Fluence Beam Monitor for High-Intensity Particle Beams Based on a Multi-Gap Ionization Chamber and a Method for Ion Recombination Correction

TLDR
In this article , a multi-gap detector (MGD) composed of three parallel-plate ionization chambers (ICs) with different gap widths, assembled to prove the capability of correcting for charge volume recombination which is expected to occur when high fluence rates are delivered.
Abstract
This work presents the tests of a multi-gap detector (MGD), composed of three parallel-plate ionization chambers (ICs) with different gap widths, assembled to prove the capability of correcting for charge volume recombination which is expected to occur when high fluence rates are delivered. Such beam conditions occur with a compact accelerator for charged particle therapy developed to reduce the costs, to accomplish faster treatments and to exploit different beam delivery techniques and dose rates as needed, for example, for range modulation and FLASH irradiations, respectively. The MGD was tested with carbon ions at the Centro Nazionale di Adroterapia Oncologica (CNAO Pavia, Italy), and with protons in two different beam lines: at Bern University Hospital with continuous beams and at the Laboratori Nazionale del Sud (Catania, Italy) of the Italian National Center of Nuclear Physics (INFN) with pulsed beams. For each accelerator, we took measurements with different beam intensities (up to the maximum rate of ionization achievable) and changed the detector bias voltage (V) in order to study the charge collection efficiency. Charge recombination models were used to evaluate the expected collected charge and to measure the linearity of the rate of ionization with the beam fluence rate. A phenomenological approach was used to determine the collection efficiency (f1) of the chamber with thinnest gap from the relative efficiencies, f1/f2 and f1/f3, exploiting the condition that, for each measurement, the three chambers were exposed to the same rate of ionization. Results prove that two calibration curves can be determined and used to correct the online measurements for the charge losses in the ICs for recombination.

read more

Citations
More filters
Journal ArticleDOI

Technical note: Measurement of the bunch structure of a clinical proton beam using a SiPM coupled to a plastic scintillator with an optical fiber.

TL;DR: In this article , the authors evaluated the capabilities of a very compact fast plastic scintillator coupled with an optical fiber readout by a SiPM and electronics sensor which has been used to provide information on the time structure at the nanosecond level of a clinical proton beam.
References
More filters
Journal ArticleDOI

Initial Recombination of Ions

TL;DR: In the absence of forces other than the Coulomb attraction, the probability of escape equals the reciprocal of the Boltzmann factor as discussed by the authors, which is the proper procedure whenever the Langevin factor equals unity, as in gases at high pressures.
Journal ArticleDOI

AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams.

TL;DR: This protocol represents a major simplification compared to the AAPM's TG-21 protocol in the sense that large tables of stopping-power ratios and mass-energy absorption coefficients are not needed and the user does not need to calculate any theoretical dosimetry factors.
Journal ArticleDOI

Ionization measurements at very high intensities. Pulsed radiation beams.

TL;DR: In pulsed radiation beams the high ion density produced by each pulse encourages recombination between ions of opposite signs, so formulae are derived for the proportion of ions escaping recombination in plane parallel, cylindrical and spherical ionization chambers under a variety of different conditions.
Journal ArticleDOI

Dosimetric commissioning and quality assurance of scanned ion beams at the Italian National Center for Oncological Hadrontherapy

TL;DR: After successful dosimetric beam commissioning, quality assurance measurements performed during a 24-month period show very stable beam characteristics, which are therefore suitable for performing safe and accurate patient treatments.
Journal ArticleDOI

The saturation curve at high ionization intensity

TL;DR: In this paper, the form of the saturation curve is investigated theoretically and experimentally in a parallel plate ionization chamber and it is shown that very high ionization intensities can be measured in such a chamber, provided the spacing of the plates is small.
Related Papers (5)