Design and Construction of a Pulsed Ultrasonic Air Flowmeter

Reads0
Chats0
TLDR
This respiratory air flowmeter is extremely fast and accurate, with low noise (below 9ml/s), with a wide flow range (bidirectional from 0 to 9 l/s and with a linear frequency response up to 70 Hz).
Abstract
The construction and specific function of a new ultrasonic flowmeter are described. The mean velocity of the respiratory air flow is calculated by measuring the transit times of short ultrasonic pulse-trains simultaneously transmitted upstream and downstream at a 650 Hz rate. The flowmeter system consists of a control unit and a separate flowhead. The former includes the power supplies, a controlling microprocessor, most of the signal processing circuitry, and three analog outputs for flow, volume, and temperature, respectively. The flowhead contains the respiratory tube with a constant circular cross section (length 90 mm, diameter 20 mm, dead space 35 cm3), a fast temperature sensor, two electronic circuits for processing of flow and temperature data, and a sound transmission channel with two capacitive ultrasonic wide-band transducers. This respiratory air flowmeter is extremely fast (response time 1-2 ms) and accurate, with low noise (below 9ml/s), with a wide flow range (bidirectional from 0 to 9 l/s) and with a linear frequency response up to 70 Hz.

read more

Citations
More filters
Journal ArticleDOI

Lung clearance index: Normal values, repeatability, and reproducibility in healthy children and adolescents

TL;DR: Reliability and robustness of equipment, protocol and analysis and the reliability of the MBW technique in general are demonstrated, and low variability of the LCI for both, within and between tests is demonstrated.
Journal ArticleDOI

A novel sensor for routine continuous spirometry of intubated patients.

TL;DR: The D-lite sensor, in a single, lightweight adaptor piece, encloses flow, airway pressure, and sidestream gas measurement, which can be handled by sophisticated computer algorithms at the host-monitor end.
Journal ArticleDOI

Continuous calculation of intratracheal pressure in the presence of pediatric endotracheal tubes.

TL;DR: Ptrach can continuously be monitored in the presence of pediatric ETT by combining ETT coefficients and the flow and airway pressure continuously measured at the proximal end of the ETT.
Journal ArticleDOI

Multiple breath washout with a sidestream ultrasonic flow sensor and mass spectrometry: A comparative study†‡

TL;DR: It is concluded, that this ultrasonic prototype system measures end‐tidal tracer gas concentration accurately and may therefore be a valid tool for MBW beyond early childhood and could be the basis for a commercial device allowing more widespread application of MBW in the near future.
Journal ArticleDOI

A Novel Sidestream Ultrasonic Flow Sensor for Multiple Breath Washout in Children

TL;DR: The aims of this study were to assess the feasibility and clinical validity of measuring lung volume and the LCI using the sidestream ultrasonic flow sensor in children and adolescents with CF in relation to spirometry and plain chest radiographs.
References
More filters
Journal ArticleDOI

Der Pneumotachograph; ein Apparat zur Geschwindigkeitsregistrierung der Atemluft

TL;DR: In this article, a pneumotachograph is genannt, beschrieben, der eine exakte, fortlaufende optische registrierung der Stromgeschwindigkeit der Atemluft gestattet.
Journal ArticleDOI

A technique for measuring frequency response of pressure, volume, and flow transducers

TL;DR: The results emphasize the need to test the response characteristics of any transducer with specific connectors and fittings that are to be used to make the actual measurements of pressure, volume, or flow.
Journal ArticleDOI

Ultrasonic Measurement of Respiratory Flow

TL;DR: This work reviews ultrasonic measurement of fluid flow with emphasis on ultrasonic pneumotachometers (UPTM's) and UPTM's, and presents four principle types: time of flight, sing around, phase shift, and phase-locked frequency shift.
Journal ArticleDOI

Design and Construction of an Ultrasonic Pneumotachometer

TL;DR: A new design which used cylindrical shell transducers to generate ultrasound and measured phase shift as a measure of transit time change due to flow was investigated, which was linear and predictable over a wide dynamic range using a single gas under controlled conditions.
Journal ArticleDOI

A coaxial ultrasonic pneumotachometer.

TL;DR: Empirical measurements of the response function and baseline stability of a prototype transducer are presented, along with detailed correction algorithms for ultrasound velocity and effective transmission path length.
Related Papers (5)