Fatigue resistance of two implant/abutment joint designs.

Reads0
Chats0
TLDR
Within the limitations of this in vitro study, the effect of joint design on the fatigue strength and failure mode of the ITI single tooth implant system was significantly better than the Brånemark single-tooth implant system tested.
Abstract
Statement of Problem. Because of reported mechanical failures, alternative implant system components with suggested optimized strength have been manufactured. However, the endurance of these products has not been well investigated. Purpose. This study was designed to assess the effect of joint design on the fatigue strength and failure mode of 2 single-tooth implant systems: Branemark and ITI, in which a hex mediated-butt joint and 8-degree internal conical implant/abutment interface are used, respectively. Material and Methods. Seven 10-mm implants from each implant system were embedded to a depth of 7 mm in cylindrical acrylic resin blocks. CeraOne and Solid abutments with cement-retained castings were assembled to the Branemark and ITI implants, respectively. The assembled units were mounted in a lever-type-testing machine that was equipped with an automatic counting device and shutoff sensors, enabling the recording of the number of cycles till failure. A cyclic load of 100 N was applied perpendicular to the long axis of the assemblies at a rate of 75 cycles/min. To investigate specimen resistance to fatigue during 6 years of simulated function, a target of 1,800,000 cycles was defined. Specimen preparation and testing was performed by the same operator. The association of the joint design with the occurrence of failure was verified by Fisher's exact probability test ( P Results. For the Branemark group, the gold alloy abutment screw in all specimens fractured between 1,178,023 and 1,733,526 cycles with a standard deviation of 224,477 cycles. For the ITI group, all specimens had no failure until 1,800,000 cycles. Statistical analysis showed a highly significant difference between the 2 groups ( P =.000582). Conclusion. Within the limitations of this in vitro study, the effect of joint design on the fatigue strength and failure mode of the ITI single tooth implant system was significantly better ( P >.001) than the Branemark single-tooth implant system tested. (J Prosthet Dent 2002;88:604-10.)

read more

Citations
More filters
Journal ArticleDOI

A systematic review of the performance of ceramic and metal implant abutments supporting fixed implant reconstructions.

TL;DR: The 5-year survival rates estimated from annual failure rates appeared to be similar for ceramic and metal abutments, and a meta-analysis of the laboratory data was impossible due to the non-standardized test methods of the studies included.
Journal ArticleDOI

In vitro differences of stress concentrations for internal and external hex implant–abutment connections: a short communication

TL;DR: Within limitations of the model study, it was suggested that fixtures with internal-hex showed widely spread force distribution down to the fixture tip compared with external hex ones.
Journal ArticleDOI

Performance of conical abutment (Morse Taper) connection implants: a systematic review.

TL;DR: It is indicated that implant systems using a conical implant-abutment connection, provides better results in terms of abutment fit, stability, and seal performance and these design features could lead to improvements over time versus nonconical connection systems.
Journal ArticleDOI

Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system

TL;DR: Within the limitations of this study, reverse torque values of the screw joint were preserved under eccentric lateral loading, as compared with centric loading.
Journal ArticleDOI

Implant-abutment interface design affects fatigue and fracture strength of implants.

TL;DR: Implant systems with long internal tube-in-tube connections and cam-slot fixation showed advantages with regard to longevity and fracture strength compared with systems with shorter internal or external connection designs.
References
More filters
Book

Restorative dental materials

TL;DR: Restorative dental materials, Restorative dental Materials , کتابخانه دیجیتال جندی شاپور اهواز
Journal Article

Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws.

TL;DR: The long-term outcome of prostheses and fixtures (implants) in 759 totally edentulous jaws of 700 patients concur with multicenter and earlier results for the osseointegration method.
Journal Article

Survival of the Brånemark implant in partially edentulous jaws: a 10-year prospective multicenter study.

TL;DR: The Brånemark Implant System is a safe and predictable method for restoring partially edentulous patients, as demonstrated by this 10-year follow-up investigation.
Journal Article

Bending overload and implant fracture: a retrospective clinical analysis.

TL;DR: It was concluded that prostheses supported by one or two implants and replacing missing posterior teeth are subjected to an increased risk of bending overload.
Related Papers (5)