Effects of Pulsatile Control Algorithms for Diagonal Pump on Hemodynamic Performance and Hemolysis.

TLDR
Comparing hemodynamic performances under different pulsatile control algorithms between Medos DeltaStream DP3 and i-cor diagonal pumps in simulated pediatric and adult ECLS systems found high pulsatile amplitudes delivered higher levels of hemodynamic energy to the patient, but high rotational speeds increased the risk of hemolysis.
Abstract
The objective of this study is to compare hemodynamic performances under different pulsatile control algorithms between Medos DeltaStream DP3 and i-cor diagonal pumps in simulated pediatric and adult ECLS systems. An additional pilot study was designed to test hemolysis using two pumps during 12h-ECLS. The experimental circuit consisted of parallel combined pediatric and adult ECLS circuits using an i-cor pump head and either an i-cor console or Medos DeltaStream MDC console, a Medos Hilite 2400 LT oxygenator for the pediatric ECLS circuit, and a Medos Hilite 7000 LT oxygenator for the adult ECLS circuit. The circuit was primed with lactated Ringer's solution and human packed red blood cells (hematocrit 40%). Trials were conducted at various flow rates (pediatric circuit: 0.5 and 1L/min; adult circuit: 2 and 4L/min) under nonpulsatile and pulsatile modes (pulsatile amplitude: 1000-5000rpm [1000 rpm increments] for i-cor pump, 500-2500rpm [500 rpm increments] for Medos pump) at 36°C. In an additional protocol, fresh whole blood was used to test hemolysis under nonpulsatile and pulsatile modes using the two pump systems in adult ECLS circuits. Under pulsatile mode, energy equivalent pressures (EEP) were always greater than mean pressures for the two systems. Total hemodynamic energy (THE) and surplus hemodynamic energy (SHE) levels delivered to the patient increased with increasing pulsatile amplitude and decreased with increasing flow rate. The i-cor pump outperformed at low flow rates, but the Medos pump performed superiorly at high flow rates. There was no significant difference between two pumps in percentage of THE loss. The plasma free hemoglobin level was always higher in the Medos DP3 pulsatile group at 4 L/min compared to others. Pulsatile control algorithms of Medos and i-cor consoles had great effects on pulsatility. Although high pulsatile amplitudes delivered higher levels of hemodynamic energy to the patient, the high rotational speeds increased the risk of hemolysis. Use of proper pulsatile amplitude settings and intermittent pulsatile mode are suggested to achieve better pulsatility and decrease the risk of hemolysis. Further optimized pulsatile control algorithms are needed.

read more

Citations
More filters
Journal ArticleDOI

Evaluation of centrifugal blood pumps in term of hemodynamic performance using simulated neonatal and pediatric ECMO circuits.

TL;DR: The RotaFlow centrifugal pump had a significantly better hemodynamic performance when compared to the PediMag and CentriMag blood pumps at identical experimental conditions in simulated neonatal and pediatric ECMO settings.
Journal ArticleDOI

Evaluation of Two Femoral Arterial Cannulae With Conventional Non-Pulsatile and Alternative Pulsatile Flow in a Simulated Adult ECLS Circuit.

TL;DR: This novel i-cor ECLS system can provide non-pulsatile and ECG-synchronized pulsatile flow without significantly increasing the cannula pressure drop and hemodynamic energy loss and should be evaluated before use in clinical practice.
Journal ArticleDOI

Artificial Organs 2019: A year in review

TL;DR: Artificial Organs provides a brief reflection of the research and progress in artificial organs intended to advance and better human life while providing insight for continued application of these technologies and methods.
Journal ArticleDOI

Centrifugal Pump Generates Superior Hemodynamic Performance Compared to a new Diagonal Blood Pump in Neonatal and Pediatric ECMO Circuits

TL;DR: The RotaFlow centrifugal pump demonstrated superior hemodynamic performance when compared to the DP3-i-cor diagonal pump in simulated neonatal and pediatric ECMO circuits.
Journal ArticleDOI

Comparison of three control strategies for axial blood pump

TL;DR: The blood pump is widely used to prolong the life of end-stage HF patients and the shortage of donated hearts means that the pump gen...
References
More filters
Journal ArticleDOI

Energy equivalent pressure.

TL;DR: In this article, the authors describe differences in steady and pulsatile blood flow in terms of an energy equivalent pressure, which is obtained by calculation from phasic flow and pressure measurements.
Journal ArticleDOI

Pulsatile Versus Nonpulsatile Flow During Cardiopulmonary Bypass: Microcirculatory and Systemic Effects

TL;DR: It is suggested the implementation of pulsatile flow can better optimize microvascular perfusion, and may lead to improved patient outcomes in high-risk cardiac surgical procedures requiring prolonged CPB time.
Journal ArticleDOI

Pulsatile flow during cardiopulmonary bypass preserves postoperative microcirculatory perfusion irrespective of systemic hemodynamics

TL;DR: Pulsatile cardiopulmonary bypass preserves microcirculatory perfusion throughout the early postoperative period, irrespective of systemic hemodynamics, paralleled by an increase in oxygen consumption during pulsatile flow.
Journal ArticleDOI

Does pulsatility matter in the era of continuous-flow blood pumps?

TL;DR: It is suggested that lack of pulsatility may be one factor that has limited the success of continuous flow LVADs and research needs to focus on methods to generate pulsatility either by the native heart or by various speed modulation algorithms.
Journal ArticleDOI

Electrocardiogram-synchronized rotational speed change mode in rotary pumps could improve pulsatility.

TL;DR: The authors' newly developed pulsatile mode of continuous-flow LVADs can produce pulsatility comparable to physiological pulsatile flow.