Cardiac Output Monitoring by Impedance Cardiography During Treadmill Exercise

Reads0
Chats0
TLDR
An ensemble-averaging technique for eliminating motion artifacts was adapted and an array consisting of four spot ECG electrodes for impedance cardiogram (ICG) monitoring and compared it to conventional encircling band electrodes was applied.
Abstract
We have modified impedance cardiography for monitoring cardiac output during stress tests. We adapted an ensemble-averaging technique for eliminating motion artifacts. We applied an array consisting of four spot ECG electrodes for impedance cardiogram (ICG) monitoring and compared it to conventional encircling band electrodes. We tested ten normal adults, compared the cardiac output obtained by our ICG monitoring system to that simultaneously obtained by the carbon dioxide (CO2) rebreathing method at rest and during three levels of treadmill exercise. The results show that the correlation coefficient between the spot electrode ICG and the CO2 rebreathing method is r = 0.90, and between the band electrode ICG and the CO2 rebreathing method is r = 0.96. If we use the peak-to-valley height of dz/dt instead of the peak height of dZ/dt in computing cardiac output, the correlation coefficient between the spot electrode ICG and the CO2 rebreathing method can be improved to r = 0.95.

read more

Citations
More filters
Journal ArticleDOI

Methodological Guidelines for Impedance Cardiography

TL;DR: The various aspects of impedance methodology are discussed, and alternative approaches described, with the objective of providing an informed basis for choosing among these methodological alternatives.
Journal ArticleDOI

Ambulatory monitoring of the impedance cardiogram

TL;DR: It is concluded that the VU-AMD is a valid device for the measurement of systolic time intervals in real-life situations, but its applicability for absolute stroke volume and cardiac output determination remains to be established.
Journal ArticleDOI

Gender differences in blood pressure control during a variety of behavioral stressors.

TL;DR: A post hoc analysis indicated an apparent association between oral contraceptive use and higher cardiovascular reactivity among the females tested, and a consequence of the decision to test all women during days 10 to 14 of the menstrual cycle when reactivity in women not using oral contraceptives may be suppressed.
Journal ArticleDOI

Motion Artifact from Spot and Band Electrodes During Impedance Cardiography

TL;DR: A new four-spot electrode array is proposed and replaced the usual encircling band electrode array with it for impedance cardiography for exercising subjects and the signal-to-noise ratio (SNR) from the spot electrode array was higher than that from a typical band electrode arrays at rest and during four levels of exercise on a treadmill.
Journal ArticleDOI

A software package for non-invasive, real-time beat-to-beat monitoring of stroke volume, blood pressure, total peripheral resistance and for assessment of autonomic function

TL;DR: The goal of the present study was to develop and evaluate algorithms for non-invasive, real-time, beat-to-beat monitoring of stroke index (SI), blood pressure (BP) and total peripheral resistance index (TPRI) which has a menu-driven interface, suitable for routine use by unskilled staff.
References
More filters
Journal ArticleDOI

A Real-Time QRS Detection Algorithm

TL;DR: A real-time algorithm that reliably recognizes QRS complexes based upon digital analyses of slope, amplitude, and width of ECG signals and automatically adjusts thresholds and parameters periodically to adapt to such ECG changes as QRS morphology and heart rate.
Journal Article

Development and evaluation of an impedance cardiac output system

TL;DR: Impedance cardiac output values simultaneously compared with dye dilution techniques under rest and exercise conditions are compared to show the difference in output values between the two techniques under different exercise conditions.
Journal ArticleDOI

Motion Artifact from Spot and Band Electrodes During Impedance Cardiography

TL;DR: A new four-spot electrode array is proposed and replaced the usual encircling band electrode array with it for impedance cardiography for exercising subjects and the signal-to-noise ratio (SNR) from the spot electrode array was higher than that from a typical band electrode arrays at rest and during four levels of exercise on a treadmill.
Journal ArticleDOI

The specific resistance of blood at body temperature

TL;DR: The specific resistance of human, canine, bovine, and equine blood with a haematocrit range extending from 0 to 70% and at 37°C was measured at 25 kHz.
Related Papers (5)