scispace - formally typeset
Journal ArticleDOI

In vivo measurement of total knee replacement wear.

Reads0
Chats0
TLDR
A Roentgen stereophotogrammetric analysis system that can measure penetration of the metallic femoral component into the polyethylene of the tibia and has the potential for predicting long-term wear problems with new designs of TKR and new materials within 2 years is developed.
Abstract
Polyethylene wear is one of the most important causes of failure of total knee replacements (TKRs). Currently, wear can only be accurately measured by retrieval studies. There is a need for a method to measure wear accurately in vivo. We have developed a Roentgen stereophotogrammetric analysis (RSA) system that can measure penetration of the metallic femoral component into the polyethylene of the tibia. We have used this system to study six AGC TKRs at 6 years postoperatively and six control AGC TKRs at 2 weeks postoperatively. The mean difference between the RSA measured bearing thickness and the manufacturer's quoted values for the control group was −0.03 mm (S.D. 0.17). The average linear penetration in the study group was 0.8 mm (S.D. 0.46). This was significantly (P<0.0001) different from the control group. The average linear penetration rate was 0.13 mm per year (S.D. 0.08). We would expect the penetration to deepen with time. In young active patients, this could be a cause for concern, particularly with a thin bearing. The current system is accurate enough to measure wear at 5 years post TKR. It has the potential for predicting long-term wear problems with new designs of TKR and new materials within 2 years.

read more

Citations
More filters
Journal ArticleDOI

Guidelines for standardization of radiostereometry (RSA) of implants

TL;DR: 6 research centers have agreed upon standards for terminology, description and use of RSA arrangement including radiographic set-up and techniques, which will form the basis of a detailed standardization protocol under supervision of ISO and the European Standards Working Group on Joint Replacement Implants.
Journal ArticleDOI

ESB clinical biomechanics award 2008: Complete data of total knee replacement loading for level walking and stair climbing measured in vivo with a follow-up of 6–10 months

TL;DR: Using the data obtained, mechanical simulators can be programmed according to realistic load profiles and musculo-skeletal models can be validated, which until now often lacked the ability to predict properly the non-sagittal load values, e.g. varus-valgus and internal-external moments.
Journal ArticleDOI

Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty.

TL;DR: The wear-related loss of thickness in gamma-irradiated-in-air polyethylene bearings from unicondylar and total knee replacements implanted in osteoarthritic knees with varus deformity is influenced mainly by the shelf age of the polyethylenes, the age of a patient, and the postoperative angulation of the knee in the coronal plane.
Journal ArticleDOI

Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty

TL;DR: In this paper, the authors measured the in vivo ten-year linear wear of the Oxford unicompartmental knee arthroplasty, using a roentgenstereophotogrammetric technique.

Ten year in-vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty.

TL;DR: The results from this in vivo study confirm that the Oxford unicompartmental knee arthroplasty has low ten-year linear wear in clinical practice, which may offer the device a survival advantage in the long term.
References
More filters
Journal ArticleDOI

Macrophages stimulate bone resorption when they phagocytose particles

TL;DR: It was concluded that implant loosening may result from bone resorption stimulated by mediators released by macrophages that have phagocytosed particles of bone cement or polyethylene.
Journal ArticleDOI

Wear of Polyethylene Acetabular Components in Total Hip Arthroplasty. An Analysis of One Hundred and Twenty-eight Components Retrieved at Autopsy or Revision Operations*

TL;DR: Wear appeared to occur mostly at the surface of the components and to be due to large-strain plastic deformation and orientation of the surface layers into fibrils that subsequently ruptured during multidirectional motion.
Journal ArticleDOI

Wear and osteolysis in total joint replacements

TL;DR: The basic strategy for addressing the problem of osteolysis is to reduce the "number" of accumulated wear particles in the interface tissues, which could be achieved either by improving the materials or the geometry of the articulating counterface.
Journal ArticleDOI

Wear of congruent meniscal bearings in unicompartmental knee arthroplasty: A RETRIEVAL STUDY OF 16 SPECIMENS

TL;DR: In unicompartmental knee replacement, careful implantation of fully congruous meniscal bearings can avoid failure due to polyethylene wear.
Journal ArticleDOI

A knee simulating machine for performance evaluation of total knee replacements

TL;DR: It was concluded that a simulating machine should be based on the input of forces and moments, rather than on displacements and rotations, in order to provide data of kinematics and wear.