scispace - formally typeset
Open AccessJournal ArticleDOI

The feasibility of a piecewise-linear dynamic bowtie filter

Scott S. Hsieh, +1 more
- 01 Mar 2013 - 
- Vol. 40, Iss: 3, pp 031910-031910
Reads0
Chats0
TLDR
A dynamic bowtie producing a time-dependent, piecewise-linear attenuation profile is possible and can be used to modulate the flux of the scanner to the imaging task, and initial simulations show a large reduction in dynamic range.
Abstract
Purpose: The prepatient attenuator (or “bowtie filter”) in CT is used to modulate the flux as a function of fan angle of the x-ray beam incident on the patient. Traditional, static bowtie filters are tailored only for very generic scans and for the average patient. The authors propose a design for a dynamic bowtie that can produce a time-dependent piecewise-linear attenuation profile. This dynamic bowtie may reduce dynamic range, dose or scatter, but in this work they focus on its ability to reduce dynamic range, which may be particularly important for systems employing photon-counting detectors. Methods: The dynamic bowtie is composed of a set of triangular wedges. Each wedge is independently moved in order to produce a time-dependent piecewise-linear attenuation profile. Simulations of the bowtie are conducted to estimate the dynamic range reduction in six clinical datasets. The control of the dynamic bowtie is determined by solving a convex optimization problem, and the dose is estimated using Monte Carlo techniques. Beam hardening artifacts are also simulated. Results: The dynamic range is reduced by factors ranging from 2.4 to 27 depending on the part of the body studied. With a dynamic range minimization objective, the dose to the patient can be reduced from 6% to 33% while maintaining peak image noise. Further reduction in dose may be possible with a specific dose reduction objective. Beam hardening artifacts are suppressed with a two-pass algorithm. Conclusions: A dynamic bowtie producing a time-dependent, piecewise-linear attenuation profile is possible and can be used to modulate the flux of the scanner to the imaging task. Initial simulations show a large reduction in dynamic range. Several other applications are possible.

read more

Citations
More filters
Journal ArticleDOI

Vision 20/20: Single photon counting x-ray detectors in medical imaging

TL;DR: The authors offer the vision for the future ofPCD-CT and PCD-XR with the review of the current status and the prediction of detector technologies, imaging technologies, system technologies, and potential clinical benefits with PCDs.
Journal ArticleDOI

Recent and Future Directions in CT Imaging

TL;DR: The prediction is that significant further improvements in speed, spatial resolution and dose efficiency can be expected in the next decade.
Journal ArticleDOI

Energy-sensitive photon counting detector-based X-ray computed tomography

TL;DR: The performance of PCDs is not flawless, and thus, it requires integrated efforts to develop PCD-CT for clinical use, and the prediction for the future is reviewed.
Journal ArticleDOI

Regularization strategies in statistical image reconstruction of low‐dose x‐ray CT: A review

TL;DR: The conceptual and mathematical bases of these regularization strategies are reviewed and their efficacies in SIR of low-dose CT are illustrated.
Journal ArticleDOI

Variation in CT Number and Image Noise Uniformity According to Patient Positioning in MDCT

TL;DR: Radiologists should treat CT number measurements with caution when patients are grossly mispositioned in the scanner, and substantial changes in attenuation values shown in the present study are large enough to warrant further investigation.
References
More filters
Journal ArticleDOI

Geant4—a simulation toolkit

S. Agostinelli, +126 more
TL;DR: The Gelfant 4 toolkit as discussed by the authors is a toolkit for simulating the passage of particles through matter, including a complete range of functionality including tracking, geometry, physics models and hits.
Journal ArticleDOI

Computed Tomography — An Increasing Source of Radiation Exposure

TL;DR: The facts are summarized about CT scans, which involve much higher doses of radiation than plain films, and the implications for public health are summarized.
Book ChapterDOI

Graph Implementations for Nonsmooth Convex Programs

TL;DR: Graph implementations as mentioned in this paper is a generic method for representing a convex function via its epigraph, described in a disciplined convex programming framework, which allows a very wide variety of smooth and nonsmooth convex programs to be easily specified and efficiently solved.
Journal ArticleDOI

K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors.

TL;DR: This paper investigates to which extent energy-sensitive photon counting devices, operated in the pulse-mode, are capable of revealing quantitative information about the elemental composition of the absorber, and addresses the question of measuring and imaging the local density of a gadolinium-based contrast agent in the framework of a generalized dual-energy pre-processing.
Journal ArticleDOI

Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements

TL;DR: It is concluded that attenuation-based online tube current control has great potential for reducing patient dose in CT and that it should be made generally available for clinical use.