Copper oxide-polyaniline nanofiber modified fluorine doped tin oxide (FTO) electrode as non-enzymatic glucose sensor

TLDR
In this article, a non-enzymatic glucose sensor based on copper oxide-polyaniline nanofiber modified fluorine doped tin oxide (CuO/PANI-NF/FTO) was introduced as a nonenzymatically glucose sensor.
Abstract
In this study, copper oxide-polyaniline nanofiber modified fluorine doped tin oxide (CuO/PANI-NF/FTO) electrode is introduced as a non-enzymatic glucose sensor. PANI-NFs were polymerized by a chemical method and through a cellulose ester membrane. These fibers were dip coated on FTO and then CuO particles were deposited on PANI-NFs by the electrodeposition method (CuO/PANI-NF). The morphology and structure were characterized by scanning electron microscopy (SEM), X-ray diffraction (XRD), energy-dispersive X-ray spectroscopy (EDS) and Fourier transform infrared (FTIR). The electrochemical response of the proposed electrode in presence of glucose was investigated in 0.1 M NaOH. In addition, comparing the performance of the CuO/PANI-NF/FTO with PANI-NF/FTO and the bare FTO indicated its superiority over the last two. The chronoamperometry (CA) linear range of the proposed sensor was between 0.25 μM to 4.6 mM with the detection limit of 0.24 μM (S/N = 3). The interference effect of some important species such as ascorbic acid, dopamine, uric acid, urea and sodium chloride was examined and no considerable interfering effect was observed. The repeatability and reproducibility of the electrode for glucose detection were 4.9 and 3.6%, respectively. Finally, the electrode stability was investigated and after 11 days, the electrode showed ∼90% of its initial signal. All results proved the proposed electrode excellence for non-enzymatic glucose sensing due to its simplicity, ease of preparation, wide linear range, appropriate selectivity and stability.

read more

Citations
More filters
Journal ArticleDOI

Electrochemical sensors and biosensors based on the use of polyaniline and its nanocomposites: a review on recent advances

TL;DR: The simultaneous use of suitable properties polyaniline and nanoparticles can provide the fabrication of sensing systems with high sensitivity, short response time, high signal/noise ratio, low detection limit, and wide linear range by improving conductivity and the large surface area for biomolecules immobilization.
Journal ArticleDOI

A Critical Review of Electrochemical Glucose Sensing: Evolution of Biosensor Platforms Based on Advanced Nanosystems.

TL;DR: The history of these biosensors based on commercialized systems, improvements in the understanding of the surface science for enhanced electron transfer, the various sensing platforms developed in the presence of the nanomaterials and their performances are described.
Journal ArticleDOI

A review of the advanced developments of electrochemical sensors for the detection of toxic and bioactive molecules

TL;DR: In this paper, the design and development of novel sensitive and cost-effective electrochemical sensors based on various classes of nanocomposites fabricated by different methods for the detection of biomolecules (glucose, dopamine, ascorbic acid and uric acid), heavy metal ions (Pb2+, Hg2+, Cd2+ and As3+) and environmental pollutants (hydrazine, nitrobenzene, nitrophenols and pesticides) are reviewed.
Journal ArticleDOI

A review on recent advances in application of electrospun nanofiber materials as biosensors

TL;DR: In this paper, the features of different types of electrospun biosensors and their developing trends were analyzed and discussed in a review article to share the current opinion with the relevant researchers on the applications of electro-spinning technology in biosensor.
References
More filters
Journal ArticleDOI

Electrochemical Glucose Biosensors

TL;DR: The major factors that play a role in the development of clinically accurate in-vivo glucose sensors include issues related to biocompatibility, miniaturization, long-term stability of the enzyme and transducer, oxygen deficit, short stabilization times, in- vivo calibration, baseline drift, safety, and convenience.
Journal ArticleDOI

Glucose oxidase: an ideal enzyme

TL;DR: The properties of glucose oxidase (GOD) are described in relation to the widespread use of this enzyme in biosensors, and the shortcomings of other enzymes that oxidize glucose are indicated.
Journal ArticleDOI

Polyaniline: The infrared spectroscopy of conducting polymer nanotubes (IUPAC Technical Report)

TL;DR: In this article, a group of par- ticipants from 11 institutions in different countries recorded the FTIR spectra of PANI bases prepared from the samples obtained in the solutions of strong and weak acids and in alkaline medium within the framework of an IUPAC project.
Journal ArticleDOI

Non-invasive glucose monitoring technology in diabetes management: a review.

TL;DR: This manuscript aims to review the various NGM techniques and devices and the challenges and future trends in NGM.
Journal ArticleDOI

Nonenzymatic amperometric determination of glucose by CuO nanocubes-graphene nanocomposite modified electrode.

TL;DR: A nonenzymatic amperometric glucose sensor based on copper oxide (CuO) nanocubes-graphene nanocomposite modified glassy carbon electrode ( CuO-G-GCE) with excellent electrocatalytic activity and high stability for glucose oxidation is reported.