An expandable system has been developed to operate multiple probes for the atomic force microscope in parallel at high speeds. The combined improvements from parallelism and enhanced tip speed in this system represent an increase in throughput by over two orders of magnitude. A modular cantilever design has been replicated to produce an array of 50 cantilevers with a 200 μm pitch. This design contains a dedicated integrated sensor and integrated actuator where the cells can be repeated indefinitely. Electrical shielding within the array virtually eliminates coupling between the actuators and sensors. The reduced coupling simplifies the control electronics, facilitating the design of a computer system to automate the parallel high-speed arrays. This automated system has been applied to four cantilevers within the array of 50 cantilevers, with a 20 kHz bandwidth and a noise level of less than 50 A. For typical samples, this bandwidth allows us to scan the probes at 4 mm/s.

read more

Content maybe subject to copyright    Report

Automated parallel high-speed atomic force microscopy
S. C. Minne, G. Yaralioglu,
a)
S. R. Manalis, J. D. Adams, J. Zesch,
b)
A. Atalar,
a)
and C. F. Quate
c)
E. L. Ginzton Laboratory, Stanford University, Stanford, California 94305
~Received 27 January 1998; accepted for publication 2 March 1998!
An expandable system has been developed to operate multiple probes for the atomic force
microscope in parallel at high speeds. The combined improvements from parallelism and enhanced
tip speed in this system represent an increase in throughput by over two orders of magnitude. A
modular cantilever design has been replicated to produce an array of 50 cantilevers with a 200
m
m
pitch. This design contains a dedicated integrated sensor and integrated actuator where the cells can
be repeated indefinitely. Electrical shielding within the array virtually eliminates coupling between
the actuators and sensors. The reduced coupling simplifies the control electronics, facilitating the
design of a computer system to automate the parallel high-speed arrays. This automated system has
been applied to four cantilevers within the array of 50 cantilevers, with a 20 kHz bandwidth and a
noise level of less than 50 Å. For typical samples, this bandwidth allows us to scan the probes at
4 mm/s. © 1998 American Institute of Physics. @S0003-6951~98!01618-0#
As the semiconductor industry continues to shrink its
design rules, new advanced metrology and lithography tools
become increasingly important. There has been much recent
work on applying the well-known techniques of scanning
probe microscopy ~SPM! to these problems. The crux of
these efforts is to maintain the resolution and precision of the
probe-based system while increasing its capacity, in terms of
area scanned in a given time, to a level that is viable for
commercial purposes. The throughput of the scanning probe
microscope system has been enhanced by moving to parallel
operation with arrays, and by increasing the scanning speed
of each probe.
The parallel system is a direct approach to increasing the
speed of the SPM system since the throughput of a micro-
scope is increased in direct proportion to the number of tips
in an array. Integrating deflection, or force, sensors onto the
cantilever simplifies the operation of the parallel system be-
cause it reduces the necessary alignment and maintenance
required for operation. The piezoresistive atomic force mi-
croscope ~AFM! sensor, developed by Tortonese
1
et al., has
been incorporated into parallel arrays and has been used for
imaging and lithography.
2
Many other groups have adopted
the piezoresistive sensor for probe systems. Lutwyche
3
et al.
has fabricated and used two-dimensional arrays of piezore-
sistive cantilevers. Parallel piezoresistive cantilevers for data
storage applications have been fabricated by Chui
4
et al. and
Ried
5
et al. Parallel optical lever cantilever sensors have
been demonstrated by Lang
6
et al. for use in imaging and
other sensing applications.
In most probe systems, throughput is limited by the
speed at which the tip can be scanned over the sample sur-
face. For constant force imaging this speed is generally lim-
ited by the speed of the feedback system, which is in turn
limited by the speed, or resonance, of the z-axis actuator.
High-speed imaging has been previously done by Barrett
7
et al. and Mamin
8
et al. These systems serve as powerful
demonstrations but they are not total solutions. Barrett’s sys-
tem operated in a nonfeedback mode, and thus required hard
samples. Mamin’s system scanned the sample with a low-
displacement high-resonance actuator, and thus limited the
mass and topography of the sample. Manalis
9,10
and co-
workers have used both piezoresistive and optical sensing on
cantilevers with integrated actuators for high-speed imaging.
In this letter, we report an automated cantilever array
design, which allows parallel constant force imaging at high
speeds. The footprint of the cantilever structure has been
modified to occupy a slice only 200
m
m in width, allowing
the devices to be placed in a one-dimensional expandable
parallel array. Improved electrical performance of the device
allows us to use minimal nonsynchronous electronics, per-
mitting simple fabrication of the control system onto a per-
sonal computer ~PC! expansion card. The integrated elec-
tronics, coupled with the PC control, provides automated
operation over the array.
Previously, we showed that the scan speed of the AFM
could be increased by an order of magnitude over the stan-
dard AFM by integrating a thin layer of ZnO on the base of
a piezoresistive cantilever. The voltage applied to the ZnO
film bends the cantilever to conform to the sample topogra-
phy, and thus maintain constant force on the tip. The sample
force was detected by measuring resistance changes in the
piezoresistor. In that study, electrical coupling between the
ZnO actuator and the piezoresistive sensor required the use
of a synchronous system. Although this technique signifi-
cantly increased the imaging speed, unwanted capacitance in
the detection circuitry limited the bandwidth to 6 kHz. Ide-
ally, the imaging bandwidth should be limited by the me-
chanical resonance of the cantilever when the tip is in contact
with a surface. Ultimately, the design must have the versa-
tility to control an electrical interaction between the tip and
sample ~for lithography, modification, etc.!. Therefore, a pro-
vision to bias the tip without interfering with the sensing and
a!
Also at Bilkent University, Ankara, Turkey.
b!
Also at Xerox PARC, Palo Alto, CA 94304.
c!
Electronic mail: quate@stanford.edu
APPLIED PHYSICS LETTERS VOLUME 72, NUMBER 18 4 MAY 1998
23400003-6951/98/72(18)/2340/3/$15.00 © 1998 American Institute of Physics

actuating properties of the device must be included in the
design.
Our cantilever, designed to address these issues, is sche-
matically shown along with a brief process flow in Fig. 1.
Inspection of Fig. 1 reveals several key features in the de-
sign. First, the bottom electrode of the actuator ~between the
ZnO and the silicon substrate! serves as a grounding plane.
This prevents the signals that are applied to the top ZnO
electrode from coupling capacitively to the piezoresistor
through the silicon substrate. Second, the piezoresistor is de-
fined by a patterned implant and this allows the cantilever to
be actuated with a single pad of ZnO. In the previous design,
the piezoresistor was defined by physical etching, thereby
requiring the ZnO to be patterned on both of the cantilever
legs. We found that the dual leg ZnO configuration reduced
the overall yield of working devices. Third, grounding bus-
ses are added for common piezoresistor and common ZnO
electrodes. This minimized the tip-to-tip spacing by eliminat-
ing two electrical contacts per cantilever. Finally, a dedicated
voltage connection is made to the tip with a heavy implant so
the tipsample biases can be applied independent of the sen-
sor or actuator signals. A complete description the old de-
sign’s process, which was followed for this process, is given
in Ref. 9.
Micrographs showing the array of cantilevers with the
corresponding bonding pad layout are shown in Fig. 2. We
have fabricated an array of 50 cantilevers that are spaced by
200
m
m yielding an array spanning 1 cm. Figure 2~a! shows
the 1 cm array next to a dime for scale. Figure 2~b! shows a
detail of five cantilevers within the array. The piezoresistor
and ZnO regions are clearly seen. The two horizontal metal
lines running across the device are the piezoresistor and the
ZnO ground busses. The three contacts per cantilever ~pi-
ezoresistor sensor, ZnO actuator, and tip bias! run vertically
down the die to the bonding pads shown in Fig. 2~d!. Each
device occupies a horizontal footprint of 200
m
m, which is
the minimum achievable tip-to-tip spacing due to the design
limitations on the bonding pads. @The bonding services cur-
rently available to us can bond two rows of pads where each
row is on a 100
m
m period. This was the basis for our con-
tact configuration as shown in Fig. 2~d!.# A close up of the
integrated tip is shown in Fig. 2~c!. The radius of curvature
of these single-crystal silicon tips is generally less than 100
A. This tip was engineered using gas variations in a plasma
etch to produce a high aspect ratio profile that is useful for
both imaging and lithography. The tips are covered with a
thin film of titanium to enhance the capability for lithogra-
phy.
For small arrays of cantilevers it is possible to manually
control all aspects of the cantilever’s operation during an
imaging or lithography experiment. However, for massively
parallel operations, or operation in a scalable system, it is
important to have a compact electronic system for automated
control. In the piezoresistive/ZnO system, the position of the
cantilever arm is detected by microvolt changes in the pi-
ezoresistive sensor’s bridge circuit. This voltage is amplified
to tens of volts to generate a suitable signal for feedback and
analysis. Since the voltage gain is high, special consider-
ations need to be given to electronic noise and phase delays
in order to create a high-speed, sensitive, electronic system.
FIG. 1. Schematic illustration of the expandable cantilever design with in-
tegrated piezoresistive sensor and integrated ZnO actuator. The perspective
view on the left shows the final device, while the exploded view on the right
shows the pattern of all of the films in the device.
FIG. 2. ~a! and ~b! An array of cantilevers with integrated actuators and
sensors with improved shielding between the actuator and sensor. Part ~a!
shows an entire array of 50 cantilevers spanning 1 cm next to a dime ~for
scale!. Part ~b! shows a detail of five of the cantilevers. The cantilevers are
spaced by 200
m
m ~c! SEM of a typical integrated single-crystal silicon tip.
The radius of curvature is below 10 nm. ~d! corresponding electrical contact
structure for the cantilevers. There are three leads per device: piezoresistor,
ZnO, and tip bias.
FIG. 3. Schematic illustration of the automated parallel high-speed AFM. A
computer-controlled PC card contains the electronics to automatically con-
trol the operational parameters of the cantilevers shown in Fig. 2 ~set point,
gain, and feedback!. A second computer is used to collect the data and
control the scan.
2341Appl. Phys. Lett., Vol. 72, No. 18, 4 May 1998 Minne
et al.

Automation is achieved through computer control of the mi-
croscope’s operational parameters: force setpoint, gain, and
feedback. A schematic of the automated microscope is
shown in Fig. 3. Two computers are used for imaging. One
contains the custom electronic PC expansion board and is
responsible for monitoring and modifying the cantilevers op-
erational parameters. The other computer is responsible for
collecting data and controlling the scanner.
The transfer function of the electronic system is shown
in Fig. 4. Figure 4~a! is a simulation of the combined elec-
tronic and mechanical system. The cantilever was modeled
as a second-order system with a resonance of 50 kHz and a Q
of 100. These parameters were determined from the open-
loop transfer function of the cantilever when the tip is in
contact with a sample @Fig. 4~c!#. The electronic system was
modeled using the written specifications of each of the cir-
cuit components. Figure 4~b! shows the measured closed-
loop response of the cantilever and electronic system. We
found excellent agreement between the simulations and the
measured data. The primary limitation on bandwidth is due
to the resonance of the cantilever. The influence of the elec-
tronic phase can be seen in the closed-loop response in the
small hump before the peak in the amplitude trace. The
bandwidth of the amplifier chain is measured to be 1.1 MHz
and the phase shift at 100 kHz is less than 10°.
A431 image obtained with the cantilevers and control
electronics is shown in Fig. 5. A detail of a memory cell on
an integrated circuit chip is shown in Fig. 5~a!. The scan size
is just over 100
m
m3100
m
m, which is the limit of a con-
ventional piezotube. The ZnO actuator is able to track the 2
m
m high topography at 1 mm/s. The images do not overlap
because the tips are 200
m
m apart while the scan is only 100
m
m. Figure 5~b! shows a parallel image acquired at a scan
speed of 4 mm/s. The minimum detectable deflection for this
image is less than 5 nm in a 20 kHz bandwidth.
The authors would like to acknowledge Babur Hadimio-
glu and Nick Ulman. Support for this work came from
DARPA and SRC. S.C.M. and S.R.M. acknowledge support
from the Leland T. Edwards fellowship.
1
M. Tortonese, H. Yamada, R. C. Barrett, and C. F. Quate, The Proceed-
ings of Transducers ’91 ~IEEE, Pennington, NJ, 1991!, Publication No.
91 CH2817-5, p. 448.
2
S. C. Minne, Ph. Flueckiger, H. T. Soh, and C. F. Quate, J. Vac. Sci.
Technol. B 13, 1380 ~1995!.
3
M. Lutwyche, C. Andreoli, G. Binnig, J. Brugger, U. Drechsler, W. Hae-
berle, H. Rohrer, H. Rothuizen, and P. Vettiger, Proceedings IEEE Inter-
national Workshop on Microelectro Mechanical Systems ~MEMS 98!,
Heidelberg, Germany, Jan 2529, 1998.
4
B. W. Chui, T. D. Stowe, T. W. Kenny, H. J. Mamin, B. D. Terris, and D.
Rugar, Appl. Phys. Lett. 69, 2767 ~1996!.
5
R. P. Ried, H. J. Mamin, B. D. Terris, L. S. Fan, and D. Rugar, J. Micro-
electromech. Syst. 6, 294 ~1997!.
6
H. P. Lang, R. Berger, C. Andreoli, J. Brugger, M. Despont, P. Vettiger,
Ch. Gerber, J. K. Gimzewski, J. P. Ramseyer, E. Meyer, and H. J.
Guntherodt, Appl. Phys. Lett. 73, 383 ~1998!.
7
R. C. Barett and C. F. Quate, J. Vac. Sci. Technol. B 9, 302 ~1991!.
8
H. J. Mamin, H. Birk, P. Wimmer, and D. Rugar, J. Appl. Phys. 75, 161
~1994!.
9
S. R. Manalis, S. C. Minne, and C. F. Quate, Appl. Phys. Lett. 68, 871
~1995!.
10
S. R. Manalis, S. C. Minne, A. Atalar, and C. F. Quate, Rev. Sci. Instrum.
67, 3294 ~1996!.
FIG. 4. ~a! Simulated closed-loop transfer function. Simulation parameters
for the cantilever were measured, as shown in part c, and for the electronics
were taken from the data sheets. ~b! Measured closed-loop transfer function
~cantilever plus electronics!. The bandwidth of the system using a 40° phase
margin is 20 kHz. ~c! Open-loop transfer function ~cantilever only! of the
cantilever in contact with a smooth sample.
FIG. 5. 43 1 images of ~a! a memory cell of an integrated circuit and ~b!
two-dimensional grating both obtained in parallel with the cantilevers shown
in Fig. 2 and the electronics shown in Fig. 3. The bandwidth of the system
in 20 kHz, and the sensitivity is less than 5 nm in that band. The tip velocity
for the images is 1 and 4 mm/s for parts ~a! and ~b!, respectively.
2342 Appl. Phys. Lett., Vol. 72, No. 18, 4 May 1998 Minne
et al.
Citations
More filters
Journal ArticleDOI

A Survey of Control Issues in Nanopositioning

TL;DR: This paper presents an overview of nanopositioning technologies and devices emphasizing the key role of advanced control techniques in improving precision, accuracy, and speed of operation of these systems.
Journal ArticleDOI

The "millipede" - nanotechnology entering data storage

TL;DR: In this paper, a new scanning-probe-based data-storage concept called the "millipede" is presented, which combines ultrahigh density, terabit capacity, small form factor, and high data rate.
Journal ArticleDOI

The Millipede: more than one thousand tips for future AFM data storage

TL;DR: In addition to data storage in polymers or other media, and not excluding magnetics, this work envision areas in nanoscale science and technology such as lithography, high-speed/large-scale imaging, molecular and atomic manipulation, and many others in which Millipede may open up new perspectives and opportunities.
Journal ArticleDOI

Science and technology of nanomaterials: current status and future prospects

TL;DR: In this article, the authors examine the important facets of nanomaterials research, highlighting the current trends and future directions, since synthesis, structure, properties and simulation are important ingredients of nanoscience, materials chemists have a major role to play.
Journal ArticleDOI

AFM: a versatile tool in biophysics

TL;DR: It is hoped that this review can help the reader in appreciating how atomic force microscopy contributes to the current explosive growth of nanobiosciences, where biology, chemistry and physics merge.
References
More filters
Journal ArticleDOI

Sequential position readout from arrays of micromechanical cantilever sensors

TL;DR: In this article, the authors demonstrate that chemically specific responses can be extracted in a noisy environment using a sensor to detect specific chemical interactions and an uncoated cantilever as reference.
Journal ArticleDOI

Atomic force microscopy for high speed imaging using cantilevers with an integrated actuator and sensor

TL;DR: A cantilever with an integrated ZnO piezoelectric actuator in feedback with a piezoresistive sensor is utilized in an atomic force microscope (AFM) to achieve a new high speed imaging technique as discussed by the authors.
Journal ArticleDOI

Low‐stiffness silicon cantilevers for thermal writing and piezoresistive readback with the atomic force microscope

TL;DR: In this paper, a low-stiffness silicon cantilevers have been developed for proposed data storage devices based on the atomic force microscope, in particular thermomechanical recording, which combine a sharp tip with an integrated piezoresistive sensor for data readback from a rotating polycarbonate disk.
Journal ArticleDOI

Atomic force microscope lithography using amorphous silicon as a resist and advances in parallel operation

TL;DR: In this paper, a 400 μm×100 μm parallel image was obtained in the time it would normally take to obtain a 100 μm × 100μm image, using a parallel array of five piezoresistive cantilevers.
Journal ArticleDOI

6-MHz 2-N/m piezoresistive atomic-force microscope cantilevers with INCISIVE tips

TL;DR: In this paper, four-legged cantilevers with both imaging and thermomechanical surface modification capabilities have been used to write 2-Gb/cm/sup 2/ data at 50 kb/s on a spinning polycarbonate sample and to subsequently read the data.