Effect of skin impedance on image quality and variability in electrical impedance tomography: a model study

K G Boone, +1 more
- 01 Sep 1996 - 
- Vol. 34, Iss: 5, pp 351-353
Reads0
Chats0
TLDR
It is concluded that the patient's skin should be abraded to reduce impedance, and measurements should be avoided in the first 10 min after electrode placement, to allow satisfactory images.
Abstract
A computer simulation is used to investigate the relationship between skin impedance and image artefacts in electrical impedance tomography. Sets of electrode impedance are generated with a pseudo-random distribution and used to introduce errors in boundary voltage measurements. To simplify the analysis, the non-idealities in the current injection circuit are replaced by a fixed common-mode error term. The boundary voltages are reconstructed into images and inspected. Where the simulated skin impedance remains constant between measurements, large impedances (> 2k omega) do not cause significant degradation of the image. Where the skin impedances 'drift' between measurements, a drift of 5% from a starting impedance of 100 omega is sufficient to cause significant image distortion. If the skin impedances vary randomly between measurements, they have to be less than 10 omega to allow satisfactory images. Skin impedances are typically 100-200 omega at 50 kHz on unprepared skin. These values are sufficient to cause image distortion if they drift over time. It is concluded that the patient's skin should be abraded to reduce impedance, and measurements should be avoided in the first 10 min after electrode placement.

read more

Citations
More filters
Proceedings ArticleDOI

Incorporating Multiple Observations in global Ultrasound Elastography

TL;DR: This paper proposes a novel framework for time delay estimation in ultrasound elastography using multiple observations (mGLUE), and primary validation of mGLUE against soft and hard inclusion simulation phantoms proves that m GLUE is capable of obtaining high quality strain map while dealing with noisy ultrasound data.
Journal ArticleDOI

A 30-MHz, 3-D Imaging, Forward-Looking Miniature Endoscope Based on a 128-Element Relaxor Array

TL;DR: This work describes the design, fabrication, and characterization of a 128-element crossed electrode array in a miniature endoscopic form factor for real-time 3-D imaging and generated preliminary 2-D images of a wire phantom using the previously described simultaneous azimuth and Fresnel elevation (SAFE) compounding imaging technique.
Journal ArticleDOI

Low-complexity adaptive minimum variance ultrasound beam-former based on diagonalization

TL;DR: Using LCAMV, not only the resolution and contrast are not changed significantly, but also the complexity is reduced from O (L 3 ) to O ( L × l o g ( L ) ) .
Proceedings ArticleDOI

Analysis of resolution for an amplitude steered array

TL;DR: The dependence of axial resolution on the length of the array is a unique feature of the amplitude-steered array, leading to an interesting tradeoff between lateral and axialresolution.
Journal ArticleDOI

Novel electrode–skin interface for breast electrical impedance scanning

TL;DR: The CFGTL-interface appears to be an effective electrode–skin interface for breast EIS examination and can decrease the ‘contact artifact’ and increase the accuracy of the examination.
References
More filters
Journal ArticleDOI

The Sheffield data collection system.

TL;DR: Any improvement in spatial resolution can only be made at the expense of speed and sensitivity which may well be the more important factors in determining the clinical utility of APT.
Journal ArticleDOI

Using compound electrodes in electrical impedance tomography

TL;DR: A sensitivity study showed that the reconstructed resistivity distributions are less dependent on the unknown contact resistance values for a compound electrode than a conventional electrode and that the use of a compound electrodes results in improved images for the reconstruction algorithm.
Journal ArticleDOI

Finite element method and reconstruction algorithms in electrical impedance tomography

TL;DR: To validate the method of data analysis, a theoretical model was developed, with the limiting case being the turn-on transient for a membrane covered spherical cathode, and the results agree with previously published data.

Assessment of noise and drift artefacts in electrical impedance tomography measurements using the sheffield Mark I system

K. G. Boone, +1 more
TL;DR: In this article, the mesures peuvent etre affectees d'artefacts dus au bruit et aux derives are discussed, e.g., les variations de l'impedance du fluide and celles des impedances d'electrode sont peu importantes.
Related Papers (5)