scispace - formally typeset
Open AccessJournal Article

A chronic intracortical electrode array: preliminary results

TLDR
It is concluded that a chronic intracortical stimulating electrode array of this geometry should have the following properties: 1) the penetrating electrodes and supporting substrate must be stiff (to maintain interelectrode spacings upon implantation), and 2) the percutaneous leads must be extremely flexible (to avoid array movement after implantation).
Abstract
Two sets of electrode arrays made of either 25- or 50-microns-diameter Teflon-insulated platinum-iridium wire and Teflon have been developed for chronic intracortical electrical stimulation. Cortical histological studies were performed following acute and chronic implantation in cats. While some neural damage resulted from the implantations of either array configuration, a unique set of problems was associated with each diameter wire. Arrays with 50-microns electrodes and lead wires tended to maintain interelectrode spacing upon implantation, but the percutaneous leads retained residual stress which made array implantation difficult. Arrays with 25-microns electrodes and lead wires suffered from changes in interelectrode spacing upon implantation, but were much easier to manipulate during surgery. Both array configurations demonstrated some movement after implantation. It is concluded that a chronic intracortical stimulating electrode array of this geometry should have the following properties: 1) the penetrating electrodes and supporting substrate must be stiff (to maintain interelectrode spacings upon implantation), and 2) the percutaneous leads must be extremely flexible (to avoid array movement after implantation).

read more

Citations
More filters
Journal ArticleDOI

A silicon-based, three-dimensional neural interface: manufacturing processes for an intracortical electrode array

TL;DR: A method is described for the manufacture of a three-dimensional electrode array geometry for chronic intracortical stimulation that consists of a 4.2*4.
Journal ArticleDOI

Wireless implantable microsystems: high-density electronic interfaces to the nervous system

TL;DR: This paper describes the development of a high-density electronic interface to the central nervous system that permits the long-term monitoring of neural activity in vivo as well as the insertion of electronic signals into neural networks at the cellular level.
Patent

Biological interface system

TL;DR: In this paper, a system and method for an improved biological interface system that processes multicellular signals of a patient and controls one or more devices is described, which may include improved communication, self-diagnostics, and surgical insertion tools.
Journal ArticleDOI

Chronically implanted intrafascicular recording electrodes

TL;DR: A newly designed intrafascicular electrode for chronic neural recording was studied by implanting 12 electrodes in the radial nerves of 6 cats for 6 months and it was suggested that this type of electrode provides information that is appropriate for feedback control in functional electrical stimulation (FES) systems.
Patent

Closed loop brain machine interface

TL;DR: In this article, a closed loop brain-machine interface is proposed to translate one or more neural signals into a movement, or a series of movements, performed by a machine and provide sensory feedback to the subject.
Related Papers (5)