Effect of skin impedance on image quality and variability in electrical impedance tomography: a model study

K G Boone, +1 more
- 01 Sep 1996 - 
- Vol. 34, Iss: 5, pp 351-353
Reads0
Chats0
TLDR
It is concluded that the patient's skin should be abraded to reduce impedance, and measurements should be avoided in the first 10 min after electrode placement, to allow satisfactory images.
Abstract
A computer simulation is used to investigate the relationship between skin impedance and image artefacts in electrical impedance tomography. Sets of electrode impedance are generated with a pseudo-random distribution and used to introduce errors in boundary voltage measurements. To simplify the analysis, the non-idealities in the current injection circuit are replaced by a fixed common-mode error term. The boundary voltages are reconstructed into images and inspected. Where the simulated skin impedance remains constant between measurements, large impedances (> 2k omega) do not cause significant degradation of the image. Where the skin impedances 'drift' between measurements, a drift of 5% from a starting impedance of 100 omega is sufficient to cause significant image distortion. If the skin impedances vary randomly between measurements, they have to be less than 10 omega to allow satisfactory images. Skin impedances are typically 100-200 omega at 50 kHz on unprepared skin. These values are sufficient to cause image distortion if they drift over time. It is concluded that the patient's skin should be abraded to reduce impedance, and measurements should be avoided in the first 10 min after electrode placement.

read more

Citations
More filters
Proceedings ArticleDOI

3D ultrasound imaging performance of a row-column addressed 2D array transducer: a simulation study

TL;DR: This paper compares the imaging performance of a 128+128 element row-column addressed array with a fully addressed 16×16 2D array via simulations of the point spread function with Field II to show that the detail resolution can be more than doubled using the row- column addressed array instead of the fully addressed array.
Journal ArticleDOI

Towards real time 2D to 3D registration for ultrasound-guided endoscopic and laparoscopic procedures.

TL;DR: A new method for multimodality (preoperative CT and intraoperative US endoscopic images) registration to guide endoscopic interventions was developed and found to be efficient using clinically realistic datasets.
Journal ArticleDOI

Accuracy of Carotid Strain Estimates From Ultrasonic Wall Tracking: A Study Based on Multiphysics Simulations and In Vivo Data

TL;DR: It is concluded that deviations are present due to the complex 3-D vessel wall deformation, the presence of specular reflections and, to a lesser extent, the spatially varying beam profile, with the error depending on the phase in the cardiac cycle and the scanning location.
Journal ArticleDOI

Resolution and brightness characteristics of short-lag spatial coherence (SLSC) images

TL;DR: Results demonstrate that the resolution and contrast of coherence-based images depend on a range of parameters, but are generally superior to those of matched B-mode images under challenging imaging conditions.
Journal ArticleDOI

Multifrequency Interlaced CMUTs for Photoacoustic Imaging.

TL;DR: It is demonstrated that the low-frequency subarrays are able to better visualize diffuse agent distributions, while the high-frequencySubarrays offer fine-resolution imaging important for microvascular imaging and structural navigation.
References
More filters
Journal ArticleDOI

The Sheffield data collection system.

TL;DR: Any improvement in spatial resolution can only be made at the expense of speed and sensitivity which may well be the more important factors in determining the clinical utility of APT.
Journal ArticleDOI

Using compound electrodes in electrical impedance tomography

TL;DR: A sensitivity study showed that the reconstructed resistivity distributions are less dependent on the unknown contact resistance values for a compound electrode than a conventional electrode and that the use of a compound electrodes results in improved images for the reconstruction algorithm.
Journal ArticleDOI

Finite element method and reconstruction algorithms in electrical impedance tomography

TL;DR: To validate the method of data analysis, a theoretical model was developed, with the limiting case being the turn-on transient for a membrane covered spherical cathode, and the results agree with previously published data.

Assessment of noise and drift artefacts in electrical impedance tomography measurements using the sheffield Mark I system

K. G. Boone, +1 more
TL;DR: In this article, the mesures peuvent etre affectees d'artefacts dus au bruit et aux derives are discussed, e.g., les variations de l'impedance du fluide and celles des impedances d'electrode sont peu importantes.
Related Papers (5)