Effect of skin impedance on image quality and variability in electrical impedance tomography: a model study

K G Boone, +1 more
- 01 Sep 1996 - 
- Vol. 34, Iss: 5, pp 351-353
Reads0
Chats0
TLDR
It is concluded that the patient's skin should be abraded to reduce impedance, and measurements should be avoided in the first 10 min after electrode placement, to allow satisfactory images.
Abstract
A computer simulation is used to investigate the relationship between skin impedance and image artefacts in electrical impedance tomography. Sets of electrode impedance are generated with a pseudo-random distribution and used to introduce errors in boundary voltage measurements. To simplify the analysis, the non-idealities in the current injection circuit are replaced by a fixed common-mode error term. The boundary voltages are reconstructed into images and inspected. Where the simulated skin impedance remains constant between measurements, large impedances (> 2k omega) do not cause significant degradation of the image. Where the skin impedances 'drift' between measurements, a drift of 5% from a starting impedance of 100 omega is sufficient to cause significant image distortion. If the skin impedances vary randomly between measurements, they have to be less than 10 omega to allow satisfactory images. Skin impedances are typically 100-200 omega at 50 kHz on unprepared skin. These values are sufficient to cause image distortion if they drift over time. It is concluded that the patient's skin should be abraded to reduce impedance, and measurements should be avoided in the first 10 min after electrode placement.

read more

Citations
More filters
Journal ArticleDOI

Two-Stage Motion Correction for Super-Resolution Ultrasound Imaging in Human Lower Limb

TL;DR: This paper investigates the feasibility of a two-stage motion estimation method, which is a combination of affine and nonrigid estimation, for SR US imaging and reduces the width of the motion-blurred microvessels to approximately 1.5-fold.
Journal ArticleDOI

Full 2D displacement vector and strain tensor estimation for superficial tissue using beam-steered ultrasound imaging.

TL;DR: This study investigates beam steering at maximally three different angles to determine the full 2D strain tensor, and outperforms conventional methods that do not use beam steering.
Journal ArticleDOI

Estimates of echo correlation and measurement bias in acoustic radiation force impulse imaging

TL;DR: Acoustic radiation force impulse (ARFI) imaging is a novel imaging modality in which pulses from a diagnostic ultrasound scanner are used to displace tissue and track its motion, leading to decorrelation of the echo signal.
Journal ArticleDOI

High Frame-Rate, High Resolution Ultrasound Imaging With Multi-Line Transmission and Filtered-Delay Multiply And Sum Beamforming.

TL;DR: In MLT images, compared to standard Delay And Sum (DAS) beamforming including Tukey apodization, F-DMAS beamforming yields better suppression of cross-talk and improved lateral resolution, and preliminary in vivo cardiac images show that the frame rate can be improved up to 8-fold by combining F- DMAS and MLT without affecting the image quality.
Journal ArticleDOI

Rapid MR-ARFI method for focal spot localization during focused ultrasound therapy.

TL;DR: This work developed a single‐shot echo‐planar imaging–based MR‐ARFI method that reduces scan time and ultrasound energy deposition and demonstrates sufficient sensitivity to phase aberrations and can serve as rapid feedback for adaptive focusing in brain treatments and other applications.
References
More filters
Journal ArticleDOI

The Sheffield data collection system.

TL;DR: Any improvement in spatial resolution can only be made at the expense of speed and sensitivity which may well be the more important factors in determining the clinical utility of APT.
Journal ArticleDOI

Using compound electrodes in electrical impedance tomography

TL;DR: A sensitivity study showed that the reconstructed resistivity distributions are less dependent on the unknown contact resistance values for a compound electrode than a conventional electrode and that the use of a compound electrodes results in improved images for the reconstruction algorithm.
Journal ArticleDOI

Finite element method and reconstruction algorithms in electrical impedance tomography

TL;DR: To validate the method of data analysis, a theoretical model was developed, with the limiting case being the turn-on transient for a membrane covered spherical cathode, and the results agree with previously published data.

Assessment of noise and drift artefacts in electrical impedance tomography measurements using the sheffield Mark I system

K. G. Boone, +1 more
TL;DR: In this article, the mesures peuvent etre affectees d'artefacts dus au bruit et aux derives are discussed, e.g., les variations de l'impedance du fluide and celles des impedances d'electrode sont peu importantes.
Related Papers (5)