scispace - formally typeset
Proceedings ArticleDOI

Initial results from a forward-viewing ring-annular ultrasound array for intravascular imaging

Reads0
Chats0
TLDR
In this paper, a forward viewing ring-annular array for intravascular ultrasound imaging has been proposed, which is appropriate for over-the-wire delivery but limits the forward-looking aperture to a ring-ANNular array.
Abstract
In this paper, we present some initial results from our proposed forward viewing ring-annular array for intravascular ultrasound imaging. We have investigated array design and image synthesis allowing simultaneous sideward imaging at 20 MHz and forward imaging at 10 MHz from the same array. This design is appropriate for over-the-wire delivery but limits the forward-looking aperture to a ring-annular array. In particular, a 1200 /spl mu/m diameter ring-annular array with 64 elements has been constructed. The measured forward- viewing radiation pattern of an individual element indicates an acceptance angle range of over 90/spl deg/. Currently, there are significant sensitivity and center frequency variations among the elements. Nevertheless, images of point targets are obtained as a proof-of-concept. Future improvements of the array will be discussed.

read more

Citations
More filters
Proceedings ArticleDOI

Advances in two dimensional arrays for real time 3D intravascular ultrasound

TL;DR: Two forward-viewing transducer intended to guide the placement and retrieval of devices in the peripheral vessels are developed and real time 3D images have been obtained from both phantoms and actual hearts.
Proceedings ArticleDOI

Forward-looking IVUS imaging using an annular-ring CMUT array

TL;DR: In this article, a 64-element, 1.15mm diameter annular-ring CMUT array was characterized and used for forward-looking IVUS imaging tests using low temperature processes suitable for CMOS electronics integration on a single chip.
References
More filters
Journal ArticleDOI

Synthetic phased arrays for intraluminal imaging of coronary arteries

TL;DR: In this article, a 64-element, high efficiency, ceramic piezoelectric array transducer operating at 20 MHz has been constructed for ultrasonic intraluminal imaging.
Journal ArticleDOI

Optimizing the beam pattern of a forward-viewing ring-annular ultrasound array for intravascular imaging

TL;DR: A practical array geometry and signal processing architecture for a forward-viewing IVUS system that can achieve side-lobes under -40 dB for on-axis situations and under -30 dB for steering to the edge of a 60/spl deg/ cone.
Journal ArticleDOI

Arterial imaging with a new forward-viewing intravascular ultrasound catheter, I. Initial studies.

TL;DR: Preliminary data suggest that a forward- viewing IVUS catheter is feasible, accurate, and useful for evaluation of arterial geometry distal to the catheter tip.
Journal ArticleDOI

Arterial imaging with a new forward-viewing intravascular ultrasound catheter, II. Three-dimensional reconstruction and display of data.

TL;DR: Preliminary data demonstrate the feasibility of 3DR of forward-viewing IVUS data and allows rapid, detailed analysis of diseased arterial segments previously unavailable with standard IVUS and may permit better targeting of interventional techniques.
Journal ArticleDOI

Coded EXcitation with spectrum inversion (CEXSI) for ultrasound array imaging

TL;DR: A formulation of the scheme, according to which CEXSI can be extended to improve SNR in sparse array imaging in general, and its potential to achieve excellent decoding with very low (<80 dB) side-lobes is demonstrated.