scispace - formally typeset
Journal ArticleDOI

The application accuracy of the Pathfinder neurosurgical robot

TLDR
A new technique to measure the overall application accuracy of image-guided stereotactic systems has been used to measureThe Pathfinder neurosurgical robot, which uses a robot-mounted camera to triangulate fiducial skin markers, virtually eliminating line-of-sight problems experienced by other optical localisation methods.
Abstract
We have developed a new technique to measure the overall application accuracy of image-guided stereotactic systems. It has been used to measure the accuracy of the Pathfinder neurosurgical robot, which uses a robot-mounted camera to triangulate fiducial skin markers, virtually eliminating line-of-sight problems experienced by other optical localisation methods. A phantom consisting of plastic spheres on stalks was used. The targets are the centre of the spheres. Points on a sphere's surface were automatically selected on CT images, from which the centre of the sphere was calculated in image space to greater precision than the CT voxel dimensions. A depth gauge was attached to Pathfinder in place of a biopsy needle and driven to the target as defined on CT. Pathfinder then moved to several points over the sphere, moving perpendicular to the trajectory in a grid pattern. By considering the depth gauge reading, 3D points on the sphere's surface were measured in robot space and the centre of the sphere was found. The difference between the actual and planned targets yields the overall accuracy. Preliminary experiments show that the application accuracy of Pathfinder is 2.7 mm. Further calibrations of the robot are expected to improve this value significantly.

read more

Citations
More filters
Journal ArticleDOI

Medical Robots: Current Systems and Research Directions

TL;DR: By examining robotic systems across time and disciplines, trends are discernible that imply future capabilities of medical robots, for example, increased usage of intraoperative images, improved robot arm design, and haptic feedback to guide the surgeon.
Journal ArticleDOI

From Passive Tool Holders to Microsurgeons: Safer, Smaller, Smarter Surgical Robots

TL;DR: An overview of the significant achievements in surgical robotics is provided and the current trends and future research directions of the field in making surgical robots safer, smaller, and smarter are identified.
Journal ArticleDOI

Current Capabilities and Development Potential in Surgical Robotics

TL;DR: The widely deployed teleoperated manipulators aim to enhance human cognitive and physical skills and provide smart tools for surgeons, while image-guided robotics focus on surpassing human limitations by introducing automated targeting and treatment delivery methods.
Patent

System and method for optical position measurement and guidance of a rigid or semi-flexible tool to a target

Gilboa Pinhas
TL;DR: In this article, a system for measuring the position of a hand-held tool relative to a body in six degrees of freedom employs the camera (115) attached via mechanical linkage to the tool so that the camera moves together with the prosimal portion of the tool and has a field of view including the distal end.
Journal ArticleDOI

A Patient-Mounted Robotic Platform for CT-Scan Guided Procedures

TL;DR: A novel robotic assistant dedicated to medical interventions under computed tomography scan guidance that is built from an original 5 DOF parallel structure with a semispherical workspace, particularly well suited to CT-scan interventional procedures.
References
More filters
Journal ArticleDOI

A versatile camera calibration technique for high-accuracy 3D machine vision metrology using off-the-shelf TV cameras and lenses

TL;DR: In this paper, a two-stage technique for 3D camera calibration using TV cameras and lenses is described, aimed at efficient computation of camera external position and orientation relative to object reference coordinate system as well as the effective focal length, radial lens distortion, and image scanning parameters.
Book

A versatile camera calibration technique for high-accuracy 3D machine vision metrology using off-the-shelf TV cameras and lenses

Roger Y. Tsai
TL;DR: A new technique for three-dimensional camera calibration for machine vision metrology using off-the-shelf TV cameras and lenses using two-stage technique has advantage in terms of accuracy, speed, and versatility over existing state of the art.
Journal ArticleDOI

Least-Squares Fitting of Two 3-D Point Sets

TL;DR: An algorithm for finding the least-squares solution of R and T, which is based on the singular value decomposition (SVD) of a 3 × 3 matrix, is presented.
Journal ArticleDOI

The Application Accuracy of Stereotactic Frames

TL;DR: The results suggest a potentially significant degree of error in the application accuracy of all stereotactic instruments, which is accentuated by but not entirely due to imaging-associated errors.
Journal ArticleDOI

The Application Accuracy of the NeuroMate Robot—A Quantitative Comparison with Frameless and Frame-Based Surgical Localization Systems

TL;DR: The NeuroMate as mentioned in this paper is an image-guided, robotic-assisted system for brain surgery, which is a commercially available system used today for stereotactic procedures in neurosurgery.
Related Papers (5)