA).
Journal ArticleDOI

Estudio comparativo de la medición del poder corneal central usando el queratómetro manual, el IOLMaster® y el tomógrafo Sirius®

TL;DR: B Bland et al. as discussed by the authors compare the valores of the queratometria corneal central obtenidos mediante el tomografo de camara de Scheimpflug with anillos de Placido (Sirius ®), el queratometric manual tipo Bausch & Lomb ® (MK1 Reichert ®) and biometro optico IOLMaster ® ; and conocer the repetibilidad intraobservador of estos equipos.
Journal ArticleDOI

Astigmatism Profile in a Large Series of Brazilian Patients.

TL;DR: In this article , the mean magnitude of the anterior, posterior, and total corneal astigmatism was 1.50 ± 1.11, 0.34 ± 0.15, and 1.29± 0.98 diopters, respectively.
References
More filters
Journal ArticleDOI

Contribution of posterior corneal astigmatism to total corneal astigmatism.

TL;DR: Selecting toric intraocular lenses based on anterior corneal measurements could lead to overcor correction in eyes that have with‐the‐rule astigmatism and undercorrection in eyes which have against‐the-ruleAstigmatism.
Journal ArticleDOI

The shape of the anterior and posterior surface of the aging human cornea.

TL;DR: The shape of the anterior corneal surface provides no definitive basis for knowing the asphericity of the posterior surface and the results show the effective refractive index is 1.329, which is lower than values commonly used.
Journal ArticleDOI

Analysis of aggregate surgically induced refractive change, prediction error, and intraocular astigmatism

TL;DR: Analytical methods for evaluating the results of keratorefractive surgical procedures and emphasize the importance of intraocular astigmatism and the limitations of manual keratometry are demonstrated.
Journal ArticleDOI

Foldable toric intraocular lens for astigmatism correction in cataract patients.

TL;DR: The results indicate that phacoemulsification and posterior chamber AcrySof toric IOL implantation is an effective option to correct preexisting astigmatism in cataract surgery.